 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmJune 12, 2025
Date PostedJuly 21, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2154-2025
Recall Event ID 97154
Product Classification I.V. start kit - Product Code LRS
ProductMedline Convenience Kits, containing BD ChloraPrep Clear - 1mL Applicator, labeled as: IV START KIT LAB DRAW-SH, REF CDS860014P
Code Information UDI/DI 10888277797819 (EA), 40888277797810 (CASE), Lot Numbers: 24EBR549
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
Manufacturer Reason
for Recall
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms for the patient.
FDA Determined
Cause 2
Other
ActionOn June 12, 2025, firm instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No product is to be returned.
Quantity in Commerce184 units
DistributionDistributed domestically to California, Florida, Pennsylvania, and Texas. No international distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.