• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall AW Server

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall AW Serversee related information
Date Initiated by FirmSeptember 05, 2025
Date PostedOctober 08, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0068-2026
Recall Event ID 97657
510(K)NumberK081985 
Product Classification System, image processing, radiological - Product Code LLZ
ProductAW Server 2.0, AW Server 3.0, AW Server 3.1, AW Server 3.2
Code Information UDI-DI 00840682102384 System IDs 907729AWS1 720238AW1 907ULMCCT 205934AWSRV 205975AWSRV 205971AWS 205510VAW 205815AWS2 UABMAWS1 749153GE3 UABMRPET 256593AW1 256265AWS4 256265AWS3 205481AWS1 251343AWS 251665MCIAWS 251316VAWS 251435AW5 251435AWSMARTSUB1 303584HP To be provided 251435AWSSMARTSUB1 256386AW1 205RCAWS 870423AWSMARTSUB1 870642AWS 718329SMART11 479314AWS1 479314AWS2 870207AW1 501686AWS31 501978AW1 501686AWSMARTSUB2 501552AW2 501202AWS1 501686AWSMARTSUB1 479338AWS 928773AWS 928214AWS 480256AWS 928757AWEN To be provided pacsaws01 To be provided 602344AWS 602651AWS 480342AWS 602277REVAW 480342AWSMARTSUB2 480342AWSMART1 480342AWSMART2 602344AWS1 602344SMARTAW 480342AWS2 602BUMC870 480342AWSMARTSUB3 928445AW1 06A6A668-AW-AWS_PRIMARY 844823AW1 844823AW 928283AWSMARTSUB1 520STJAWS 818RIAWSVR 661CHAWS 661KMCAWS 13123 13767 650696AWS To be provided 530UCVETAWSERV To be provided 760633HDCT To be provided 714FVAWSERVER 714FVDISC 510792AWH 559326AWS1 714447AWS AVHHFCTAW1 858657AWSMARTSUB2 858626AWS32 858824AWSMARTSUB2 858824SMARTSUB1 To be provided 909558SMART 805737AW2 562826AW46 9694 9693 323361AWSMARTSUB1 13256B To be provided 13526B 559353AW1 949367AWS 323BEVAWS 951704AWS 949764AWS 949764SMARTAW 510534AWS 510869AWSERV 510752AWS 916SUTABAWS 510437AWDL360 5496962AW11 510869APEX 760940AWS32 760940APEX 714771AWSRV 714456AWSMART 714456AW 650725AWS10 650725AWS30 650725AWSVM1 650725AWSVM2 650725AWSVM3 808341AW2 760446AWS 949278AW2 916SUTAWS 916SUTAWS2 916UCDAWS To be provided 858966SERVER 619532SERVER 9907 858657AWS32 858939EHLAWS 858626EHLAWS 858805AWSVM01 415600AWSERVER 415833AWSF 415476AWV 415MBCREV1 UCSFAWS4 UCSFAWS3 650725AWS20 408GSHOGEAW 408259EDCT 510784AWSERVER 9200 13195B 13196B 650696PET 805AWVM 310264AWS SHCGEAWSVM2 To be provided 818881AWSVR 805434AWS AWSHT1 310222HPAW 310823EHL1 310423EHL2 310423EHL1 To be provided 818782AWS1 To be provided 818676AW4 310423AWS 310423AWS2 To be provided 970544AWS 720848AWSRV 719365AWS DHAWSERV1 MCAAWSERV 303602AWS1 303DHREVCT 970945AWL03 970945VAW 307FRMIPCT4 719955REVO1 303603AWSRV 720321AWS1 303730AW 719584VPSW 203384AWSERVER 203739AWS 860679AWS 860679AWSSCT To be provided 860JRAWS 860972AWSMARTSUB2 203694AWSMARTSUB1 203694RTCT16 203694CTREV 860344CTFR1AW 860358AW1 To be provided 860224AWSVR1 860224EDAWSSCT 860224APEXAWSSCT 203737AWSVR 203276AWS1 203380AW1 860496AW02 203261AWS 860456APEXAWSSCT 202715AWS 202476AWS 202877AWS32 202745CT1 To be provided 302744AW 302744AWSMART 302645AWS1 302733AWSERVERL02 302733AWSMARTSUB2 To be provided 302645AWS DUPONTPETMRAW 732996AW1 727398AWS2 727398AWS 561955AWS1 561955SMARTSUBAWS1 561955AW2 954450AWS 561955SMARTSUBEVO1 561736AWS To be provided BRANDONAWS To be provided 352799AWLOW 352597AWS2 813635MPAWS 321799CTT03 321799AWS 314868AWS1 386HAWSERVER2 AM1436W320_DUP To be provided RADIOLOGY20072529AWS 772LMCAWS 850863AW1 352376AW 333SERV MEMAWSERVER MEMAWS 786243AWS1 BAYONETPTAWS To be provided 904202VAWS 904953AWS2 904953AWS 904202AWS 904953AWS3 904244AWS1 904ASHCHIVAWS 904996VAWS AM13D0X149_DUP 561747AW49 OSCAWSERVER 407PCSERVER 863577MRAW WATSONAWS2 321HFSERV To be provided 305575AWS 305596AWS 786596AWS 786596AWS1 305223AWS 305575REVO 786596WKSMART To be provided To be provided To be provided To be provided To be provided HCIT5218773AWS 305535AWSR NCHAWSERV NCHAWSERV2 239NHAWSERV USONCAWSRV 352816AW 850383AWE2 850383AWE1 OVAWSERVER 561346AW 386TLSERVER 954844AWSMARTSUB1 850494CTAWSER 941883AWSERV 772335AWV AM141C095 To be provided 904819AWS 407321AWL02 FAWCETTAWS LARKINAWS32 305284EVO To be provided 850431AWSERVER 813635AWSRV MACDILLAWS BAYCAREAWS1 BAYCAREAWS2 MOFFITTAWS2 954767AWE2 954767AWE1 AM15P4F42_DUP HLEEMOFF20867329 561969AWS1 404616AWLTSERVER 770903AW3 404686AWSMART 404686REV80AWS 404616AWSMART2 404616REV_ES_AWEHL 404616AWS 404616AWSMARTSUB2 404616CASSAPEX2EHLAWS 404616CASSAPEX To be provided 404686APEX_EHL_AWS 404686REV_80_AWS 706774AWSER 912466OAW 912466AWS 770812HP 706544MAAWS 706787AWSERV 706787PCT To be provided 770219AW1 770535REV 404712AWSERVER2 478633AWSMARTSUB1 478633UWAASTDV2 478633UWAASTDV 912350CTREV2AW 229228AWS43 229228AWCL4 229353AWS To be provided 808YMAWHT 808522AWHT 808432AWHT2 808432AWSMARTSUB1 808522AWS1 808522AWHT1 808233AWSMARTSUB1 808KMAPCT 808432REV 808433AWHT 515282CTAW 319338AWSXL 641428AWS3 712737AWSMARTSUB2 319272AWSERVA 208422REVAW1 208954AWS 208205AW1 630264AWS1 847381AW 630268AWSRV 1000005354AWS 1000005358 630266EHLAW 312864PAWSW 773296AWSRV 773292AWSRV 312942AWSRV 773665AWS 773702AWS 312454SMART 312996CTRV2EHLAWS 312996AWS1 312864AWS 773962AWS 312996CTASCAWS AM13DKD98 773665D710 815756AW To be provided 630275AWSRV 847437AWS 847316AWS 1000003355EHLAWS 708202AW1 70824SMARTAWS 815741AWS 630364AWS1 630268AWSVIR 618998AWS1 708216AWSRV 708216NM870 To be provided 618244AWS1 630527AWS 815300AWSVR 309454AWS1 708425AWSRV 708425AW32 708425CTRV2EHLAWS 708383AWS AM15A0C2292_DUP 708747DIS610 1000005331 708274AWS 847723AWSRV 847723AWS 309691PSERV 1000004500 1000000098 1000003733 561589AW 309683AWS1 309655AWS 309683AWS3 1000005353EHL 1000000010AWS 309672REV1 309691HD750 1000005502 To be provided To be provided 1000003732_AWS 217223AWS1 815226AWSR 815971AWS 779696AW1 1000004006AWS 847305AWS 217528AWS 217383AWS1 217337AWS1 To be provided 630824AWSMARTSUB1 765298AWS1 765298AWS2 260665AWSERV 812934AW2 574546MAXCTNANO MSKAWCUSTOM50 812485SERVAW 812450SERVAW 260266AWSERV 574364AWS 317887AWSERVER 317621AWS1 574204AWS 765747AWS1 812471PETCT To be provided 574647AWS 574647REVCT1SSAW 574647AWSUB 574267AWS1 913588AWS 913588AWS2 913588NREVAWS 913588APEX1AWS To be provided 913588APEX2AWS 4514429AW11_DUP 913676SAWSERVER 913588WWAWS 316262AWS To be provided 270745AWSRV 502875AW 859233AWS 859276AWL02 859313AWL03 502897IGS740AW1 502287PETAW 270767AWSMARTSUB2 270752AWSMARTSUB4 270767CTREV 270444AWS1 270575AWS1 859879DIAG 859737AW3 225755XR656 985NOAWSERV To be provided 504842AWSERV1 504842ERCT 318395AWS1 504349AWS 504347AWTIER1 504988AWS1 318410AW4 318214AWSMARTSUB1 318214SMARTSUB1 504529CTAW 504TOURMRAW 504988AW2 337948AW 985446HPAW 985493PET1 0850210870AS1 617726AWSCTMR 6177263DAWS 617414SERV1 617754AWSMARTSUB3 617754AWSMARTSUB2 857880AW 617754SMARTSUB2 To be provided To be provided To be provided To be provided 617665SERV1 617665AWSMARTSUB1 617665AWVEN1 516252AWSMART 617381AWSMART1 STEWARDAWSRVR 508457AWS1 508971AWS1 978466AWSVR1 781306AWSMARTSUB1 781213UWAASTDV 781453AWSMARTSUB1 41344716K 978741AWS1 978741AWSMARTSUB1 To be provided 978741VIRAW1 781624AW 781624AWSMART1 781624AWSMARTSUB1 518243AW45 781624AWSMARTSUB2 413748AW 413794AW2 413748AASW 413568AWSMART 413794AW1 413748PVSW-AWSVMTEST 508334AWTIER 410333AWS To be provided 443643CTAW 443643SMARTSUB1 443643AWSMARTSUB2 301496AWS1 410822AWS 240964AW 412607AWS1 412607AW1 240637REVEVOAW 808341AW1 410337AWSMARTSUB1 207621AWL03 207563AWS1 207795AWS 207795HPD1 207795AWSMARTSUB1 207255AW 207662AWHTS 207885REVCT5AWS 207662REVCT4AWS 