| Date Initiated by Firm | May 02, 2025 |
| Date Posted | June 05, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1910-2025 |
| Recall Event ID |
96823 |
| Product Classification |
ChannelCheck Convenience Pack - Product Code MNL
|
| Product | ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample) |
| Code Information |
Lot Numbers: 200014
200015
200016
900153
900154
900155
900156
900157
900158
900159
900160
900161
900162
900163
900164
900165
900166
900167
900168
900169
900170
900171
900172
900173
900174
900175
900176
900177
103564
103312
103293
102807
102096
101673 |
Recalling Firm/ Manufacturer |
Healthmark Industries Co., Inc. 18600 Malyn Blvd Fraser MI 48026-3496
|
Manufacturer Reason for Recall | During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An "URGENT PRODUCT RECALL" notification letter was sent to customers beginning 5/2/25.
Recommended Mitigations/ User Recommendations
Please discontinue use of the Pre-Filled Water Syringes containing non-sterile water, which can be identified by a red cap. Any unused product can be returned for credit or replacement.
Customer are requested to take the following actions
1. Ensure all affected product in your facility s inventory is no longer in use and has been properly quarantined.
2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected products have been transferred.
3. Contact Customer Service via email at HMCS@hmark.com or 800-521-6224 to return your unused affected product or if you have any questions.
4. Complete and return the response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.
|
| Quantity in Commerce | 193035 units |
| Distribution | US Nationwide and Internationally to countries of: Canada, Malaysia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|