 Class 2 Device Recall Prismaxsee related information
Date Initiated by FirmJuly 28, 2025
Date PostedAugust 26, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2437-2025
Recall Event ID 97233
510(K)NumberK190910 
Product Classification Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
ProductPrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701
Code Information UDI 00085412639499, All Serial Numbers
FEI Number 3038195011
Recalling Firm/
Manufacturer
VANTIVE US HEALTHCARE LLC
510 Lake Cook Rd
Deerfield IL 60015-4964
For Additional Information ContactVantive's customer service
800-284-4060
Manufacturer Reason
for Recall
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
FDA Determined
Cause 2
Under Investigation by firm
ActionVantive issued an Urgent Medical Device Correction notice to it consignees on 0728/2025 via USPS first class mail. The notice explained the issues with the devices, hazard involved, and requested the following: Actions to be Taken by Customers 1. Operators may continue to safely use the PrisMax Systems and TherMax Blood Warmer Units per the associated Operator s Manuals and on-screen instructions. If further assistance is required, please contact your local Vantive account representative. Please ensure that all operators of these devices are made aware of this Urgent Medical Device Correction notification. Vantive recommends that a copy of this notification is posted within the appropriate location of your facility. 2. As corrections become available, a local Vantive representative will contact your facility to determine the correction plan and schedule the associated correction(s) for impacted devices. 3. If you received this communication directly from Vantive, complete the enclosed Vantive Customer Reply Form and return it to Vantive by scanning and e-mailing it to FieldActions@vantive.com. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive an email and/or phone call from Vantive to confirm your receipt of this notification. 4. If you purchased this product from a distributor, please note that responding using the Vantive Customer Reply Form is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your
Quantity in Commerce4772 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KDI
-
-