207662REVCT3AWS 734936AWSRVM 734936AWSRVR3 UMREVCT2 To be provided 734844AWS1 517353AWS 810732AWS 810606ASW1 616391AWS 616AWS 616685D870CZT 616392AWS 616546AWS1 906486AW 269341AW1 269341AWS1 269341AWS2 517364AWS To be provided 269789AWS 248465AWS 248849AWS2 248898AWS 989583VAWS 989907AWL To be provided 248849AWS 231213AWS 231935AWSCT 586576AWS 586573VMS 734763AWS1 763792AWS 507385AWS1 507794EHLAWS 612725PETAW 507284AWS2 507284AWS1 NM1069915 WS278814 WS529288 314344HORAWSRVR To be provided 816698SMARTSUB2 816691AWSERVER 816932AWS 816276PRIAWS 816276AWS 816282NM870 573686ASVR 816271AWSMARTSUB2 816271AWSMART1 816271AWSMART2 816271AW 314251MMR5AW1 3148652AWSMARTSUB2 314251AWS4 314865RECT 417269AWSMART1 417269AWSMARTSUB2 417820AWSMARTSUB1 901227AWS1 901227AWS2 228GPAWSRV 228575CTAWVM 601288AW1 601264EVOAWS 601264AWS 601288AWV3X To be provided 601968AWS1 601982SERV 601982AWS1 333332AW1 MSKAWCUSTOM35 MSKAWCUSTOM34 662841AWSVR 662337AWS1 406731AW1 406455AWS1 406455CTREV 406329AWSRVR 336884AWS1 828213AWSVR 828257AW1 910721CTMAX 828262AWS 910450AW 704783AWS 704597AW1 To be provided To be provided 828894SMARTSUB1 828894AWSMARTSUB1 704403CT1SSAW 704801APEXAWS 910892GLDSLAW 919684AWSERH 919684AWSERL To be provided 704834AWS To be provided 336832APEXAWS 252744AWSERV 828696AWS 828681PRAWS 828681AWSMARTSUB2 828681CTSMARTSUB To be provided 336884APEXAW 704612AWSMARTSUB1 336846REVOAW 704403AWSMARTSUB2 704403SMARTSUB1 336564CT1 828757REV1 980993AWSMARTSUB1 980993SMARTSUB1 704355AWS 336719REV1 336719CT3 CEAWSERVER 910715AWS1 252908AW 704512APEX1AWSMART 919938AW1 910642AW1 910667AWSMARTSUB1 910667AW2 910772AW2 336713AWS1 336716AW 336397CT11 336765AW 605328AWS AM137HD87 701857AWS 308537AWS2 308865PETAWS2 308455REVSMARTNANO 308865PETMI 402481AWS1 402481AWS2 402484AWS1 To be provided To be provided To be provided 402559AWS1 603650AWS1 603629AWS 603668AW 603668AWSMARTSUB1 603629CT660 603526AWSMARTSUB1 603356AWS1 703689AWS To be provided 732840AWS 856886APEXAWSMART 609463AW532 856488AWSMARTSUB1 609922AWSMARTSUB1 973676AWS 732321AWS 732294LAWS 609652AWS 609914AWS1 609584AWS 732739AWS 201635MSKCCVM To be provided 732775AWS 973971AWS2 973971AW2 973971AW11 973971ASCEND To be provided To be provided 732776AWS1 To be provided 973972AWS 973926AWS 973877AWS 973972AWV 4211061AW41 609380AWS AM14EGX46_DUP 732360AWS 973754AWS 609537AW1 973831AWSR 201634AW 201447AWSERV2 908522AWS1 732557CTAW 973754AWSMARTSUB1 973534AWSSMART 505609AWS SJRM20570373 575556AW1 575627AWS 505368AWS 702616AW 702961AWS 702724AWS 702724PET 518262AWSMARTSUB2 518373AWS 518262AW1 518262AWSMART1 518262SMARTSUB4 518626AWS1 51262AWSMARTSUB4 518262AWSMART3 518842AWS 718918AWS 914349AWS13 BLH20533405 BLH20081761 718918AWSMART 718920SMARTSU1 718904AWSMART1 718904AWVM 914349AWSMARTSUB1 718519SMARTAWS 718960AWS To be provided To be provided To be provided 718920AW2 914787AWS 914787EVO To be provided To be provided KCAWSERVER 718567AWS 718250AWSVIRT 718780APEXSMARTAW 718238DISCIQ 718240EDCT 718836AWS To be provided 718250AWS 716845AWMR42 716845REVOCT 516877AW11 518292AW 585755AWS3 631638AWSMART 631CMKTEMPO 914737AWS1 718334AWSERVER ELMVIRTAWS 607776PRISTINA4 914457AWSMARTSUB1 To be provided 516674AWSMARTSUB1 518773AWS 631670AWS 516563AWS 570529AW1 718206AWS 718206AWSVIRT 607WMCAWS1 845334AWS 516622AWS To be provided 646962AWSMARTSUB2 646962AWSMARTSUB1 516622SMARTSUBAW1 631548AWS 631291APEXSS 607535REVCT 914666AW3 914666AWSMARTSUB1 914666REV 718920AWSMARTSUB1 718920SMARTSUB1 EHAW2SERVER LIJ066505AWS 516NYCBAWSMARTSUB2 LIJ071291AWS NSLI21213406 NSLI21213404 NSLI21213403 NSLI21213402 NSLI21213401 NSLI21213400 NSLI21213399 212241AWSERVER 212562AWSERVER 212369AWSERVER MSKAWSDL580 NYPHLTSERVER MSKAWCUSTOM30 MSKAWCUSTOM29 MSKAWCUSTOM28 MSKAWCUSTOM27 MSKAWCUSTOM26 MSKAWCUSTOM22 CORAWRSH HSSAWSERVER NYPHAWSVR05 NYPHAWSVR04 NYPHAWSERVER2 NYPHAWSERVER 212305MH3SMART 212639AWS1 BELVIRTAWS 568185AWSMARTSUB1 NYPHJOSUB1AW NYPHDHKSUB1AW 212305MH3SMARTAW CHONY3SMARTAW 212312AWSMART HSSM3CT750A MSK68IR3ANGIO To be provided MSKAWSCUSTOM31 CHONY3CTAWSMARTSUB1 paviradgeaw8 paviradgeaw7 paviradgeaw6 paviradgeaw5 paviradgeaw4 paviradgeaw3 paviradgeaw2 paviradgeaw1 AM126TR857O 212241AWVCT MSKAWCUSTOM33 516632AWS2 516632SMART1 To be provided 516719AWSMART 631751AWS 518926AW 631PBAWS 516734AWSMARTSUB1 To be provided 585275AWS 585922AWS 585275AWS1 To be provided 516705AWSMARTSUB1 518897AW 212305AWS 518243AWSMART 518243AW46 516266SMARTSUBAW1 631726AWS To be provided 718226VZAWS 718226AWSMART1 718226AWSMARTVZ 631444AWS 845368AWS30 315464AWS2 802342AWSMART1 802342AWSMART 315464AWSMARTSUB1 212305AWSMARTSUB2 315473AWS 607865REVAW 914682AWS MSKAWCUSTOM54 MSKAWCUSTOM53 MSKAWCUSTOM52 MSKAWCUSTOM51 MSKAWCUSTOM39 MSKAWCUSTOM38 516705AWSMART1 914681WPHAWS 914681AWS 212608AW 914681AWSVM 914681AWS1 212608CT 716688AWS 716597REV1 914666AWSMARTREV1 330792AWS 216293AWS2 216293AWSMART 330729SMART2AWS 330729AWS1 740439AWS1 513584AWS1 859371MFAWS 513584AWS 513569AW11 513636AWS To be provided To be provided 216791AWSERV 614293AWS 614293DAWS 419526AWSERV 614DHWAWS 614RMHSSAW2 614RMHSSAW1 614GMCER2AWS 614HOBRT2 614CANASAWS 937208AWS1 740GDYEVOAWS DMHAPEX1AWS To be provided DMHEVO2SSAW 513290AW1 513803AWS2 419227AWSMARTSUB1 419526AWSMARTSUB2 419526SMART3AWS 614RMHAWS 740383REVOAWS 740633AWS To be provided To be provided 740348UWAASTDV 419394AWS 419291AWSMARTSUB2 419291AWSMARTSUB1 419578AWSMARTSUB1 330841AWS1 To be provided 614WMCERAWS 330480AWSMART1 330480AWSMARTSUB3 330481AWSMARTSUB1 330480AWSSMARTSUB2 580559APEX160 580559APEX160AWS 918577AWS1 405773AWS1 918758AWSMARTSUB1 918838AWSERVER 918494SERVER 918233AWSERVER 918640AWSMARTSUB1 918505IGSAW 541789AWS2 503418PETMRAW 503216AW1 541MCAWSVR 610402AWSERVER2 484526AW 724773AWS1 484526APEXAWS 484526ASCENDAWS To be provided 717262LTAWS 570645AWS 272212EHLAWS 610954AWSERVER 484503REVAW1 484822AWSMARTSUB To be provided 814877AWHT 814838PVSW1 814868AWSMARTSUB1 724450AWS 724357AW 635217GE3 866785AWS 215707AWSERVER 215537SERVER 215CHOPAW2 215503AWSERVER 215728AWS1 215728FCCCREV 412623AWSRV3 412942AWSERV 412802AWS3 412802AWS2 412802AWS4 412942AWHT 412802AWS6 412802AWS1 412942AWS1 To be provided To be provided To be provided To be provided To be provided To be provided To be provided 412802AWS 610327AWSERVER 267985NANO 215453AWLT 814676AWSERVER 814231AW3 610869AWSERVER 570552AWS 570842AWSERVER 570824CT To be provided 412802AWS5 717855HTAWS AM15RJE119_DUP 401444HPD1 401444AWSERV 401782AWLT 401348CTAWS1 864512APEX1AW 843789DISCAW 843789DISC 0004375151 To be provided 843347AW1 864255AWS 864725AW2 843692AWSERVER 910269AWS12 864560AWS1 843209AWS1 803774AWS 803936AWS 803791AW3 605322AWVS 605328AWH 605333SMARTAWS 605333AWSMARTSUB2 To be provided 423844SERVER 423708PET 931528CT 615435AWSBJ 731541AWS1 731541CTAWS 731541AWS4 423433FCTAWSEHL 865584AWS 865588CT1 865977AWSERV 916EXPOAWS 901522AWS 901595AWS 901523AWS 901516AW346 901287AWS1 AM13DLV175 TBD 615349AWSERV 615327AWS 615284AWSERV 615327AWS2 615AWSERV 706291SERV 615322AWS3 615463AWSERV 615222REV1 To be provided To be provided 325670AWS 325670AWS3 817465AWSR 512324AW 512TRAVISAWS 512UTAWS 979MGAWS 832325AWS1 979SJAWS 979KSAWS 936539PRAWS 361RISTAWS 361761AWS2 281477AWS1 281477AWS2 214857VAAWS1 214645PTMIAW 214345AW1 469HPAWS 214645AW1 214820870NM 915544AWSMARTSUB1 915532HPTIER 210BAMCAWS32 956389AW 713792AWS 713442AWS2 713563AWS 713HCAHAWS 713566AWL02 713790AW2 MHTMCAWS2 MHTMCAWS 713790AW1 