 Class 2 Device Recall Chameleon PTA Balloon Cathetersee related information
Date Initiated by FirmSeptember 19, 2024
Date PostedOctober 29, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0257-2025
Recall Event ID 95447
510(K)NumberK163231 
Product Classification Catheter, angioplasty, peripheral, transluminal - Product Code LIT
ProductChameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm x 75cm (CFN CH07-40-75US), 8mm x 40mm x 75cm (CFN CH08-40-75US), 9mm x 40mm x 75cm (CFN CH09-40-75US), 10mm x 40mm x 75cm (CFN CH10-40-75US), 12mm x 40mm x 75cm (CFN CH12-40-75US)
Code Information 1. 5mm x 40mm x 75cm (CFN CH05-40-75US): UDI-DI: 10884521812260; Lot Numbers: 22201047, 22201424, 22201425, 22202802, 22203008, 22203009, 22203040, 22300536, 22300970, 22300971, 22301116, 22301270, 22400129, 22400572, 22400573, Added 12/13/2024: 22202643, 22300000, 22300969, 22301117. Expansion 12/05/2024 with additional serial numbers: 22202643, 22300000, 22300969, 22301117. 2. 6mm x 40mm x 75cm (CFN CH06-40-75US): UDI-DI: 10884521812277; Lot Numbers: 22201337, 22201908, 22202412, 22202413, 22202611, 22202612, 22202801, 22300151, 22300318, 22300319, 22300320, 22300862, 22300965, 22301285, 22301494, 22301661, 22301895, 22301896, 22302180, 22400243, 22400244, 22400574, 22400576. 3. 7mm x 40mm x 75cm (CFN CH07-40-75US): UDI-DI: 10884521812284; Lot Numbers: 22200635, 22201897, 22202223, 22202426, 22202427, 22202800, 22203055, 22300321, 22300864, 22301028, 22301648, 22301649, 22301889, 22301890, 22302033, 22302034. 4. 8mm x 40mm x 75cm (CFN CH08-40-75US): UDI-DI: 10884521812291; Lot Numbers: 22201015, 22201046, 22201087, 22201785, 22201786, 22201813, 22202144, 22202145, 22202644, 22202645, 22300165, 22300230, 22300534, 22300535, 22300876, 22300877, 22300966, 22301133, 22301134, 22301299, 22301495, 22302032, 22400154, 22400164, 22400165. 5. 9mm x 40mm x 75cm (CFN CH09-40-75US): UDI-DI: 10884521812307; Lot Numbers: 22201364, 22202428, 22203026, 22300152, 22300539, 22301123, 22301124, 22400698, 22400699, 22400858. 6. 10mm x 40mm x 75cm (CFN CH10-40-75US): UDI-DI: 10884521812314; Lot Numbers: 22201814, 22201871, 22202646, 22300185, 22300219, 22300231, 22300863, 22301300, 22301301, 22302026, 22302031, 22302044, 22400153, 22400247, Added 12/13/2024: 22200739, 22201815, 22202176, 22202177, 22202799. Expansion 12/05/2024 with additional serial numbers: 22200739, 22201815, 22202176, 22202177, 22202799. 7. 12mm x 40mm x 75cm (CFN CH12-40-75US): UDI-DI: 10884521812321; Lot Numbers: 22201391, 22201872, 22202781, 22300301, 22300302, 22300878, 22300967, 22300968, 22301269, 22400128, 22400279, Added 12/13/2024: 22202212, 22202213, 22203025, 22300300. Expansion 12/05/2024 with additional serial numbers: 22202212, 22202213, 22203025, 22300300.
Recalling Firm/
Manufacturer
Covidien LP
15 Hampshire St
Mansfield MA 02048-1113
Manufacturer Reason
for Recall
Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn "URGENT: MEDICAL DEVICE RECALL" notification letter dated September 2024 was sent to customers. Required Actions: 1. Quarantine and discontinue use of unused affected product lot numbers immediately (see Attachment A). 2. Return all unused products from your inventory to Medtronic. Please contact: rs.covidienfeedbackcustomerservice@medtronic.com for the Return Good Authorization (RGA). a. Credit for the returned affected product will be issued based on the RGA number. b. If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. 3. Complete and return the enclosed Consignee Confirmation Product Retrieval to rs.gmbmitgfca@medtronic.com even if you do not have unused inventory. Please copy your local sales representative on this email when possible. 4. Share this notice with those who need to be aware within your organization or to any organization including, but not limited to, surgeons, interventionalists, or other staff. Please maintain a copy of this notice in your records. If you have questions regarding this communication, please contact your local representative or Customer Service at 800-962-9888, option 2 (Monday Friday 8:00 a.m. to 6:30 p.m.).
Quantity in Commerce16,336 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Canada, Canary Islands, China, France, French Guiana, Germany, Greece, Guadeloupe, India, Ireland, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Spain and Turkey.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.