| | Class 1 Device Recall Medtronic Pipeline Vantage Embolization Device with Shield Technology |  |
| Date Initiated by Firm | January 30, 2025 |
| Date Posted | March 05, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1245-2025 |
| Recall Event ID |
96168 |
| PMA Number | P100018 |
| Product Classification |
Intracranial aneurysm flow diverter - Product Code OUT
|
| Product | Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30, PED3-027-450-12, PED3-027-450-14, PED3-027-450-16, PED3-027-450-18, PED3-027-450-20, PED3-027-450-25, PED3-027-450-30, PED3-027-450-40, PED3-027-500-14, PED3-027-500-16, PED3-027-500-18, PED3-027-500-20, PED3-027-500-25, PED3-027-500-30, PED3-027-500-40, PED3-027-550-16, PED3-027-550-18, PED3-027-550-20, PED3-027-550-30, PED3-027-550-40, PED3-027-550-50, PED3-027-600-16, PED3-027-600-18, PED3-027-600-20, PED3-027-600-30, PED3-027-600-40, PED3-027-600-50 |
| Code Information |
All unused devices. Product lots including and between B317266 and B817292.
REF/UDI-DI/Lot:
PED3-027-350-12/00847536035263, 00847536030770/B409106, B637287, B700548;
PED3-027-350-14/00847536035287, 00847536030787/B409234, B700550, B794880;
PED3-027-350-16/00847536035294, 00847536030794/B409692, B697962, B700551, B795136, B795142, B809808;
PED3-027-350-20/00847536035317, 00847536030817/B410645, B558444, B623124, B672921, B700552, B777630;
PED3-027-350-25/00847536035324, 00847536030824/B410718, B672530, B700553, B771934, B771938;
PED3-027-400-12/00847536030862, 00847536035362, 00847536036680/B321585, B336094, B349668,B371137, B409814, B414591, B437740, B482271, B594119, B597085, B598465, B612514, B616043, B616052, B647374, B649853, B655648, B658620, B658685, B666385, B666418, B671805, B720634, B720635, B744619, B744621, B756639, B765446, B765448, B780611, B786108;
PED3-027-400-14/00847536030879, 00847536035379, 00847536036697/B318079, B335275, B339540, B350391, B364366, B372416, B375523, B410879, B415080, B422400, B438538, B448437, B459694, B460349, B464650, B469525, B474550, B556978, B601007, B608852, B616050, B622657, B625452, B625453, B629552, B683800, B689019, B691414, B698756, B714840, B715051, B721683, B745065, B753732, B763635, B763642, B802419;
PED3-027-400-16/00847536030886, 00847536035386, 00847536036703/B326193, B330879, B330881, B357080, B401867, B402034, B403012, B417737, B419736, B420127, B422990, B423257, B426548, B438677, B439156, B439301, B445198, B449040, B449149, B449591, B453559, B454843, B459846, B460523, B465307, B469682, B474417, B560503, B575892, B578340, B597083, B603196, B608221, B611268, B615571, B623135, B623136, B625944, B625945, B672865, B678987, B694225, B694934, B699098, B715043, B721778, B721782, B730828, B730829, B738422, B738423, B739914, B744842, B744860, B766155, B766701, B769315, B776922, B776933, B777378, B791207;
PED3-027-400-18/00847536036710/B795091;
PED3-027-400-20/00847536030909, 00847536035409, 00847536036727/B318459, B329314, B329316, B338468, B368540, B374265, B380481, B381434, B383600, B384092, B384233, B388982, B391816, B420182, B420318, B423735, B439714, B441852, B449699, B450256, B460915, B461092, B461510, B465420, B470193, B475106, B478030, B485264, B487677, B557537, B565556, B571634, B575122, B576711, B577356, B578341, B631830, B632949, B635238, B636674, B649884, B654937, B657799, B658273, B658540, B659188, B659211, B666546, B671140, B679007, B680109, B692381, B692536, B692916, B693469, B705994, B722574, B722592, B730946, B739972, B741664, B741666, B746709, B752946, B753532, B756807, B756808, B763701, B778174, B778192, B778193, B778534, B778535, B781327, B781347, B791327, B811796;