713791AW 281929AWS 713790AWS 713442AWS7 713442AWS8 713ACT48SAWS 713ACT49SAWS 713486AWCAW 713441AWSMARTSUB1 832522AWS1 713790AWS2 713791DISC AM12X6P114_DUP To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided To be provided 414559AWSMART1 AM13YTC307 281319AWS1 To be provided 281737AWSMARTSUB1 432221AW 713790AWS1 713790APEX1_NANO 903243AW1 463814AW2 601310AWSMARTSUB2 USONCAWS2 210450AWS 210351AWS 210575AW41A1 210617REVOAW 210MHSWSMARTSUB2 210MHSSMARTSUB1 210MHSSMARTSUB2 210575AW1 713442AWS5 713442AWS3 254965AWS1 713442AWS11 210219AWSMART 254SWAW2 936270AWSMARTSUB1 936270AWS 830278AW 361572SSASW 361572SSASW-AWN 361573UWAAVMSW 254TSICT 713442AWS6 801855AMX4 435716AWSVM 801314AWS1 801507REV 801507CTSMARTAW 801479AW6 AM13J3U3 435658AWS1 435658REV1 435637SERVER CUMC15111209 801357AWS1 801357SMART1AW 801718AWS1 435251SMART1 435251NM 801581D710AWS 801662AWS 801662AWS1 385531CT2SMAWS 801268AWS1 To be provided 801501AWS1 703504AWS1 703664AWREV1 703849AW41 703558AWSRT 434SMJHS1 804281AWS 703698AWS 703391AWS 703858AWS 703698AWS2 703391AWS1 703698SMARTSUBAW1 FH120609295AW2 FH120609295AW1 703776AMX3 703776AWS1 434315AWS 571231AWS1 757722AWSERVER To be provided 703369AWSL 276666AWSER 804594AWS3 757388SERVER1 To be provided 804SRMCAWS 703648AWSL 703648AWSH 571367AWSH 804VCUPETAWS 804281AWS3 501387AW 5156567PT11 804289CT660 AM14B89292_DUP AM16KDW1458_DUP To be provided 540776AWS2 703776AWS 757455AW1 540536AWSL 540536AWS 703523AWS 802238CT1 802775CTAW 802775AWSMARTSUB1 To be provided 802775UWAASTDV 425688SMARTSUBAW To be provided To be provided To be provided To be provided To be provided To be provided 425261AWS 425899OPTSMARTAW 206545EESW 206215AWS 206598AWSVR 206987AWS 206598AWSVR2 206668AWSMARTSUB1 206277AWS2 425258AWSMARTSUB 206598AWSMARTSUB3 206598AWSMARTSUB4 206598AWSMARTSUB2 206598VAWS 206598AWSMARTSUB1 206368AWSMARTSUB2 888825AW1 253968AWS 253968AWS1 MAMC2NM870 To be provided 360SWSERVER 920454AWS1 920738AWSMARTSUB1 920361AWS1 608930ASCENDCT 414326AWSERV2 608775AWSRVR1 To be provided 801581MIAWS 608263AWSERV 262415CT520 414326CSMAWS 414266AWSERV2 414805AWSERV To be provided 920729AWSMARTSUB1 111111111 262569AWSMART 262656AWSERV 262656AWSERV1 681342PCTAWS1 304526SMAWSERV INVIVOPET6 304264AWS 304293MR3TAW To be provided To be provided 307444AWSMART 307777AWLOW AL4093AW01 AL4004AW05 276954478 DZ1009AW05 DZ1004AW09 DZ3053AW03 DZ1863AW03 DZ2684AW02 DZ1335AW01 DZ1038AW01 DZ1045AW07 DZ1045AW06 DZ2673AW05 DZ2673AW04 DZ1062AW03 DZ4291AW02 DZ5026AW01 DZ1015AW03 DZ4258AW02 DZ1070AW03 DZ4016AW01 DZ1193AW06 DZ4609AW01 DZ1090AW06 DZ3723AW01 DZ3724AW01 DZ4149AW01 DZ1061AW07 DZ3972AW02 DZ3445AW02 DZ3635AW01 DZ2092AW02 DZ1128AW05 DZ1856AW06 DZ1027AW06 DZ4619AW03 DZ2416AW01 DZ3053AW02 DZ2773AW02 DZ4227AW01 DZ3805AW01 DZ1908AW05 DZ3343AW02 DZ3953AW01 DZ5076AW01 DZ5054AW01 DZ5038AW02 DZ1255AW03 DZ1235AW02 DZ3610AW02 DZ1638AW09 DZ1147AW02 HCIT4308951AWS MR7949ARAWS2 MR7949ARAWS1 HCIT4706078AWS HCIT4731149AWS HCIT4986733AWS HCIT1536851AWS HCIT4919173AWS HCIT4936778AWS HCIT4825134AWS HCIT4890853AWS HCIT4430774AWS HCIT4430774AWS3 HCIT4430774AWS2 HCIT4430774AWS1 HCIT4532706AWS3 83153230682 832AW000002CT1 0910492056 AUAWX23204492 83199163254 834AWS000176CT1 83179605857 83151839074 83093285391 83098956691 AUAWX24490813 0910492069 83162992905 83162511620 83163140157 83095978437 832AW000002MI1 AUAWX24438408 AUAWX25000297 10385790 832AW000162CT2 83200117961 83170652378 83154708501 83174466616 83179605493 832CT000044AWVS AUAWX25299969 832AW000163CT3 83188558879 0910492057 832AW000230CT1 83155768301 83160724893 83158812739 0910492050 83180275812 83178031902 832AW000104CT1 832AW000161CT1 83174658399 83163398084 834AW000034CT1 834AW000067CT1 83155582566 834AW000046CT1 834AW000055MI1 834AW000048CT1 834AW000049CT1 83141966577 834CT000048 83183490835 83200761748 83214197335 11975249 83181869974 83119339558 834AW000106MI1 835AWS000184CT1 83193919822 0910555038 83188375786 837AW000003MI1 837MI000003 833AW000075CT1 833AWS000211CT1 833AWS000211CT2 833AW000124CT1 833CT000136 833AW000156CT1 83208456156 832AW000131CT1 83132679954 5423120C 5423116B 5423103B 83182012655 83214033402 833AW000042CT1 833AW000043CT1 83131723729 83151034893 833AWS000181CT1 83164804848 833AW000155CT2 836AWS000202CT1 83163553521 83118756718 836AW000070CT1 836CT000213 83202635598 836AW000110CT1 5358793C 83155964206 AH5047AW01 AH4001AW07 AH4001AW13 AH7084AW04 AH7084SV02 horadspacsaw1 horadspacsaw2 AH1410AW26 AH1410AW25 AP0364AW01 AH5372AW04 AH1923AW01 AH0132AW04 12128169 12162554 AH5644AW01 AH0127AW01 AH5703AW03 AP7676AW02 AH0129AW02 AH5362AW03 AW51339232 AW52991582 AW52325242 HCIT5307016AWS BY4200AW01 BY4196AW01 BY4005AW05 BY4203AW01 BY4120AW01 BY4194AW01 BY4033AW15 BY4033AW16 BY4006AW06 BY4002AW04 BY4059AW01 BY4197AW01 AW52744012 BY4063AW01 BY4036AW03 BY4195AW01 BY4075AW01 BY4198AW01 BY4061AW01 140024OT05 130083OT10 613169AW01 190194OT02 110002OT06 190052OT14 080026OT06 030355OT10 190059OT06 AW220078SS02 AW220078SS01 110020ST12 110020ST13 110020OT02 220078OT08 AW220078S3 220078SS03NANO 220078SS02NANO 220078SS01NANO 030020AW01 030130OT01 060011OT01 030151OT03 068A6B51 030232OT01 190207SS02 190207OT04 0685cccd 06855ad5 0685429c 068536f7 06859e90 110025OT02 080038OT03 030388OT01 110022OT06 020056OT03 110022OT10 210001OT01 190226OT04 270617OT02 019022OT01 019022OT05 080004OT03 050045OT01 060055OT02 110009OT03 150019OT02 220826OT07 210002OT04 527944OT01 190747OT08 210013OT06 220096SS01 140029OT01 030091AW01 12628398 110901SS02 180041OT01 300013OT03 030155OT01 110094SS01AWS 110094SS06 110013SS02 110094OT02 110094SS04 110094OT04 110059OT03 13137868 080100OT01 110014OT08 080041OT03 110017OT05 150043OT04 150043OT06 030078OT10 030028OT03 190171OT03 110019OT11 110012OT05 AWS55110162 441236AWS HCIT4691924AWS HCIT4691921AWS HCIT5206230AWS HCIT4759526AWS BA4080SV01 BA4177AW01 BA4080AWS1 1018QAC014 BA4008AW02 BA4043AW02 BA4017AW02 AW55342162 AW53585393 AW54590032 BA4188AW01 AW5135988 BA4147AW03 HCIT4214173AWS CBDMG2300086HM PET14043PT4AW MRRD0927 HCIT4179120AWS MRUA2387AWS 85000 HCIT4185270AWS HCIT4450231AWS MRHM1985 CBDMG2300088HM HCIT4214173AWS2 HCIT7125379AWSVM HCIT4664266AWS PET13394PT2AW HCIT5296719AWS HCIT4101603AWS CBDWG2300012HM HCIT4517785AWS HCIT4751399AWS HCIT1511657AWS CT443038CN5 HCIT4044016PAWS PET13330PT6 HCIT5024502AMS 010120OT12 010120ST62 HCIT5531518AWS HCIT5051390AWS TAIXX2200152CN HCIT1612214AWS TAIXX2200012CN HCIT4345622AWS HCIT4685236AWS 0036 HCIT1399777AWS HCIT1386516AWS HCIT4414933AWS HCIT4410860AWS HCIT7120151AWS HCIT4807791AWS HCIT4670363AWS HCIT4713917AWS CT425487CN6AW HCIT4374861AWS HCIT1401982AWS USE230A7X4AW 2613126AWS HCIT4694426AWS PT4152200024PT HCIT4321260AWS HCIT4285294AWS HCIT4777744AWS HCIT4719083AWS MRR11187AWS HCIT4568867AWS HCIT4194744AWS HCIT0073507AWS HCIT4033533AWS HCIT4361797AWS HCIT4013040AWS HCIT1386665AWS HCIT7103885AWS HCIT4757007AWS CBDMG2100136HM HCIT5227412AWS HCIT5467696AWS HCIT5393888AWS HCIT5518780AWS HCIT4952394AWS HCIT4096486AWS 2584646 4651769MR11 BG4600AWS1 13479446 13377481 BG4609AWS1 BG4033AWS1 BG4037AWS1 BG4786AWS1 BG4009AWS1 BG4927AWS1 BG4020AWS1 BG4693AWS1 BG4549AWS1 12754289 12817704 13065291 AW51201171 AW5124155 AW51201621 AW51201322 6073AW0044MR1 6073AW0053MR2 6073AW0053MR1 604557APEXRTEHLAW1 613256SMARTSUBAWS 902667AWS 613623SMARTSUBAWS 204726AWS 514262AWS 604678AWS 403777AWN1 403944AWS1 403777D870A To be provided 780639AWS 613938AWS 250426D870 902473AWS2 204638AWS 