• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Introducer Kit Coaxial Dilator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Introducer Kit Coaxial Dilatorsee related information
Date Initiated by FirmMarch 27, 2025
Date PostedMay 07, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1735-2025
Recall Event ID 96622
Product Classification Percutaneous sheath introducer kit - Product Code OFL
ProductIntroducer Kit- Coaxial Dilator REF: KIT-002-28, 00841268104730 KIT-002-34, 00841268104792 KIT-002-35, 00841268104808 KIT-011-40, 00841268105553 KIT-011-62, 00841268105614 KIT-038-04, 00841268104556 KIT-039-03, 00841268104587 KIT-039-13, 00841268106055 KIT-081-02 / PS2-38, 00841268108394 These Micro-Introducer Kits are intended to introduce up to a .038 inch guidewire or catheter into the peripheral vascular system following a small gauge needle stick.
Code Information Catalog Number: KIT-002-28 UDI-DI code: 00841268104730 Lot Number: G25038156 KIT-002-34, UDI-DI code: 00841268104792 Lot Number: G25064259 KIT-002-35, UDI-DI code: 00841268104808 Lot Number: G25029077 KIT-011-40, UDI-DI code: 00841268105553 Lot Number: S25031227 KIT-011-62, UDI-DI code: 00841268105614 Lot Numbers: G25052163 G25080165 KIT-038-04, UDI-DI code: 00841268104556 Lot Number: G25092186 KIT-039-03, UDI-DI code: 00841268104587 Lot Numbers: G25050190 G25078192 KIT-039-13, UDI-DI code: 00841268106055 Lot Number: G25050202 KIT-081-02 / PS2-38, UDI-DI code: 00841268108394 Lot Number: G25056231
FEI Number 3000718472
Recalling Firm/
Manufacturer
Galt Medical Corporation
2220 Merritt Dr
Garland TX 75041-6137
Manufacturer Reason
for Recall
Due to a potential open seal in the sterile barrier packaging.
FDA Determined
Cause 2
Process control
ActionOn 03/27/2025, the firm emailed customers an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that affected products may have an open seal in the sterile barrier packaging that will render the product non-sterile and will increase the risk of patient harm due to microbial and environmental contamination. Customers are instructed to: 1.DO NOT USE OR DISTRIBUTE THE DEVICEs LISTED IN THIS RECALL. Identify and segregate the affected lot(s) that are in their possession as well as those in the possession of their end-users. 2. Complete the enclosed Field Correction Reply Forms and email or fax them to the attention of the Corrections Coordinator at quality@galtmedical.com or 214-778-1433. We have enclosed two forms: one to account for inventory at their facility and the other for capturing these exact details from their end users. The form lists the product number, lot number, and quantity that the firm's records indicate their facility has received. It is essential that, even if they do not have any product remaining in their possession, they complete the attached forms (Attachment A & B), indicating zero quantity to be returned, and email or fax the forms to GALT MEDICAL CORP. 3. Ship the returned product to GALT MEDICAL CORP. using Galt s carrier account information listed on the forms. 4. Reference Return Authorization Number RMA# 1200 on the outside of the shipping box and include a copy of the Field Correction Reply Forms with their shipment. 5. Once the completed Field Correction Reply Forms have been received and processed, Galt will issue a credit to customers for the returned product. Please place a new purchase order (PO), noting Replacement Parts for new products. For questions, contact Quality Director at 1-214-778-1306
Quantity in Commerce4,280 kits
DistributionWorldwide - U.S. Nationwide distribution in the states of AL, FL, GA, IL, MA, OH, PA, and UT. The countries of Canada, Ireland, Israel, Korea, Qatar, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-