| | Class 1 Device Recall Atlan |  |
| Date Initiated by Firm | October 15, 2024 |
| Date Posted | December 18, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0667-2025 |
| Recall Event ID |
95759 |
| 510(K)Number | K230931 |
| Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
| Product | Brand Name: Atlan
Product Name: Atlan A300
Model/Catalog Number: 8211300
The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. |
| Code Information |
Model/Catalog Number: 8211300;
UDI-DI Number: 04048675556343;
Serial Numbers shipped directly from Draegerwerk AG & Co., KGaA to OUS:
ASTA-0314; 0315; 0325; 0338; 0340; 0384; 0411; 0425
ASTB-0002; 0010; 0018; 0151; 0160; 0161; 0165; 0175; 0176; 0177; 0178; 0179; 0180; 0181; 0182; 0214; 0215; 0218; 0232; 0233; 0234; 0235; 0236; 0248; 0249; 0250; 0262; 0263; 0264; 0265; 0266; 0276; 0277; 0278; 0279; 0280
ASTC-0008; 0025; 0036; 0037; 0038; 0039; 0040; 0041; 0042; 0046; 0047; 0048; 0049; 0050; 0062; 0063; 0064; 0077; 0078; 0079; 0080; 0081; 0089; 0090; 0091; 0092; 0093; 0102; 0103; 0104; 0123; 0124; 0125; 0126; 0127; 0133; 0139; 0140; 0150; 0151; 0152; 0153; 0154; 0169; 0176; 0177; 0178; 0179; 0182; 0183; 0184; 0185; 0186; 0187; 0192; 0195; 0196; 0200; 0201; 0202; 0203; 0242
ASTD-0028; 0051; 0052; 0080; 0081; 0082; 0083; 0084; 0085; 0086; 0087
ASTE-0012; 0013; 0014; 0015; 0041; 0042; 0043; 0044; 0063; 0064; 0065; 0084; 0176; 0177; 0178; 0179; 0243; 0244; 0282; 0299
ASTF-0005; 0006; 0019; 0027; 0028; 0029; 0048; 0049; 0050; 0072; 0073; 0074; 0075; 0082; 0093; 0110; 0111; 0162; 0166; 0167; 0197; 0198; 0220; 0221; 0222; 0301; 0406; 0434
ASTH-0001; 0002; 0003; 0024; 0025; 0041; 0042; 0043; 0044; 0097; 0098; 0099; 0115; 0116; 0117; 0118; 0149; 0150; 0151; 0152; 0153; 0195; 0196; 0197; 0198; 0220; 0221; 0222; 0227; 0228; 0229; 0230; 0257; 0258; 0259; 0260; 0261; 0274; 0275; 0276; 0277; 0297; 0298; 0299; 0304; 0305; 0328; 0329; 0355; 0356; 0357; 0393; 0394; 0395; 0396; 0407; 0408; 0409; 0410; 0423; 0424; 0425; 0426; 0427; 0428; 0445; 0446; 0447; 0448; 0449; 0465; 0466; 0467; 0468; 0469; 0470; 0484; 0485; 0486; 0487
ASTK-0033; 0034; 0035; 0062
|
Recalling Firm/ Manufacturer |
Draeger Medical, Inc. 3135 Quarry Rd Telford PA 18969-1042
|
| For Additional Information Contact | Michael Kelhart 01-267-6641131 |
Manufacturer Reason for Recall | The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use. |
FDA Determined Cause 2 | Process control |
| Action | On October 15, 2024 Urgent Medical Device Correction letters were sent to customers. Necessary Action: The corrective action is to replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly replacement. Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Monitor the patient s condition specifically their oxygenation status as a brief cessation of ventilation can potentially lead to e.g.; hypoxia, loss of lung recruitment, bradycardia and cardiac arrest. If preferred, the alarm priority of the "Ventilator error!!!" can be downgraded with "ALARM RESET" after switching to Man/Spont ventilation mode. Please ensure that all users of the Draeger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. Complete and return the Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Draeger Service Technical Support 8AM-8PM EST at 1-800-437-2437 (press 2, then 2, then 2 again). Questions regarding this Urgent Medical Device Correction notice, contact Michael Kelhart 8AM-4:30PM EST at 267-664-1131 or via email at mike.kelhart@draeger.com. |
| Quantity in Commerce | 280 units (OUS only) |
| Distribution | No Atlan A300 or A300XL devices were imported into the US.
Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries:
Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = BSZ
|
|
|
|