780489AWS 306637AWS1 780998AWS 506452AWS02 709651NM870 To be provided 519822AWS 902473AWS1 905527AWS 905527AWS1 905522AWS2 905522EHLAWS 780532AWS 250318AW1 250979AWS1 250862AWS 250712EHL1AWS1 250979DMI1 613258VMAWS 902678AWS 613548AWSHT1 418665AWS 604514APEX1EHLAW1 450662AWS 519322AWS1 519322AWS 519663AWS 519646AWS 905883XELSV 506857AW78 To be provided 514376AWS 514934BAWS 514934AWS 514340D870C 514252D870ASCT 514866AWS1 613354AWLT 705647AWS 905895AWS 604988AWS E416249AWS1 905845REVCTEHLAWS 905845AWS 519941AWS1 705327AWS 905433AWS 613761AWS 705746AWS 403627SMSUBAW 306765AWS1 250565AWS To be provided 403343APEX1AW1 To be provided 306766AWS3 306766RCT1 418862AWS2 418862AWS To be provided 819324AWS 519464AWS 306966AWS1 306931AWS 416284AWS3 204785AWS 807737AWS 250565AWN2 4726798AW21 250565AWN1 250565AWN3 450746AWS 450746D870 905684AWS 506648AWS1 506648AWS 709737AWS 204326AWS 519272AWS 519245AWS 705522AWS 705674NM870 604582AWS 604588AWN1 604582APEX1EHLAW1 604930APEXRT1AW1 604588LSVR 306778AWS 902567AWS 450654AWS 450581AWS 416813AWS 416530VAWS 416581AWS 416581AWSMARTSUB2 416530VAWS2 416813AWS1 604875AWS 604874AWS 604885AWS 604682AWSERVER 604877AWS 604875AWS1 604739AWS 604877AWS2 To be provided 250519AWS1 250370AWS 780842AWSMARTSUB 519881AWS 514934CAWS 519253AWSV 519253AWS3 204787AWS1 902742AWS 306786AWS1 AM149YK203_DUP HCIT4942651AWS HCIT4293925AWS3 HCIT1756569AWS HCIT5393605AWS HCIT5314063AWS HCIT5424884AWS HCIT1716036AWS HCIT1702537AWS HCIT5377947AWS HCIT1649785AWS HCIT1686143AWS HCIT1659298AWS HCIT1730670AWS HCIT4096494AWS HCIT4920063AWS fc9a1db5-AW-AWS HCIT4505278AWS HCIT4567313AWS HCIT5026141AWS HCIT5183890AWS HCIT5340690AWS 082421140450AWS 082499180101 082421110138AWS 082421111102AWS 082421111172AWS 082421111326AW1 082421040294AWS 082421040629AW1 082421040678AW1 082421040532 5054637 082421040638AW1 AWS302 082421220272AWP 082426150055X1 082421150313AW1 082421150390 082421150352AWP 082421150289AWP 082421160796AWP 0684D7F7-AW-AWS_PRIMARY 082421160429AWS 082421160671AWS 082421161064 082421161100AWP 082421160671 082421160880AWP 082403160170 082421160865AW1 082421160779AWP 082421160684AW1 082421180123AWS 082421180270W11 082421180277M1 082421180343AWP 082426180026 082421240324AWP 082421240353AW1 255B00E6-AW-AWS_PRIMARY 082421070604AWP 082421070694AW1 082421010527 082421010574AW1 082421070664AWP 082421091065AWP 082421090975AWP 082421210389AWS 082421210600AWS 082421210692AW1 082421210651AW1 082421210665AWP 082421210575AWP 082421190387AWS 082421190593W11 06a2e7bb-AW-AWS_Primary 082445120013AWS 082499130821 25BDB08B-AW-AWS_PRIMARY 082421120675AWS 082445120041 082421121073 082427120342 082421121239AWP 082421200448 082421200357AW1 082421200379AW1 082421200528M1 082421020180AWS 082421020385AW1 082421020226AWS 5107269 LANLING 082421230771 082421230698AW1 082421060121AWS 082421030661 082421030687AW1 082421190258 008500000 06A8C80F-AW-AWS_PRIMARY 4839807 082421300094 082421300117AW1 082421120788AWS 082421111184 082421111124AWS 082421111108AWS 082421111172 082499110172 082421111185 082445110005 082421250311W11 082421040292AWS 082421040302AWS 082421040329AWS 082421040340AWS 082445040019AWS 082421040300AWS 082421040281AWS 082421070509AWS 082421070568AW1 082455200034 082421160581AWS 082455040360 082421040608AWS 082421040733AWS 082455040362 082455040363 082421040647AW1 082421190317AW1 082421040334AWS 082421040335AW1 082421160425AWS 082421040407AWS 082421040378AWS 082421300086AWS 082421040505AWS 082421040499AWS 082421150328AWS 069B9916-AW-AWS_PRIMARY awsfuxing 082421040492AWS 082421111187AWS 082421111187 082421120446AWS 082421120449AWS 082421120672AWS 082421190446AWS 082421160539AWS 082421160564AWS 082421280071AWS 082499180123 082499180122 082421230497AWS 082427230116AS1 082421280086AWS 082421060195AWS 082421060231AWS 082421220322AWP 082421220338AW1 082421150184AWS 082421150216AWS 082427150116AWS 082421160145AWS 082421160545AWS 082455160205 082499180086 082421160679 082421160605AWS 082421180137AWS 082421180135AWS 082421180352AWS 082421160334AWS 082499150188 082455160256 082421160335AWS 082421100733AWS 082499170033 082427170039AWS 082421170111AWS 082421140296AWS 082421140324AWS 082421140834AWS 4778487 082421140427AW2 082421010395AWS 082421010410AWS 082421111146AWS 082421111201AWS 082421111201 082421090981M0 082421090698AWS 082421101700AWP 082421190566AWP 082421111196AWS 082421190279AWS 082427190180 082421120437AWS 082427200140AWS 082421020328AWP 082421140715W11 082421101204AWS 082421101150AWS 082421101434AW1 PINGDU 069CCDDF-AW-AWS_PRIMARY 06BFDE6E-AW-AWS_PRIMARY 082427100322AWS 082421101031AWS 082421101031AW1 082421010291AWS 082421101486AWP 082427030216AWS 082421260055AWP 082421150131AWS 0828550057 0828550063 0828210146 AWSKWUH 082421111167 082421200338AW1 082421200353AWS 082421200354AWS 082421200344AWS 082421200440AWS 082421090747AWS 082421120670AWS 082499130808 2595E4DB-AW-AW31 06A1A25E-AW-AWS_PRIMARY 06A10F3B-AW-AWS_PRIMARY 082499130842 082421230845 082421300094AWS 082421101424 082421150105AWS 082421100767AWS 082421100979AWS 082499250146 082421150276AWS 082427310142AWS 082421310416AWS 082421101494AWS 082421100710AWS 082421101400AW1 082421101137AWS 082427100358AWS 082421130160AS2 082421130160AS1 082421130267AWS 082421130266AWS 082445230008AWS 082421130204AWS 082421130265AWS 082421130259AWS 082421130272AWS 082427130118AWS 4878383 082499230063 082499180112 082499180080 082445160068AWS 082421160487AWS 082421270121AWS 082421070555AWS 082421230515AWS 082421230670AWP 082421120429AWS 082499130793 082421180263AWP 082421070586AWS 082421150303AWS 082499050084 082499050076 082421050207AW1 082427250070AW2 082421140276AWS 082421030426AWS 082421130157AWS 082421050125AWS 082421100955AWS 082427040142AWS 082421310342AWS 082421060214AWS 082421160580AWS 082455040329 082421120851 082421050146AWS 082421180225AWS 082421160793AWS 082421210495AWS 082421160519AWS 082421140786AWS 082427220102AWS 082427250097AWS 082427220136AWS 082427070319AWS 082421100706AWS 082421091109AW1 082421140332AWS 082421230697AW1 082421210424AWS 082421100764AWS 082421120842AWP 082421121097AWP 082421121253 253EFB71-AW-AWS_PRIMARY 082421121078AW1 082421180139AWS 082427150134AWS 082421070564AWP 082421270227 082421121018AWP DFGD 082421140448 082421140815AWP 082421300122AWP 082421140719AWP 082499250139 082421280101AWP 082421070320AWS 082421210732AW1 082421070635 082421140384AWS 082421140746AWS 082421140399AWS 082421140806AWS 082499140328 082421140775AWP 082421140925AWP 082421140748AWP 082421140738AWP 082421140862AWP 082421090648AWS 082421101462AW1 082421180195AWS 082421280090AWP 0683D12E-AW-AWS_PRIMARY 082421121029AWP 06B14BE4-AW-AWS_PRIMARY 082421290186AWP 082421010599AWP 082421310430AW1 082421310487W14 082421310435AW1 082421160748AWP 082421240316AWP 5020167 082455040337 082421160725AWP 082421120991AWP 082421160809AWP 082421100835AWS 082421140780AWP 082421300121AWP 477849201 06A88779-AW-AWS_PRIMARY 082421101327AWP 082421101538AW1 082421101564AW1 082421130247AW1 082421310220AWS 082421210659AW1 082455050085 069C3A9F-AW-AWS_PRIMARY 082421200557 082421091062AWP 082421230640AWP 06AC3220-AW-AWS_PRIMARY 082421310438AWP 082421250316 082421250330W11 082421250386 082421250376AWP 082421120969AWP 082421070679 082421140928 082421290223AWP 082421240302AWP 082421160716 082421111347AWP 082421220234AWP 082421160724AWS 082421060214 082421280100AWP 082421140335AWS 082421160840AW1 082445140034AW1 082426140050 082421140716W11 082421140729AWP 