• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BioPro Femoral Head

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BioPro Femoral Headsee related information
Date Initiated by FirmSeptember 09, 2025
Date PostedOctober 14, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0148-2026
Recall Event ID 97605
510(K)NumberK090208 
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
ProductBioPro Femoral Head (Metal 22, 28, 32, 36, 40) cobalt chromium and ceramic heads Part ID/Description: 18593 HEAD FEMORAL 22MM -3 18594 HEAD FEMORAL 22MM +0 19003 HEAD FEMORAL 28MM-6 19004 HEAD FEMORAL 28MM-3 19005 HEAD FEMORAL 28MM+0 19006 HEAD FEMORAL 28MM+3 19007 HEAD FEMORAL 28MM+6 19008 HEAD FEMORAL 28MM+9 19053 HEAD FEMORAL 36MM -3 19054 HEAD FEMORAL 36MM +0 19055 HEAD FEMORAL 36MM +3 19056 HEAD FEMORAL 36MM +6 19057 HEAD FEMORAL 36MM +9 19130 HEAD FEMORAL 32MM-6 19131 HEAD FEMORAL 32MM-3 19132 HEAD FEMORAL 32MM+0 19133 HEAD FEMORAL 32MM +3 19134 HEAD FEMORAL 32MM+6 19135 HEAD FEMORAL 32MM +9
Code Information Part ID/ UDI-DI: 18593 M20918593 18594 M20918594 19003 M20919003 19004 M20919004 19005 M20919005 19006 M20919006 19007 M20919007 19008 M20919008 19053 M20919053 19054 M20919054 19055 M20919055 19056 M20919056 19057 M20919057 19130 M20919130 19131 M20919131 19132 M20919132 19133 M20919133 19134 M20919134 19135 M20919135 All lots and serial numbers affected
Recalling Firm/
Manufacturer
BioPro, Inc.
2929 Lapeer Rd
Port Huron MI 48060-2558
For Additional Information ContactMr. Patrick Pringle
810-982-7777
Manufacturer Reason
for Recall
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
FDA Determined
Cause 2
Package design/selection
ActionConsignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Quantity in Commerce147 total
DistributionUS Domestic distribution to Texas and Michigan.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LPH
-
-