• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BioPince TruCore II

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BioPince TruCore IIsee related information
Date Initiated by FirmJuly 02, 2024
Date PostedSeptember 20, 2024
Recall Status1 Open3, Classified
Recall NumberZ-3192-2024
Recall Event ID 95119
510(K)NumberK904987 K982960 
Product Classification Instrument, biopsy - Product Code KNW
ProductBioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument, REF: 763114100x, 763116100X, 763116160X, 763118100X, 763118200X, 763120100X, 763120160X, 763418200X, 763418250X, 763114200X
Code Information REF/Box UDI-DI/Unit UDI-DI/Lot(Expiration Date): 360-1080-01/20886333004373/00886333004379/11563436(22-Mar-2027), 11568010(16-Apr-2027), 11568790(1-May-2027), 11569552(7-May-2027), 11572560(20-May-2027) 360-1080-02/20886333224009/00886333224005/11563678(22-Mar-27), 11562475(05-Mar-27), 11562476(14-Mar-27), 11566932(08-Apr-27), 11567139(11-Apr-27), 11568361(16-Apr-27), 11569013(23-Apr-27), 11569844(29-Apr-27), 11571094(02-May-27), 11572044(08-May-27), 11572886(26-May-27); 360-1080-03/20886333224016/00886333224012/11569845(22-Apr-27), 11562477(04-Mar-27), 11564861(26-Mar-27), 11567888(12-Apr-27), 11568554(18-Apr-27), 11570075(04-May-27); 360-1580-01/20886333004359/00886333004355/11563438(21-Mar-2027), 11563439(19-Mar-2027); 360-1580-02/20886333224023/00886333224029/11562478(14-Mar-27), 11563440(18-Mar-27), 11563682(20-Mar-27), 11565238(03-Apr-27), 11566882(09-Apr-27), 11570132(01-May-27), 11570814(09-May-27), 11573339(20-May-27); 360-1580-03/20886333224030/00886333224036/11570288(02-May-27), 11563442(20-Mar-27), 11566173(03-Apr-27), 11566883(10-Apr-27), 11568153(16-Apr-27), 11570813(08-May-27); 360-2080-01/20886333004342/00886333004348/11563564(26-Mar-2027), 11566036(8-Apr-2027), 11567992(16-Apr-2027), 11570142(1-May-27), 11570863(8-May-27); 360-2080-02/20886333224047/00886333224043/11562964(07-Mar-27), 11563922(15-Mar-27), 11564189(22-Mar-27); 360-2080-03/20886333224054/00886333224050/11563443(15-Mar-27), 11566936(11-Apr-27), 11567889(22-Apr-2027); 370-1080-01/20886333004380/00886333004386/11564795(26-Mar-27); 370-1080-02/20886333224061/00886333224067/11562608(05-Mar-27), 11562748(04-Mar-27), 11564843(26-Mar-27), 11566037(08-Apr-27), 11571101(30-Apr-27), 11571360(06-May-27); 370-1080-03/20886333224078/00886333224074/11564844(22-Mar-27), 11566690(16-Apr-27), 11571100(01-May-27), 11573030(13-May-27); 370-1580-01/20886333004366/00886333004382/11564366(26-Mar-27); 370-1580-02/20886333224085/00886333224081/11571824(06-May-27), 11562114(07-Mar-27), 11566038(08-Apr-27), 11571824(06-May-27); 370-1580-03/20886333224092/00886333224098/11569414(22-Apr-27), 11571104(01-May-27); 763114100x/20886333006643/00886333006649/11563673(11-May-2029), 11570356(23-Apr-2029), 11571058(4-Apr-2029), 11572456(29-May-2029), 11572868(13-May-2029), 11573300(22-May-2029); 763116100X/20886333006612/00886333006618/11562815(7-Mar-2029), 11571061(29-Apr-2029); 763116160X/20886333006605/00886333006601/11571063(1-May-2029); 763118100X/20886333006582/00886333006588/11564330(26-Mar-2029); 763118200X/20886333006650/00886333006656/11562085(5-Mar-2029), 11571065(4-APR-2029); 763120100X/20886333006568/00886333006564/11564331(26-Mar-2029); 763120160X/20886333006575/00886333006571/11565618(10-Apr-2029); 763418200X/20886333006803/00886333006809/11566213(10-Apr-2029); 763418250X/20886333006810/00886333006816/11564594(22-Mar-2029), 11565619(4-Apr-2029), 11572175(22-May-2029); 763114200X/20886333006629/00886333006625/11566106(4-Apr-2029)
Recalling Firm/
Manufacturer
Argon Medical Devices, Inc
1445 Flat Creek Rd
Athens TX 75751-5002
For Additional Information Contact
903-675-9321
Manufacturer Reason
for Recall
Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
FDA Determined
Cause 2
Packaging change control
ActionOn 7/2/24, recall notices were mailed and emailed to Physician, Clinician, or Hospital Administrators who were asked to do the following: 1) Discontinue use of affected devices. 2) Segregate, quarantine, and return affected devices to the recalling firm. 3) Share this recall notice with the following parties to ensure your consignees receive a copy of this notice and the associated response form: a. Anyone who needs to be aware within your organization b. Any organization where the potentially affected products have been transferred or further distributed c. Anyone within your organization who may use this device 4) Complete and return the response form via email to Arbee.cummings@argonmedical.com
Quantity in Commerce95,430
DistributionWorldwide - US Nationwide distribution including in the states of AZ, FL, OH, NY, MO, CO, IL, NC, CA, AK, PA, MN, ND, TX, SC, AR, WI, OK, TN, KS, UT, NJ, VA, MA, KY, IN, AL, DE, MT, NE, WV, NV, HI, OR, WA, NH, LA, CT, MI, GA, MD, ID, MS, IA, RI, NM, VT and the countries of ITA, TUN, ROU, LTU, BEL, GBR, CAN, TUR, POL, PAN, THA, PRT, TWN, AUT, NZL, AUS, BGR, MEX, CHE, NLD, PHL, PER, FRA, SGP, ZAF, CZE, EST, ESP, OMN, NPL, COL, PRI, ISR, JPN, KEN, BHR, ARE, CHN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KNW
510(K)s with Product Code = KNW
-
-