Class 2 Device Recall Argus PB3000

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Argus PB3000see related information
Date Initiated by FirmAugust 08, 2025
Date PostedSeptember 29, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2672-2025
Recall Event ID 97451
510(K)NumberK221056 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductArgus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006
Code Information REF: 1A.701307: UDI- 07613365003024 REF: 3.921002, UDI: 07613365001914 REF: 3.921030, UDI: 07613365003468 REF: 3.921031: UDI: 07613365003475 REF: 3.921006, UDI: 07613365000115 Affected Serial Numbers: 7010.000160 and lower, 7012.000356 and lower, 7013.000256 and lower, 7014.000152 and lower
FEI Number 3002808182
Recalling Firm/
Manufacturer
Schiller, Ag
Altgasse 68
Baar Switzerland
Manufacturer Reason
for Recall
Vital sign monitoring instrument may trigger an error message during blood pressures measurements, Specifically, when the initial inflation pressure is set in the high-pressure range.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn August 8, 2025, Schiller AG issued an Field Safety Notice to affected consignees via E-Mail. On September 6, 2025, Schiller issued a "Urgent Medical Device Recall Expansion their recall to additional products. Schiller asked consignees to take the following actions: 1. Read the Field Safety Notice (FSN) provided by the system manufacturer carefully 2. Wait until your system manufacturer contacts you for device testing or to arrange the remediation of your device(s). 3. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (As appropriate) 4. Please transfer this notice to other organizations on which this action has an impact. (As appropriate) 5. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents
Quantity in Commerce266 units
DistributionUS Nationwide distribution in the states of MT, WI, IL.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MHX
-
-