| | Class 2 Device Recall Impella CP Smart Assist |  |
| Date Initiated by Firm | August 05, 2024 |
| Date Posted | September 04, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2968-2024 |
| Recall Event ID |
95129 |
| Product Classification |
Temporary non-roller type left heart support blood pump - Product Code OZD
|
| Product | Impella 5.5 Set AU; Product Code: 1000361; |
| Code Information |
Product Code: 1000361; UDI-DI: 00813502012927; Serial Numbers: 425570
425332
425331
463954
463955
476358
476357
481131
481130
481129
485279
485278
485280
480079
483437
466172A
491829
491830
491831
491832; Batch Numbers: 2024286281
2024286289
2024286292
2024375505
2024375511
2024398832
2024398833
2024413728
2024413879
2024413882
2024431078
2024431158
2024431171
2024431175
2024431650
2024431652
2024431653
2024431656
2024431673
2024431720; |
Recalling Firm/ Manufacturer |
Abiomed, Inc. 24 Cherry Hill Dr Danvers MA 01923-2575
|
| For Additional Information Contact | Morgaine Johnson 612-437-5734 |
Manufacturer Reason for Recall | A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On August 5, 2024 URGENT MEDICAL DEVICE CORRECTION (NOTIFICATION) letters were sent to consignees. ACTIONS TO BE TAKEN BY CUSTOMER/USER:
" Product is NOT being removed from the field and does not need to be returned.
" Review, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 2) to Abiomed2860@SEDGWICK.COM.
" Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products).
" If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
" Post a copy of this notice in a visible area for awareness of this field safety notice.
" As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reporting-serious-problems-fda
At Abiomed, our priority is to our customers and their patients, and that includes the safe and effective use of our products. If you have questions or concerns regarding this notice, please contact your local clinical field staff. Thank you for your cooperation. |
| Quantity in Commerce | 20 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|