PED3-027-400-25/00847536030916, 00847536035416, 00847536036734/B322422, B394134, B400793, B411501, B415161, B424676, B432804, B442338, B470284, B474964, B481516, B483308, B568647, B576712, B577358, B634985, B636192, B648012, B649383, B661189, B675951, B689989, B704525, B706229, B706230, B709931, B712123, B726423, B741049, B741051, B747537, B747541, B754030, B765519, B776777, B777166, B780003, B780520, B786557;
PED3-027-400-30/00847536030923, 00847536035423, 00847536036741/B394830, B401265, B413855, B415343, B415743, B444760, B470728, B475566, B484529, B594121, B603201, B608917, B616555, B622661, B623126, B637143, B675952, B681380, B681596, B711437, B711438, B723555, B731515, B731516, B741809, B764098, B778206, B778207, B798796;
PED3-027-450-12/00847536030961, 00847536035461, 00847536036789/B322748, B456666, B471328, B486866, B575963, B601277, B637185, B666423, B703226, B706228, B708791, B711095, B750794, B751129, B758665, B758666, B758667, B794235;
PED3-027-450-14/00847536030978, 00847536035478, 00847536036796/B320465, B336565, B340244, B351032, B374880, B404541, B425289, B432042, B462206, B470900, B477408, B481700, B483148, B601181, B604476, B617556, B637191, B669563, B669568, B675953, B692841, B694794, B696125, B697448, B704465, B709288, B711828, B719080, B719084, B741791, B745142, B745143, B751967, B753463, B772618, B814124;
PED3-027-450-16/00847536030985, 00847536035485, 00847536036802/B318562, B331383, B331528, B340320, B351038, B374958, B384705, B385273, B395007, B405061, B407737, B424599, B450972, B465971, B471455, B483899, B485855, B535234, B552285, B568744, B571592, B585330, B596585, B622035, B625593, B626044, B628700, B629503, B643420, B659204, B659718, B660616, B661809, B666550, B666589, B668758, B669369, B675954, B682011, B698182, B723071, B723072, B723787, B723791, B735828, B735829, B735830, B746751, B746752, B753483, B757640, B757645, B762226, B762228, B764394, B787789, B789383, B803301;
PED3-027-450-18/00847536036819/B797308;
PED3-027-450-20/00847536031005, 00847536035508, 00847536036826/B325710, B328706, B328709, B355336, B357077, B357672, B377120, B381729, B389289, B390355, B397986, B419639, B421094, B423116, B425691, B440190, B440356, B440859, B442521, B449796, B450384, B451092, B452205, B461652, B462339, B464094, B480010, B481041, B487030, B549631, B550980, B557508, B569276, B573461, B574475, B584490, B585331, B594727, B632801, B635651, B650035, B651312, B651584, B652220, B669427, B673379, B674005, B681938, B696501, B696960, B700714, B700961, B703735, B705508, B728234, B729519, B732220, B752592, B758912, B758914, B760821, B774880, B776153, B794615, B794620, B798765, B810476, B811668;
PED3-027-450-25/00847536031012, 00847536035515, 00847536036833/B319170, B352790, B370633, B395978, B403146, B408241, B419571, B425960, B443719, B453438, B453676, B463202, B467023, B471895, B472002, B488714, B558531, B590206, B596591, B597131, B600999, B604474, B604478, B611115, B613070, B616042, B645444, B645874, B646698, B652027, B654357, B681223, B681901, B693373, B696977, B697584, B699639, B699762, B730899, B751210, B751305, B752452, B759553, B759554, B775532, B800797;
PED3-027-450-30/00847536031029, 00847536035522, 00847536036840/B396144, B404416, B411353, B434096, B436536, B535757, B561797, B574352, B616037, B644866, B645276, B651451, B652656, B654460, B658656, B659930, B675955, B688941, B713232, B714797, B714798, B717097, B730298, B731601, B731603, B784372, B785363, B810615, B815818;
PED3-027-450-40/00847536031043, 00847536035546, 00847536036864/B353358, B377744, B411399, B448343, B549634, B559739, B615408, B623125, B637017, B675956, B683288, B684783, B699735, B699776, B704439, B710043, B710051, B733018, B761384, B761385, B761386, B764292, B778738, B778836, B786518;
