| | Class 2 Device Recall medex TranStar |  |
| Date Initiated by Firm | September 24, 2024 |
| Date Posted | October 23, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0192-2025 |
| Recall Event ID |
95438 |
| Product Classification |
Probe, blood-flow, extravascular - Product Code DPT
|
| Product | medex TranStar 84in (213cm) Single Monitoring Kit 10/EA, Product Code REF MX9505T |
| Code Information |
UDI/DI 50351688503990 (case), 10351688503992 (pouch), Lot Numbers: 4277670, 4287645, 4287649, 4299776, 4277696, 4308930, 4299811, 4308934, 4299812, 4323148, 4264943, 4264939, 4277694, 4264950, 4299779, 4308962, 4323175, 4328515, 4299788, 4299789, 4299790, 4299801, 4299802, 4308954, 4308959, 4308963, 4321917, 4308947, 4334499, 4299786, 4355732, 4355735, 4346115, 4299780, 4355730, 4355727, 4334501, 4346096, 4346097, 4346124, 4355733, 4346109, 4355728, 4355740, 4355750, 4355749, 4355746, 4346120, 4355757, 4355753 |
| FEI Number |
3012307300
|
Recalling Firm/ Manufacturer |
Smiths Medical ASD, Inc. 6000 Nathan Ln N Minneapolis MN 55442-1690
|
| For Additional Information Contact | Mrs. Mary J. Ferreira 845-907-1429 |
Manufacturer Reason for Recall | Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip. |
FDA Determined Cause 2 | Process control |
| Action | Smiths medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 09/24/2024 via traceable mailing. The notice explained the issue, potential risk, and requested the destruction of all affected product. Distributors were directed to forward the notice to their customers. |
| Quantity in Commerce | 1160168 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|