082421140625AWP 082426140049 082421140705AWP PT2M2D1F00WZ HCIT4623117AWS HCIT4913844AWS HCIT4637029AWS HCIT4668793AW HCIT1516613AWS HCIT4946501AW2 HCIT4946501AW1 HCIT5140384AWS AW32-25E76071-AWS_PRIMARY HCIT5224600AWS HCIT5369359AWS HR4052AW01 HR4050AW02 HR4183AW01 HR4156AW01 HR4232AW01 HR4032AW02 HR4008AW04 AW9999 AWS5355992 aw54276182 CY1028AW04 CY1163AW01 AW52109721 AWS5263439 AW52634482 CS1027AW02 CS1027AW01 CS1004AW03 CS1138AW05 CS1013AW01 CS1033AW02 CS1052AW07 CS1042AW03 CS1134AW02 CS1324AW02 CS1090AW11 CS1021AW08 CS1054AW09 CS1024AW10 CS1318AW06 CS1006AW13 CS1006AW15 CS1006AW14 CS1275AW04 CS1046AW01 CS1328AW02 CS1346AW02 600102AW03 600102AW04 CS1356AW03 DK1001AW02 DK1284AW01 DK1044AW08 DK1044SS02 DK1007AW07 DK1044EHLAW DK1044SS01AW DK1065AW02 DK1086AW01 DK1068SSAW DK1318AW01 DK1031AW03 DK1031AW06 DK1031AW04 DK1547AW07 5539363AWS DK1023AW01 DK1044AWS2 12349812 xlnxawnod04 DK1007SSAW DK1007MICRO DK1031SSAW DK1368AW02 DK1368AW01 DK1368AWS6 DK1368AWS5 DK1368AWS3 DK1368AWS2 DK1368AWS1 DK1108AW06 DK1108AW04 DK1549AW01 DK1106AW01 DK1054AW24 DK1054AW28 DK1054AW25 DK1054AW30 DK9054NOD6 DK9054NOD4 DK9054NOD3 DK9054NOD2 DK9054NOD1 DK9054NOD5 DK9054AW01 DK1298AW01 DK1051SSAW DK1051SS02AW DK1059AW01 HCIT4441148AWS HCIT4694283AWS HCIT4761659AWS HCIT4883185AWS 3b05e210 HCIT5042671AWS HCIT5151743AWS HCIT4940566AWS HCIT4983938AWS HCIT4796839AWS HCIT5238829AWS2 EG1526PP04 EG2563AW03 EG2676AW01 EG1540AW04 EG7628AW02 EG1194AW04 EG1194AW05 EG2650AW06 068fc458_AWS_32 EE4012AW02 EE4073AW9 EE4073AW8 EE4073AW7 EE4073AW6 EE4016AW03 EE4073AW4 EE4073AW3 EE4073AW2 EE4073AW1 EE4073AW0 AW52212882 EE4019AW09 AW52576812 AW52261701 AW55465571 AW55214862 AW52261703 EE4037AW5 2528601E-AW-AWS_PRIMARY-publisher_def AW51565422 850AW000098CT1 FI1064AW06 FI1023AW15 FI1153AW03 FI1033AW02 FI1021AW11 FI1025AW09 FI1025AWS1 FI1017AW15 FI1003AW15 FI1003AW17 FI1003AWS1 FI1090AW01 FI1003SS04AWS FI1003UP05AW FI1052AW12 B1529613 X122205817 M4159255 X121122505 M4159250 09600205 M4159480 X307050806 M2813915 M2002633 06239408 M4159717 M4160732 11960503 X62856204 M342976902 10929315 M4159806 M151671014 X151407607 06379910 S4160092 S4160080 06863195 13699726 A6036414 08075859 M4160479 X1640708 10928726 M4476536 A2978305 M4161662 M257272203 M4161852 M4471529 M187068306 M4162219 B6010834 B5483457 M44770AWS01 05993445 M4135325 X167162413 X240779702 06121642 M44849112 M4484987 08133337 M2645712 M1441709 X628532202 M2842601 M1523910 M4163794 M86323611 M4437831 M4500723 05840942 M2045018 M4409434 06529155 13749251 GP1046AW11 GP1081AW05 AW52650252 X171897208 08025090 M4478234 M4164354 M4164351 M359673904 M373312703 M4164834 M216648209 M4478904 M5505619 08792079 M4165613 M4135193 M4135183 A1039636 A1765928 5521067CMICRO X61014012 X143039507 13037380 M466439102 M4166405 M4166669 X122627807 M4018554 15169966 M2072110 G4200414 M9409554 M1754008 X236471602 M4166998 K1680818 11259956 11259955 10181906 M8322906 M4016049 M9431012 M2369016 A4167361 X4479207 X706835018 M4479145 11382586 A41676_AWS 25A3A384-AW-AWS_PRIMARY 48FFFF74-AW-AWS_NODE 48FFFF73-AW-AWS_PRIMARY M4479023 08886744 M196514410 M4167721 M9024418 M6038707 10785023 M4167843 X634445003 X221151202 M221151201 AWS03 AWS04 AWS_EMBEDDED_UV7_PRIMAR 48519C12-AW-AWS_PRIMARY 08259215 M6013228 M4400435 M4400429 M6013236 X05717116 M4168093 X6002819 V2728114 X6002836 06447483 13709812 M4168434 M4168619 M1891710 M4168833 M4168835 X699868010 M4472519 M4500930 M61421922 M61421906 AW53545242 M453783207 M453783206 M453783205 M453783204 M4461519 09252917 M220648615 B1303715 X256517205 12142101 M4169041 13864828 M2311513 A4169319 M2311514 11715304 M4169354 B1768015 07909501 M2055808 M189104206 M9202628 06541937 M551060601 M1220511 M4170019 X129032411 11259949 M4171234 MM4480410 M4458508 F4171510 D2A40755 11730541 08248217 08880015 M4143833 M4143828 A4171576 M4143841 X454682003 M4016571 M4234251 08890679 M4171824 M2753443 AWS7ACRIM M683032103 M4172335 M4172343 08425696 06971425 M347028305 M1558722 X59655542 X59655542_AWSEMB M4206534 AWCREIL1 M2748737 M40230101 M40230112 12294763 10811143 10451561 A62925714 M738353902 M24455805 M4173329 X151054615 M5172561 M5172589 X483037402 X143327902 NV5279AWS P41728AWS2 P41728AWS1 06213440 M4467035 09663352 M6044525 M4173508 M472235202 M411063304 10041105 M4174552 M171587021 08953623 X198234204 M4464822 X76305716 M4175529 M23084343 M4409519 M178185407 M178185414 D4858923-PreProd M4471611 M4481616 13277623 M4500620 M2663618 L2202008 14794938 MQ1234506 M388157304 X387005310 X387171508 MQ1000AW10 MQ1008AW04 MQ1038AW07 AW52305482 AW54068982 AW51694185 05840902 X141549409 M4480269 06529150 06442794 A1570011 A1570014 10714564 X306788202 M4176533 AW50952561 M4176615 09553460 M4482638 08270500 M4399317 M7411120 M2153108 X1617003 M4177345 M1265007 M280444802 M4485118 M4498916 M75608702 X106415410 X106415411 M4489819 X106415413 M4499412 5424157B M4178707 M2427740 M4213408 12753607 M4438416 X4180114 NC1075AW01 GF1006AW07 M386578309 M4548024 M4179507 M1425523 M4500508 A6039717 A4183054 M4181486 X472561004 M1566807 09167425 X832010010 MQ1007AW05 M4481525 M197296409 M4486929 06442798 10136468 M1807243 M4216458 M1807245 M4548545 07948303 X220908811 M4017070 M4017077 MQ1027AW05 A4182040 A4182047 M5173640 M4500409 RE1197AW01 X756087019 M218669012 14551059 M4182835 09698501 RE1044AW04 M4182445 GP1000AW07 M2364527 L1650220 13879637 M5483013 11773551 M4183858 M1261613 B1650233 A4185529 X1590902AWSSPECT X1590902AWS03 X1590902AWS02 X1590902AWS01 11525313 M4184414 M4184876 X90281620 A128275407 13640956 X185801211 M1656638 X687234012 M4186546 X401087502 11524282 M24313808 A5485265 X401876905 A5485281 M8321418 X189112208 25c34935 25b3e41b X180869705 M185142903 07948311 M2399412 12079284 AW51793051 M4187029 M4187732 M4045568 M2080328 M40480202 M4045579 M40480232 X128968011 A1279219 M40480248 M4046587 M4144353 M94838140 M4046565 M4045579AWS4 M4045579AWS3 M4045579AWS2 M4045579AWS1 M4045579AWS5 11525141 M4045579AWS8 M4045579AWS7 M4045579AWS6 13676815 13416997 M9483441 M4188438 M116375119 M41884AWS 15724461 NC1040AW01 M123456 06127968 M9886864 11923460 M4189265 M4189273 X2181510 M2312314 X197096002 M4165452 M194554202 M4207321 M4490114 X116370512 M4191211 M1322134 X117934014 A4190710 M4191128 12162540 M2437611 08264763 09960569 D1215122 M4191041 13271115 M4191513 C4191769 D4191721 X993729009 M2788823 M2800933 M7417004 X255501006 C4191769-CHUAWSX2 C4191769-CHUAWSX1 C4191769-chuawsx4 C4191769-chuawsx3 C4191769-AWS2 C4191769-AWS1 13879630 X201514819 I9237528 08039891 X179963602 M2855815 M4644042 M4644039 M4644050 N4192953 07846722 M4111233 06538714 M4490636 11960510 M54807110 M54807109 M54807AWS02 M54807AWS01 M4194051 12162556 A4194438 M9167004 M1354407 A4194458 M4134651 M6065330 M2838411 M2844565 M2844556 M464254002 X156593006 A4194866 A4194860 08868020 13706179 13848507 M1567003 M4195146 AW49413133 NC1028AW05 A4461322 11259963 M2387123 M2208514 07023730 A1111510 09970640 B4195743 A4493010 M4490804 M4490808 CZ2247020N PF1000AW10 M4491125 X365212303 M2533227 M4037078 M2637320 M54606125 M5484953 M5481251 M5400574 XM5480903 M54849109 M4040092 M40180136 X288892405 M4152263 M1151009 M40180134 A9190174 M40345212 M40345206 M5481257 M4503513 A9502031 A9502016 M2808016 X365212313 M2722627 M40330123 M1400937 M4040556 M40240120 M40240114 M54606123 X976365015 M4019521 M2533233 M1400958 05840927 06956307 06956306 06956305 07960965 06968825 08085681 11730551 10396526 11124555 