| | Class 2 Device Recall |  |
| Date Initiated by Firm | August 12, 2024 |
| Date Posted | October 03, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0009-2025 |
| Recall Event ID |
95215 |
| Product Classification |
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis - Product Code PER
|
| Product | BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V 1.4.1.
Intended use: to prepare human specimens to be analyzed. |
| Code Information |
REF/UDI-DI/Serial Numbers:
666340/382906663405/9230010101, 9230010102, 9230010103, 9230010104, 9230010105, 9230010107, 9230010108, 9230010109, 9230010112, 9230010113, 9230010114, 9230010115, 9230010116, 9230010118, 9230010119, 9230010120.
666339/9230000102, 9230000103, 9230000104, 9230000105, 9230000106, 9230000112, 9230000113, 9230000114, 9230000118, 9230000120, 9230000121, 9230000123, 9230000124, 9230000126, 9230000128, 9230000130, 9230000132, 9230000134, 9230000135, 9230000138, 9230000144, 9230000145, 9230000131, 9230000152, 9230000149, 9230000030, 9230000101, 9230000108, 9230000109, 9230000110, 9230000111, 9230000115, 9230000116, 9230000117, 9230000119, 9230000122, 9230000125, 9230000127, 9230000129, 9230000136, 9230000139, 9230000141, 9230000142, 9230000143.
663128/382906631282/9220030004, 9220030005, 9220030008, 9220030016, 9220030017, 9220030027, 9220030029, 9220030031, 9220030043, 9220030053, 9220030057.
662588/9220010119, 9220010179, 9220010182, 9220010232, 9220010236, 9220010250, 9220010265, 9220010275, 9220010293, 9220010311, 9220010313, 9220010316, 9220010324, 9220010327, 9220010335, 9220010338, 9220010341, 9220010344, 9220010348, 9220010349, 9220010292, 9220010112, 9220010120, 9220010125, 9220010127, 9220010128, 9220010130, 9220010133, 9220010138, 9220010154, 9220010161, 9220010183, 9220010188, 9220010197, 9220010215, 9220010225, 9220010231, 9220010238, 9220010240, 9220010242, 9220010244, 9220010256, 9220010257, 9220010260, 9220010263, 9220010268, 9220010269, 9220010279, 9220010280, 9220010289, 9220010290, 9220010294, 9220010296, 9220010329, 9220010332, 9220010345, 9220010354, 9230000031, 99220010292.
|
Recalling Firm/ Manufacturer |
Becton, Dickinson and Company, BD Bio Sciences 155 N Mccarthy Blvd Milpitas CA 95035-5102
|
Manufacturer Reason for Recall | Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy. |
FDA Determined Cause 2 | Software change control |
| Action | On 8/12/2024, correction notices were sent to customers who were informed of the following;
1) Cease use of the sample preparation system with software version 1.4.1. until the
Multidispense feature has been disabled. Patient samples can be prepared using single dispense on the sample preparation system or manually.
2) It is recommended that laboratories upgrade affected sample preparation systems to version 1.4.2 to correct the issue as soon as it is available.
3) Ensure the contents of this notification are read and understood by those within your organization.
4) Complete and return the response form via email to BDRC38@bd.com
If you require further assistance, contact: 1-844-823-5433, Mon-Fri 8:00am and 5:00pm CT or email productcomplaints@bd.com. Recall related questions can be emailed to BDRC38@bd.com |
| Quantity in Commerce | 130 |
| Distribution | US: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA
OUS: Ireland, Austria, Italy, Germany, Belgium, Finland, Canada, France, Spain, New Zealand, United Kingdom, Japan, Australia, Norway, Singapore, Sweden, Turkey, Brazil, Colombia, Netherlands, United Arab Emirates, Saudi Arabia, Mexico, Switzerland, Costa Rica, China, Azerbaijan, Algeria, Jordan, India
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|