PED3-027-500-14/00847536031067, 00847536035577, 00847536036895/B353371, B411952, B425824, B432210, B447358, B455035, B455951, B601160, B603379, B606137, B624428, B637037, B651443, B674478, B675957, B685652, B701387, B706116, B706164, B710564, B710571, B711705, B729531, B764751, B765289, B785958, B790520;
PED3-027-500-16/00847536031074, 00847536035584, 00847536036901/B320910, B353996, B375682, B425116, B426932, B432683, B441009, B479353, B536475, B558194, B568648, B622122, B622123, B638485, B638487, B644714, B645376, B647467, B649263, B652167, B652663, B654363, B655727, B683524, B703758, B704315, B704999, B710455, B729572, B730327, B785458, B785792, B785793, B807437, B809811;
PED3-027-500-18/00847536036918/B797650;
PED3-027-500-20/00847536031098, 00847536035607, 00847536036925/B326325, B331608, B341953, B354635, B370450, B377872, B381636, B386124, B396652, B416028, B426418, B443827, B456058, B457830, B466125, B472700, B476324, B479966, B485115, B566387, B590194, B596594, B603380, B645491, B645898, B645962, B647222, B650050, B651347, B677890, B678397, B678959, B679068, B684999, B695686, B705596, B705610, B707567, B707606, B717655, B727541, B734039, B737927, B737928, B752681, B763087, B772659, B773371, B773372, B779449, B779450, B779633;
PED3-027-500-25/00847536031104, 00847536035614, 00847536036932/B354639, B382191, B387287, B396748, B413971, B417560, B472568, B478679, B536229, B570707, B603199, B646445, B648039, B680655, B688975, B689445, B693010, B694374, B695076, B700554, B702913, B720212, B720213, B720231, B747329, B747330, B747332, B751359, B761968, B763427, B800923;
PED3-027-500-30/00847536031111, 00847536035621, 00847536036949/B337320, B342643, B358320, B378426, B382386, B382844, B401379, B427074, B428049, B428522, B429875, B435865, B436340, B445405, B452880, B456544, B596584, B598708, B606136, B623128, B655004, B655034, B655744, B663561, B666422, B668512, B669368, B669370, B707244, B728433, B743037, B743038, B743039, B750658, B751192, B752314, B752315, B779538;
PED3-027-500-40/00847536031135, 00847536035645, 00847536036963/B319346, B340946, B373012, B373062, B374156, B379194, B397871, B398496, B456742, B457215, B464764, B467126, B473757, B482175, B567777, B573469, B585305, B590192, B677757, B682747, B693594, B701050, B703777, B703949, B707108, B710395, B711021, B711022, B736159, B747948, B749861, B753183, B753184, B762889, B773500;
PED3-027-550-16/00847536031166, 00847536035683, 00847536037007/B358321, B369526, B452971, B497779, B575962, B580922, B622743, B623127, B686266, B689577, B707045, B707126, B708154, B708159, B720912, B720913, B739371, B749920, B782490, B806751;
PED3-027-550-18/00847536037014/B803199;
PED3-027-550-20/00847536035706, 00847536031180, 00847536037021/B412256, B417797, B446018, B454303, B496490, B560727, B565445, B578965, B615897, B623137, B623138, B690738, B691319, B692483, B695336, B713205, B713214, B717726, B731786, B731789, B740657, B741946, B764254, B764286, B766393, B766406, B780554, B781697, B800693, B808083;
PED3-027-550-30/00847536031203, 00847536035720, 00847536037045/B397247, B408458, B414706, B416571, B473971, B475784, B479273, B494116, B608410, B611882, B615380, B663584, B666410, B695406, B696119, B697734, B709920, B713734, B713735, B729710, B738501, B738502, B746449, B783057, B783239, B783566;
PED3-027-550-40/00847536031227, 00847536035744, 00847536037069/B358931, B378536, B389916, B409027, B433336, B437033, B454059, B493075, B494673, B612534, B615604, B616547, B652056, B654234, B655595, B671064, B672455, B672476, B684986, B704973, B706035, B707059, B717259, B717261, B742095, B750469, B761584, B768849, B771892, B771893, B773963, B778320, B778326, B780176;
PED3-027-550-50/00847536031241, 00847536035768, 00847536037083/B397414, B412119, B431531, B435644, B437187, B616028, B637697, B658014, B705133, B717320, B717336, B717341, B724924, B745389, B745405, B745974, B749926, B785200, B785874;