A91901_24484487_2 A91901_24484487 A91901_24484485_2 A91901_24484485 M2865823 C1282414 M4196652 08486358 08830631 M9185632 6845238 M156048202 M41980125 06875320 M41980110 M41980111_VAW2 M41980111_VAW1 M41980111_AWS2 M41980111_AWS1 M4190111 14523062 13784596 M203785603 A4166970 M252200602 08025087 X86070203 M414408717 M4499115 M188273402 NC1028AWS3 NC1028AWS2 NC1028AWS1 M9490525 M2748224 M41981125 M2717512 M2717523 X731650012 M4491939 M472561002 A4499221 A1588533 X216471208 M4200638 M4008410 X354818503 R4199049 M401913308 A41993102 X453783203 12452750 08228703 M4491319 10926568 X304369404 B1081917 X141389609 M4494810 C4200452 C4200451 M4145028 M4145036 C4200454 11420649 11420648 M1230017 M2036013 X225779102 M2621027 M2621030 08270492 M190091215 RE1045AW03 A4201353 12627098 M408598102 SXAPPLAWS01 07960966 AWSE03 AWSE02 A83754AWS6 A83754AWS5 A83754AWS4 A83754AWS1 M4484754 M22339606 M4484758 M2322273 M2322264 M204493826-2 X970562015 10016937 M8317618 X70053502 M171582012 06913048 06242122 07244263 M4144635 M4144638 C4193359 06373424 13757222 A4463350 X2266306 06842690 06355838 068b3d30 A4495717 08368573 F1669605 M1020821 06632679 GP1022AW05 AW52650292 M484331504 13216847 X220845203 M1377510 11525150 X226587707 X226587709 X226587711 08089971 M315152204 RE1026AW16 A190136612 M2199912 M4202931 M4487547 M4203320 M1460213 X681864012 M4201127 M129645808 X201470802 M2397006 AW53515293 M4204526 M42045AWS A1808110 M4490511 M4476123 MQ1098AW01 X4205007 13061121 M4854029 M4206225 11886229 M4206631 M227137506 M4206641 07919524 M218772806 M4472027 X140985811 RE1200AW01 M1481113 AW52211542 M4202317 srvawschog M440007403 RE1134AW02 AW51632001 M215631210 M098472002 D4858923 H4858989 M1860621 M883443912 D48589AWSRV D4858923-IRC M43920AWS D48589AWS4 D48589AWS3 D48589AWS2 D48589AWS1 gies-ge-aws01 05803158 M91647113 M91647137 PF1029AW01 M386271504 M9624255 M4487713 11524273 M4496962 M2718172 M4209510 M2875119 08310778 M4209933 M203097403 X490049106 M4433017 M220243304 M4475680 X4211010 A1064511 M220243306 X4211070 13848498 M348992903 M4473204 M348992925 X751968028 M172471314 V1083928 X976413010 M1083907 M2228048 M348992932 M348992935 X971285013 08124344 X167162410 X751968039 M3489929AWST M3489929AWSC2 M3489929AWSC1 M3489929AWSCR M3489929AWSC5 M3489929AWSC3 09535820 07946055 13786528 M41718AWS2 M41718AWS1 X129112305 M2848442 M1138214 08328821 06278433 M4215007 M245413304 PF1005AW01 M4497279 M2311412 M42155117 X247551504 M42155102 M205063004 O9885035 M1631421 M4215746 11069585 M136684819 B75162214 D190136602 08025088 M4216044 M6015638 M4216723 05803160 M4016707 M4217638 M4217641 M45052126 M45052126-AWSNODE05 M45052126-AWSNODE04 M45052126-AWSNODE03 M45052126-AWSNODE02 M45052126-AWSNODE01 06461560 M59012303 M1279754 A2943408 X738353908 X738353910 07313353 13699714 M24582603 M4497425 M463396702 M190773209 M41980111 PF1000AW06 GE1222AWS1 RU5308AWS1 682243AW01 geaws22 HC2821AW04 129619AW01 PC3383AW02 PC3383AW01 CC3081AW02 714235SV01 UC7655AW03 530019AW01 VETGEPACSAW 11456923 13410692 UC1590SV03 169766AW01 744459AW01 509982AW01 19670326 13360184 HC0220AW08 221780AW03 HC7511AW04 PC6132SV01 PC5340AW04 HC4370AW03 509940SV04 HC4017AW01 11058430 HC6051AW02 HC6051AW04 HC0074AW10 HC1627SV05 12060725 HC4438AW04 124130SV02 PC4054SS02 PC8118AW17 351482AW01 HC0881AW06 HC0881AW07 250A256B-AW-AWS UC2847AW01 UC2847AW02 UESS01AW02 CC1173AW02 259056AW01 UKFAWS01 GEAWS0 362759AW01 561545AW02 HC1217AW01 HGEL01AW03 416005CS01 HGEL01AW04 PC3741AW04 13236175 12761798 600791AW03 622329AW01 HC3810AW01 556162CS01 805534AW01 HAL406AW01 HC0727AW01 485135AW01 12360876 362520AW01 UHAN01AW19 12204583 HC1384AW01 HEIAW61 UC1989AW04 HDESXAW01 125521AW02 iskh-pa-hoq0008 HJJH01AW01 P00042AW03 HC4929AW04 PC1455AW06 410014AW06 410014AW07 HJJK01AW06 HC1991AW04 HC3055AW02 PC0396AW10 PKOB02AW04 HKOB01AW05 HC7498SV02 HC7498AW01 379798AW01 MHAWS2 MHAW 5396681B HLAH01AW01 PC3827AW05 HLUE01AW03 10863845 624751AW01 505458AW01 5392551A PC0542AW02 507903AW03 826251AW02 HC3295AW05 06808154-AW-AWS_PRIMARY 06805F94_USPAWS02 0689337E HC7669AW03 HCNE01AW03 387758AW01 553160AW02 HPAD02AW03 434487AW01 604529AW02 124448AW01 CVIS-AW01 CVIS-AW02 13756359 487692AW05 UC1559AW01 HC0521AW02 408810AW03 13243786 HC3076AW04 355495AW01 HC2252AW03 12349800 UC4372PC02 509585AW01 HC0565AW01 PC0352AW05 615010AW04 12473045 5492047B AWSVR01 493519AW02 TP5555AW01 126538AW01 HC1282-AWS H73651ADW1 H2036ADW36 H2036ADW40 H2075ADW14 H2192ADW11 H2008AW32 H2060AW13 H2189AW05 H2009ADW80 H2060ADW22 H2036ADW43 H10645AW33 AWSPrimary4860483 H2027ADW30 H2026ADW31 H17641AW37 H2021ADW16 H1141ADW25 H2126AW17 H2126AW13 H2103ADW31 H2049ADW11 H2101ADW28 H2101ADW41 13668527 H1066AW36 H2034ADW22 25A073DE-AW-AWS_PRIMARY GRKOZ1AWS 13791894 13131448 H1787IIS20 H44435AW09 H2197ADW34 H5049ADW9 H1209ADW04 06A376D6-AW-AWS_NODE H2193ADW30 H2095ADW11 H1039ADW54 H1383ADW41 H4676ADW04 H2066AW05 H2044ADW8 H10391AW19 H2032AW19 H1943AW24 H5350ADW07 H4443IIS22 H2017ADW29 H2017ADW12 250F5E6B-AW-AW31 H2014ADW09 H2010ADW19 H2016ADW23 H2014ADW44 H1105ADW33 H2015ADW28 H10392AW55 H10392AW46 H10392AW43 H7490ADW02 H0005ADW09 H4478CT07 H4478ADW08 H4478IIS15 H2073ADW18 H7274ADW1 H6252ADW5 H1197ADW12 H2063AW17 HCIT4900960AWS HCIT4494109UV GF1005AW04 HCIT4961536AWS HCIT4981607AWS 0828210088AWS1 0828210110AWS1 0828550083 0828550071 0828550099 0828210125 0828550072 0828550082 0828550074 0828550079 0828550094 0828550075 0828550105 0828550114 0828550127 0828550066 PMH00000 5055485 CA084B70-AW-AWS_PRIMARY 5104206 0828550070 0828550077 0828550085 0828550089 0828550095 0828550086 0828550073 0828550117 0828550052 0828550052IB01 251E6AEA-AW-AWS_PRIMARY 25D07BB8-AW-AWS_PRIMARY 25C7065C-AW-AWS_PRIMARY 0828550067 600117AWS1 HU2008AWS1 HU1098AWS1 HU1099AWS1 HU1087AWS1 HU1083AWS1 HU1573AWS1 HU1098AWS2 HU1571AWS1 HU1038AWS1 HU1099AWS2 HU1638SW07 HU1028AWS1 HU1028AW02 HU1697AWS1 HU1687AWS1 HU1506AWS1 HU1488AWS1 HU1939AWS1 HU1433RW02 HU1749AWS1 HU1243AWS1 HU1180AWS1 HU1766AWS1 HU1294AWS1 HU1754AWS1 HU1635AWS1 HU1103AW01 HU17095825AWS01 HU1008AWE02 HU1008AWE01 HU1008AWS1 HU1113AWS1 HU1113AWS2 HUGYOAWS1 HU1036AWS01 budgeawh1 BUDGEAWH01 HU1711AWS1 HU1674AWS1 HU1672AWS1 HU1729AWS1 HU1137AWS1 HU1047AWS2 HU1838AWS1 HU1039AWS1 HU1677AWS1 HU1677AWS2 HU1677AWS3 HU1677AWS4 HU1688AWS1 HU1500AWS1 HU1641AWS1 HU1160AWS1 HU1162AW02 HU1162AWS1 HU1831AWS1 AW52982104 083045820668221 083045067531522 083049241749523 083045828489822 083049822150013 083027603507818 083021863624122 083021869528222 083045047339623 083045046480422 083049249412022 083055247345716 083049241089623 083049240825123 083049045853321 083049862149224 083049025445319 083049820009914 083049840009514 083049824996815 083049821002812 083049823888717 083049828489822 083049609351622 083049246166124 083049289279520 083049289269620 083049806709414 083049805881615 083055803467216 083049208307324 083049805213421 083055804454717 083049809186321 250F7AF9 INAWX26381827 083049024450218 083049003436820 083049006621222 083049060676524 083049864867815 083049022842422 083049004465025 083049007217323 083049280426725 083049280425025 083049281851225 083049841233721 083049840083822 083049049673523 25E2BE76-AW-AWS_PRIMARY 083049045753717 083049040124722 083049600149523 083049602404423 083049608170521 083049042728625 083049022071225 083049602071025 083049602768225 083049242220225 083049822219525 AWBUCLAB5 AWBUCLAB4 083049860991623 083049024926414 083049024212823 083049027563220 083049802459923 