| | Class 2 Device Recall VASOVIEW HEMOPRO (VH3500) Endoscopic Vessel Harvesting System |  |
| Date Initiated by Firm | May 31, 2024 |
| Date Posted | July 11, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2337-2024 |
| Recall Event ID |
94802 |
| 510(K)Number | K153194 |
| Product Classification |
Electrosurgical, cutting & coagulation & accessories - Product Code GEI
|
| Product | The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization. |
| Code Information |
UDI-DI: 00607567700345
Lot Numbers;
3000371260 3000378288 3000362934 3000371798 3000378554 3000372477
3000378950 3000376542 3000380314 3000377320 3000380847 |
| FEI Number |
2242352
|
Recalling Firm/ Manufacturer |
Maquet Cardiovascular, LLC 45 Barbour Pond Dr Wayne NJ 07470-2094
|
Manufacturer Reason for Recall | There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Maquet Cardiovascular, LLC (MCV), a subsidiary of Getinge issued Urgent Medical Device Removal Letter to consignees via FedEx 2-Day Delivery with Signature Proof of Delivery (SPOD) on 31 May 2024. Letter states reason for recall, health risk and action to take:
Please examine your inventory immediately to determine if you have any of the HemoPro
devices with the product code/lot number(s) listed in this notice.
" Prior to use, always refer to Instructions for Use, Section Warning and Precautions ,
specifically:
o HemoPro 2
Always advance the C-Ring and Harvesting Tool under endoscopic
visualization. Ensure adequate visualization of the HEMOPRO 2
Jaws and the surgical site prior to application of electrosurgical
energy. If visualization of the surgical site is impaired, do not
initiate or continue activation of energy to the HEMOPRO 2 Jaws
to avoid thermal injury to tissue.
o HemoPro:
Always advance the C-Ring and Harvesting Tool under
endoscopic visualization. CAUTION: Take care to avoid
engaging the components of the C-Ring assembly with the
VASOVIEW HEMOPRO jaws. Doing so may damage portions of
the C-Ring assembly causing the components to separate from
the assembly and fall into the body.
" Should you have any affected product as listed in this notification, do not use the
product and remove it from areas of use. You will receive credit upon the return of
any affected devices.
" Please forward this information to all current and potential HemoPro and HemoPro 2
users within your hospital / facility.
" If you are a distributor who has shipped any affected product(s) to customers, please
forward this document to their attention for appropriate action.
" Please contact Getinge Customer Service at 888-880-2874 between 6:00 a.m. and
5:00 p.m. Pacific Standard Time to request a return material authorization (RMA) and
shipping instructions to return any affected product. Pack the product to be returned
with the appropriate return documents and, using the shipping |
| Quantity in Commerce | 600 units |
| Distribution | NationwideForeign:
AT
DE
DK
ES
GB
HK
IT
NL
ZA
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = GEI
|
|
|
|