| | Class 2 Device Recall 25LHP928, HeNe Laser System, 25 mW |  |
| Date Initiated by Firm | March 01, 2024 |
| Date Posted | September 10, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2976-2024 |
| Recall Event ID |
95290 |
| Product Classification |
General purpose laser products - Product Code RDW
|
| Product | 25-LHP-928, HeNe Laser System, 25 mW |
| Code Information |
25-LHP-928 |
| FEI Number |
3015195092
|
Recalling Firm/ Manufacturer |
PACIFIC LASERTECH, LLC 215 Bingham Dr # 110 San Marcos CA 92069-1403
|
Manufacturer Reason for Recall | These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position. |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | The firm notified all purchasers of the defect. Customers were instructed to return all units for replacement of the switch free of charge. |
| Quantity in Commerce | 4 |
| Distribution | US Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|