| Date Initiated by Firm | April 24, 2025 |
| Date Posted | May 27, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1846-2025 |
| Recall Event ID |
96752 |
| 510(K)Number | K053110 |
| Product Classification |
Electrode, ion based, enzymatic, creatinine - Product Code CGL
|
| Product | ABBOTT POINT OF CARE TEST CREA I-STAT (25/BX) ABBTPT CARTRIDGE. |
| Code Information |
Catalog No. 03P8425; GTIN: 10054749000125; Order No. 46085650. |
| FEI Number |
3004386693
|
Recalling Firm/ Manufacturer |
Mckesson Medical-Surgical Inc. Corporate Office 9954 Maryland Drive Deep Run Iii Ste. 4000 Richmond VA 23233
|
| For Additional Information Contact | Alexia Allina 984-292-4539 |
Manufacturer Reason for Recall | transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result. |
FDA Determined Cause 2 | Storage |
| Action | An URGENT PRODUCT RECALL notification dated 4/21/25 was mailed to consignees. The notification instructs consignees to immediately examine their inventory and quarantine all product identified as subject to this recall action. Consignees are asked to return the completed response form to MMSQARecalls@McKesson.com and to destroy any quarantined devices. Questions about this recall can be directed to 1-800-688-8840 from 8:00 am to 5:00 pm Monday through Friday. If product has been further distributed, consignees are asked to forward the provided recall notice to those customers. |
| Quantity in Commerce | 25 units |
| Distribution | US Nationwide distribution in the states of AL, AR, IL, KS, MT, NC, NJ, OR, SC, TN & WV.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = CGL
|