| Date Initiated by Firm | September 09, 2025 |
| Date Posted | October 14, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0151-2026 |
| Recall Event ID |
97605 |
| 510(K)Number | K912641 |
| Product Classification |
Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Product Code LZO
|
| Product | BioPro Femoral Head (Ceramic heads 28-30)
Product ID/Description
19023 HEAD FEMORAL CERAMIC 28MM -3
19024 HEAD FEMORAL CERAMIC 28MM +0
19025 HEAD FEMORAL CERAMIC 32MM -3
19026 HEAD FEMORAL CERAMIC 32MM +0
19027 HEAD FEMORAL CERAMIC 32MM +3 |
| Code Information |
Product ID/UDI-DI
19023 M20919023
19024 M20919024
19025 M20919025
19026 M20919026
19027 M20919027
All lots, all serial numbers |
| FEI Number |
1832656
|
Recalling Firm/ Manufacturer |
BioPro, Inc.
2929 Lapeer Rd
Port Huron
MI
48060-2558
|
| For Additional Information Contact | Mr. Patrick Pringle 810-982-7777 |
Manufacturer Reason for Recall | Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted. |
FDA Determined Cause 2 | Package design/selection |
| Action | Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro. |
| Quantity in Commerce | 147 total |
| Distribution | US Domestic distribution to Texas and Michigan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|