 Class 2 Device Recall VersaCross RF Wiresee related information
Date Initiated by FirmJune 10, 2025
Date PostedJuly 10, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2104-2025
Recall Event ID 97036
Product Classification Catheter, septostomy - Product Code DXF
ProductVersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.
Code Information UDI-DI 00685447006121 Lot 35935598
FEI Number 2124215
Recalling Firm/
Manufacturer
Boston Scientific Corporation
4100 Hamline Ave N
Saint Paul MN 55112-5700
For Additional Information ContactClara Johnson
651-582-4000
Manufacturer Reason
for Recall
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
FDA Determined
Cause 2
Packaging change control
ActionOn June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.
Quantity in Commerce14,724
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-