0004410586F 083049823408018 083049823428618 083049201816716 083049206522519 083055201766451 083049202671920 083049204674620 083055207778621 083049203474520 083049207526721 083049202841422 083049214735622 083049205726922 083049028284925 083049204986924 083049204198024 083049217722224 083049062312824 083049064985524 083055211206517 083049000974917 083055004223219 083049066530918 083049005023718 083049000011318 083049004075620 083049006972719 0004819173 083049001558421 083049009590421 083049046434521 083049006164323 083049008126819 083055009658025 INAWX26596166 INAWX26625265 INAWX26369609 083055806203119 083055806300419 083049843326720 083049020893221 083049203223320 083049049374219 083049846311918 083049025445418 083049843326818 083049603222619 083049843326919 083049826312120 083049603222720 083049282198819 083049211390522 083049008098319 083049105445522 083049028920021 083049603327121 083049102488421 083049001409223 083049248848223 083049006156021 083049288925124 083049288899524 083049108893524 083049108894924 083049008647724 083049601367924 083049066611319 083049103407523 083049606223420 083049206614024 083027603405819 083049212952419 083049008557723 083049283416824 255B1312-AW-LICENSESERVER 083049280081224 083055601980920 083049067033623 083049247288618 083049068392721 083049244661325 083049244661225 083049248698116 083055861021012 083049207770423 083049603507818 083049103407623 083049284039818 083049860018714 083049864282615 083049806685319 083049861088020 083049869888823 083049867220024 083049609220321 083049869574523 083049869610725 083049867983525 25CD42B5-AW-AWS_PRIMARY 083049044118924 083049801928023 083045066827522 083021066611319 083045049899222 083021803587721 083027805213421 083049801114511 083021805701523 083021801367322 083045843075522 083055868428112 083049863640823 083021602668522 083049860129622 083049860129122 083049862540324 083045028335823 083021603327120 083021609220321 7982AW0098CT1 7982AW0091CT1 2586D363-AW-AWS_PRIMARY RSHSPACSAWS JAMPANG 2547B0A4-AW-AWS_PRIMARY 2565AE7C-AW-AWS_PRIMARY RSMKCKR RSMKBTI 2572208A-AW-AWS_PRIMARY MIKATEGAL 253DA2AC-AW-AWS_PRIMARY 068356CA-AW-AWS_PRIMARY 2512963C-AW-AWS_PRIMARY 25A66FAB-AW-AWS_PRIMARY TCH 5167974 5135068 255378B7-AW-AWS 5449765 253DED7A-AW-AWS_PRIMARY GLDEMO 25E1D9A5-AW-AWS_PRIMARY 25A03A91_AW_AWS_PRIMARY RSUNPADJATINANGOR_AWS 7982AW0021MI1 25A67FEA 0696B47A 251718A0-AW-AWS_PRIMARY 7982AW0022MI1 FC1639F6 TempSysId1 RSMKWR RSMKKMY RSMKGSE-AW RSMKBNT RSMKSBY RSMKCBR RSMKBKS 258898E5-AW-AWS_PRIMARY 25372E99-AW-AW31 25BB0EAC-AW-AWS_PRIMARY 2550D75C-AW-AWS_PRIMARY RSPSAW 4809558 25A607EA-AW-AWS MANOKWARI 253BE4FE RSPOLRI RSBETHSAIDA 254528D5-AW-AWS_PRIMARY 015R67 RSUD_Sijunjung RSWaled 015R71 25149E96-AW-AW31 RSUDSijunjung RSUDSintang 254E3D5B-AW-AWS_PRIMARY 25F53471-AWS_PRIMARY 25D410F6-AW-AWS_PRIMARY 250B6DE4-AW-AWS_PRIMARY 25AEFE3A-AW-AWS_PRIMARY 25A49131-AW-AWS_PRIMARY RSALRML RSUDWN 25B6BE64-AW-AWS_PRIMARY 2598D142-AW-AWS_PRIMARY 25C2B04A-AW-AWS_PRIMARY RSUDMERANTI RSUDPDGN RSUDPASIRIAN RSUDSANGGAU RSUDDEPOK RSJAW 4856057 KOESMAAWS 257BC702 4794756 25D06B06-AW-AWS_PRIMARY 251C7AD5-AW-AWS_PRIMARY 5169592 06AD050A-AW-AWS_PRIMARY 25DB67A7-AW-AWS_PRIMARY 25EC4B82-AW-AWS_PRIMARY 2561a5cd 2556B764-AW-AWS_PRIMARY 257071EF-AW-AWS_PRIMARY 25264970-AW-AWS_PRIMARY 2529F3FB-AW-AWS_PRIMARY 7982AW0026MI1 MHTG 5378714 2533D6F2 AW52081272 01PAC10 5435069 IQ1885PT01 IE1263AW01 IE1010AW12 IE1010AW11 IE1005AW07 IE1026AW02 IE1038AW03 81100-AWSERVER-02359611 IE1083AW01 IE1031AW07 IE1219AW01 IE1228AW03 IL1510AW01 IL1061AW01 IL1020AW11 IL1067AW01 IL1018IT07 IL1062AW06 IL1062AW05 srvp7aw srvge-aw 069C238A-AW-AWS_PRIMARY AsAWS3220 06AC76D2-AW-AWS_PRIMARY AWSERVER32_EXT42 IL1041AW14 IL1484AW01 IL1484IT09 IL1484IT08 IL1028AW12 ASLX160V 06AC869C-AW-AWS_PRIMARY aslx1670v IL4761AWS1 IL1048NM08 IL1048AW01 A5671006 A5220726 A5461130 B51673AWS32N1 B51673AES32N2 01429315 10760535 A51043206 A51043291 A5512810 A51272145 A6184806 B5368633 B53686 06239152 D5343404 D5343402 CASTELFRANCO 06AB062F-AW-AWS_PRIMARY B5359422 C5103515 09526339 A156230203 09553462 13043140 C5343139 A5200933 A5174430 A5162303 09647366 H153303404 13784606 A5714631 A5360483 A53604101 A51273189 A5992716 A5109144 12218978 12247813 A5128563 B5368914 C5294651 A5216884 11435365 C5398431 B5332015DC1 B5809218 11960525 12123846 itclaspsrvawsv A417206502 A5714107 A5649022 A5366631 12952414 A402682102 A5383317 A5179210 A171995616 A171995619 10686279 K5345907 A119731006 10931332 13640986 A5912919 A5912925 A5137517 A6470605 13313540 A534303503 AWCT01 11961029 12068145 B5810532 09553472 14795013 A103987102 A103987103 H5382020 A5124549 A5124553 A5325435 A5811619 A51560 CSAW01 B227017702 08933544 12114543 09909103 A5274210 13432860 ITLOM3AW ITLOM2AW ITLOM1AW A5364713 A5251328 09721018 A192528604 09957583 B5352010 A343134112 4402942 12427457 A5128766 A5128772 A5128774 10863847 A104057305 A5810833 A5124930 13352739 A5112422 A5165734 A5165742 A209695502 B5342739 A556284803 A58122129 B5813720 A5207235 A58122162 SiteID-AWServer-01454808 suc-cis-ap01-p04 suc-cis-ap01-p03 suc-cis-ap01-p02 suc-cis-ap01-p01 D5803108 B5811466 B5811472 B5811471 09761998 A5111152 A5176617 13024445 A5537303 A5236259 A5228524 08957085 B5332705 09564199 AWS32 AWSBIO01 HQV442 HQV441 HQV440 25E7CC9E-AW-AWS_PRIMARY 10686272 GEAWSOEI-01 09631087 C5237833 A5123564 A5123572 12421057 09970935 A5113645 E5803916 B5801740 B5332015DC3 09740720 11960537 12218715 09632507 A5219606 A5219617 A5300752 A5361309 A53613115 09167430 B5659127 09495991 10249094 A204116602 A5805563 10915321 A5327030 A5129824 A5367431 10249099 A5111921 ITTOR3AW ITTOR2AW ITTOR1AW ITPAD3AW ITPAD2AW ITPAD1AW A5265814 F5367807 D5117615 09897961 09577015 11826589 C5354803 A6142013 A56364178 A51299131 A5381511 A5315591 A51205105 A6055605 A5108735 A53155114 A5624155 A5124164 A5147207 A5158822 A51299178 10931328 A5292331 D5114918 11960500 A5483707 08903131 B5367328 A5145935 A54013128 D5111652 A5820147 A5820154 A5320165 B5320436 A5320426 C5126836 A5359622 A5820185 A5697726 B5320127 B5140642 A53796127 09553467 09657915 09772554 D5820808 11960458 A5600103 A5348014 A670060225 A5362308 C5346712 D5109654 A5129131 A255888103 A173433245 AWS32IOV ITIRM1AWS 08980221 AW32CAMPO A5719870 A5163331 A5163335 A23086008 A139476003 A258001702 09435697 09692300 B5300713 A127292416 11960448 11960447 08953461 A5189370 A5366325 10729497 06B6B374-AW-AWS_PRIMARY C5210510 09599964 13634472 C5323539 C5323542 A51142156 B51142165 A51142154 10198977 08915985 A5367323 A159405610 A5229915 A5730913 C5367808 10041154 A5115183 11960466 A6340103 A5300208 09943706 5314024A 14490653 A224857357 A5139657 A51581134 A51581129 A5269719 A5391009 A5196822 A5118556 A5196830 A256767602 A58091SS01 A5332121 A5332120 09672472 A5379318 B5603950 A5902010 A422328501 A5138133 A5690010 A5678733 A5333083 A5743104 A51529176 C5358903 A56183195 A56183193 A5144931 C5618323 A5721915 A139569305 A5688113 A5721922 A53330149 ROMAIDI A57219 sfnaws01 09641114 5293842A2 5293842A1 09052350 12764113 11961146 pacsaw01 pacssen02 pacssen01 11388609 slp-pacsaw02 slp-pacsaw01 A5691235 B5320627 B5320634 B5320633 A5640203 10249097 12952416 A5383110 A5155169 A5138520 B5332015 E5373404 A5261115 A5261111 A196626506 A196626503 12123856 A5238826 A5238830 A665033302 A5725520 11960962 A51026120 A51026133 10729505 A5267112 12918968 A5141735 A5402121 A5402119 C5343508 A5103544 A757633202 A5189735 A5128420 B5580616 