PED3-027-600-16/00847536031289, 00847536035805, 00847536037120/B368960, B454127, B603197, B607478, B611280, B611281, B690544, B691908, B693533, B709546, B720780, B732968, B762785, B801322;
PED3-027-600-18/00847536037137/B801452;
PED3-027-600-20/00847536031302, 00847536035829, 00847536037144/B317266, B338536, B376423, B412355, B446526, B491096, B596581, B608223, B624427, B624430, B681791, B682681, B698133, B701028, B708016, B708642, B709539, B713242, B720316, B720318, B747892, B747893, B761449, B769429, B805791;
PED3-027-600-30/00847536031326, 00847536035843, 00847536037168/B339559, B412787, B457289, B467222, B489389, B594600, B597081, B611169, B611170, B626128, B652141, B653242, B654415, B681946, B682043, B701346, B711916, B712145, B719705, B726352, B726417, B728348, B735171, B748081, B748545, B748565, B758585, B808813;
PED3-027-600-40/00847536031340, 00847536035867, 00847536037182/B320318, B337325, B360226, B369513, B379020, B419419, B480964, B485981, B485982, B549767, B552409, B559740, B567778, B608855, B618581, B622005, B655601, B655798, B669603, B694091, B695569, B699270, B711618, B716111, B720752, B721600, B742928, B745966, B758046, B758050, B760363, B767594, B790633, B800873;
PED3-027-600-50/00847536031364, 00847536035881, 00847536037205/B327190, B332681, B342121, B403885, B415948, B428706, B429330, B433507, B434230, B457419, B557587, B617560, B623582, B626133, B651590, B652749, B653261, B653391, B658476, B658484, B670055, B670398, B673424, B680095, B695531, B703875, B713824, B714234, B726282, B732261, B732262, B732459, B733700, B746625, B758339, B758371, B768841, B783879, B783888, B788852, B790033, B794845
|
| FEI Number |
1000220862
|
Recalling Firm/ Manufacturer |
Micro Therapeutics, Inc. 9775 Toledo Way Irvine CA 92618-1811
|
| For Additional Information Contact | Kelli Lynch 203-500-3328 |
Manufacturer Reason for Recall | Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age |
FDA Determined Cause 2 | Device Design |
| Action | On 1/30/2025, Medtronic Neurovascular mailed "Urgent Safety and Labeling Notification" recall notices to healthcare professions who were asked to do the following:
Removal (retrieval) is isolated to unused inventory of the Pipeline Vantage 027 devices:
1) Remove and quarantine all unused impacted products in your inventory.
2) Return the impacted products to firm. Firm representative can assist with identifying suitable replacement product.
Correction of Instructions for Use applies to Pipeline Vantage 021 devices:
The key updates are:
1) Appropriate device diameter and length selection to account for complex anatomy.
2) Techniques to deploy Pipeline Vantage compared to Pipeline Shield using a balance of device tension and compression to achieve adequate wall apposition and landing around curves.
2) Warnings about the consequences of incomplete wall apposition and suboptimal deployment and the increased risk of braid deformation in females, especially in females less-than-or-equal-to 45 years of age.
Updated IFU M067713CDOC2 Rev. A (01/2025) was sent with the notice and can be found at: https://manuals.medtronic.com/manuals/
Share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Please maintain a copy of this letter for your records and the records of your patients.
Complete and return the customer confirmation form via email to neuro.quality@medtronic.com
Questions can be directed to the firm at rs.nvquality@medtronic.com or call at 1(800) 633-8766 (US Toll free) or (763) 514-4000 (Worldwide), available 24 hours a day, 7 days a week. |
| Quantity in Commerce | 18,245 |
| Distribution | Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = OUT
|
|
|
|