A5389825 A99263314 B5580651 09698520 09993033 09897949VM A5634730 A51043 SRV-AWS02 A5739715 A5407812 A5164528 12242336 A5105147 A6215807 A378428333 B5120746 09970636 A5208214 A51014122 A51014154 09647489 A57198 A5635622 B5146041 A6195611 A255223004 10050106 069C84C2-AW-AWS_PRIMARY 069C79A0-AW-AWS_PRIMARY A5728826 B203186402 A5105758 A5196350 A99155715 A5343519 A5117595 EW0112 SS0003 AW00107 EW0017 EW0030 SS0019 ST0010 EW0212 ST0030 SS0031 EW0306 EW0036 ST0046 EW0116 SK0030 EW0213 EW0004 EW0377 EW0376 SK5001 SR0002 ST0024 EW0114 SK0032 SS0029 EW0033 SS0039 ST0028 EW0370 SK0012 EW0370-AA-AW_Node EW0356 ST0059 SS0012 SL0022 EW0313 EW0384 EW0358 EW0357 SS0014 EW0013 EW0060 EW0035 AW00282 ST0050 SH0021 EW0353 SK0028 SL0013 SS0023 SS0022 SS0030 ST0054 SK0011 ST0045 SS0038 EW0206 EW0383 EW0364 ST0017 ST0015 EW0024 SL0020 EW0220 EW0105 SL0018 EW0031 EW0054 ST0051 ST0062 EW0007 SS0021 SL0033 ST0016 SS0027 EW0333 EW0393 SL0028 SS0026 EW0120 EW0104 SH0027 SS0004 SS0006 EW0324 EW0318 EW0214 EW0326 EW0305 ST0003 EW0053 EW0052 EW0387 EW0390 EW0390-AA-AW_Node SS0002 SH0010 AW00330 SK0031 ST0058 SR0003 ST0002 EW0113 EW0011 EW0382 EW0382_node EW0027 EW0045 SL0019 EW0335 ST0053 EW0389 SH0001 SR0004 ST0021 ST0034 EW0341 EW0340 EW0038 EW0039 AW00139 SS0024 EW0332 EW0363 JV0939 EW0325 EW0391 EW0321 SK0003 EW0339 EW0338 AW00027 ST0039 EW0019 ST0049 SS0025 SS0034 ST0052 EW0319 SS0035 EW0345 EW0002 EW0351 ST0040 ST0042 EW0020 SS0040 ST0057 EW0051 EW0041 EW0034 ST0007 SS0005 SS0020 ST0023 SS0008 ST0048 EW0320 EW0006 EW0316 EW0219 EW0050 EW0379 EW0378 EW0107 EW0347 PT0052 SH0014 EW0223 AW00225 EW0322 EW0329 EW0115 EW0349 ST0055 EW0349-1 EW0836 EW0209 AW00172 SK0014 SK0002 ST0029 ST0043 SS0028 SS0018 SL0027 SK0027 EW0025 EW0121 SL0030 SS0041 EW0359 ST0019 ST0060 ST0018 SL0021 AW00073 ST0012 ST0020 EW0001 SS0016 EW0003 SK0006 EW0344 EW0222 EW0037 EW0016 CT00346 CT00102 CT00160 NM00035 TAI106 BK0079 NM00042 CT00501 CT00177 CT00146 TAI062 BK0160 CT00308 TAI075 CT00147 TAI020 NM00024 CT00161 CT00251 TAI099 TAI066 PET00028 PET00045 CT00284 PET00030 000037 HD9230 CT00264 NM00063 CT00211 EW0312 CT00035 NM00014 CT00156 CT00163 CT00153 CT00210 CT00157 PET00041 PET00026 CT00134 CT00214 CT00125 CT00307 CT00165 CT00051 NM00023 CT00121 CT00434 PET00049 CT00268 TAI093 EW0131 CT00471 KW0007 AW00040 SL0012 AW00084 SS0013 EW0327 EW0392 SH0019 EW0350 AW00024 SL0014 SS0015 ST0044 SK0013 EW0123 SH0023 ST0041 EW0314 ST0006 SR0006 SL0025 SR0005 SL0017 EW0317 EW0331 EW0330 EW0023 EW0303 EW0032 EW0385 EW0388 EW0388-AA-AW_Node SK0034 SK0021 EW0301 EW0059 EW0103 EW0015 AW00130 SL0029 AW00045 SK0029 EW0352 EW0352-2 SL0031 ST0032 SS0046 SK0022 EW0315 ST0047 SK0017 SK0005 EW0101 ST0005 EW0368 EW0367 SR0001 EW0012 AWS00285 EW0348 ST0026 ST0061 EW0355 ST0035 EW0056 EW0102 SK0015 SK0035 EW0057 SK0009 ST0022 EW0221 AW00067 EW0217 EW0361 ST0038 EW0373 EW0372 EW0043 SK0020 AW00209 SK0010 EW0336 SL0016 EW0124 SK0001 SS0009 SS0036 SL0026 EW0310 EW0224 SK0025 EW0366 EW0365 SS0045 SS0033 SH0018 EW0062 SH0018H SS0044 EW0055 SH0016 EW0346 ST0013 SK0023 ST0056 EW0005 SL0010 EW0047 SK0016 EW0106 SH0024 SH0002 SH0025 SH0028 SS0011 SS0001 SS0017 EW0202 EW0010 EW0302 EW0328 EW0211 EW0337 SK0019 EW0026 EW0018 SS0032 EW0309 EW0354 EW0028 EW0203 EW0362 SS0037 EW0342 EW0334 ST0001 EW0046 EW0369 EW0042 EW0044 EW0029 SK0018 AW00167 EW0058 EW0048 SL0015 AW00321 SH0017 EW0343 EW0040 AW00044 EW0216 EW0311 EW0381 EW0380 EW0049 EW0014 ST0025 ST0011 ST0027 ST0014 AW00302 KA2737AWS1 KA3091AWS1 KA1174AWS1 KE6280AW01 To be provided 0850210938AS1 0850210994AS1 0850270650AS1 0850211007AS1 0850211012AS1 0850211063AS1 0850270756AS1 0850211069AS1 0850450226AS1 0850450223AS1 0850211135AS1 0850210984AS1 0850211023AS1 0850450215AS1 0850210923AS1 0850211035AS1 0850210887AS1 0850210933AS1 0850060773AS1 0850270612AS1 0850270606AS1 0850450202AS1 0850210962AS2 0850210962AS1 0850270720AS1 0850270775AS1 0850270521AS1 0850211064AS1 0850210959AS1 0850450222AS1 0850210953AS1 0850211001AS1 0850211017AS1 0850270658AS1 0850450208AS1 0850211129AS1 0850270584AS1 0850270509AS1 0850210875AS1 0850550300 0850210900AS1 0850210932AS1 0850210899AS1 0850210918AS1 0850210936AS1 0850210939AS1 0850450200AS1 0850210948AS1 0850450198AS1 0850270573AS1 0850270546AS1 0850450203AS1 0850270614AS1 0850260184AS1 0850210966AS1 0850210991AS1 0850210989AS1 0850270662AS1 0850211044AS1 0850211048AS1 0850211022AS1 0850210947AS1 0850270719AS1 0850450219AS1 0850211071AS1 0850450220AS1 0850270776AS1 0850211104AS1 0850211084AS1 0850270577AS1 0850450229AS1 0850211115AS1 0850270799AS1 0850450230AS1 0850211116AS2 0850211116AS1 0850211128AS1 0850450225AS1 0850210946AS1 0850060777AS1 0850211013AS1 0850211050AS1 0850270792AS1 0850210944AS1 0850211068AS1 0850270613AS1 0850211121AS1 0850210815AS1 0850211052AS1 0850450224 0850210849AS1 0850210853AS1 0850210876AS1 0850210874AS1 0850210890AS1 0850450187AS1 0850450186AS1 0850210942AS1 0850270558AS1 0850210957AS1 0850270574AS1 0850270561AS1 0850210985AS1 0850210963AS1 0850260196AS1 0850210990AS1 0850210914AS1 0850211008AS1 0850270698AS1 0850211043AS1 0850211015AS1 0850450207AS1 0850450206AS1 0850211003AS1 0850211053AS1 0850211065AS1 0850211056AS1 0850211075 0850211029AS1 0850450221AS1 0850211079AS1 0850211081AS1 0850211083AS1 0850211103AS1 0850211094AS1 0850270590AS1 0850211108AS1 0850211123AS1 0850450231AS1 0850270781AS1 0850211038AS1 0850060762AS1 0850210764AS1 0850210732AS1 0850210757AS1 0850210755AS1 0850210807AS1 0004971617A 0850211128AWSEHL KRAS54877111 0850210735AS1 0850210746AS1 0005435859A 0005452095A KRAS55440261 KRAS55567071 0850210806AS1 0850210835AS1 0850210751AS1 0850210720AS1 0850270481AS1 KRAS54893311 KRAS55395581 KRAS55199661 KRAS55749261 0005470350A KRAS55252361 KRAS54972221 KRAS55567021 KRAS55567011 0850450175AS1 0850210938AWSEHL 0850210727AS1 0850210750AS1 0850210773AS1 0850210690AS1 0850210728AS1 0850210725AS1 0005443399A 0005420279A 0850211121AWSEHL 0850211054EHLAWS KRAS55424681 KRAS55188901 0850270436AS1 0850210733AS1 0850210793AS1 0850210710AS1 0850210740AS1 0850210830AS1 0850210752AS1 0850270504AS1 AW52642138 KW1013AW15 KW1100AW04 KW1067AW01 KW1011AW06 KW1005AW20 KW1005AW16 25207406 KW9580AW 96501PAC91 96501PAC90 KW101 KFASAW 965K20 KW1007SV30 AWS9216 KHAWS9600 KW9755 KW9278 256E6168-AW-AWS_PRIMARY KW1004AW36 KW1155AW06 LV4041AW03 AW55509562 LV4000AW09 AW51814481 AW51900641 LV4001AW07 LV4001AW10 AW52182292 AW54274691 AW54892842 LB1331AW09 961PAC10 BMCAW 4496308 OM1007AW03 LT4001AW16 LT4129AW01 LT4001AW17 AW52186341 AW53459591 LT4110AW01 ltdatvln-aw01 LT4006AW03 LT4124AW02 AW54465401 LT4003AW02 AW54465301 LT4028AW02 aw55718721 LT4089AW02 LT4069AW02 LT4069AW01 LT4122AW01 LT4000AW15 LT4026AW07 LT4079AW02 AW52568872 PLIMCRZE1AWS AW54934902 110807OT13 200012OT01 1553AW0035CT2 0004190980D 06AF654F-AW-AWS 1553AW0020CT2 1553AW0022MR1 sdgpacvpaw03 sdgpacvpaw02 06969957-AW-AWS_PRIMARY 1553AW0025CT1 0005449852A 0834550071AWSER 1553AW0011CT1 1553AW0029CT2 PPUMAW35 PPUMAW34 PPUMAW33 1553AW0044CT1 HPUAWS 1553IT0014 IIUMMC-AWS 486CD21A-AW-AWS_PRIMARY 1553AW0030CT1 1553AW0023MR1 0834490066AW2 069CE52E-AW-AWS_PRIMARY 0834490066AW1 0691E817-AW-AWS_PRIMARY 0005456789A 1553AW0034CT1 IOCKL-AWS 1553AW0024CT1 06BAA061-AW-AWS_PRIMARY 0005456791A 0691699D-AW-AWS_PRIMARY 0005456793A 1553AW0033CT2 MT1014AW04 MALTA10 2M223804MW HCIT4753028AW HCIT4376494AWS CJRBX1800011AW HCIT4665855AWS01 HCIT4665855AWSSUR HCIT4665855AWSMEX HCIT4820102AWS HCIT4665855AW HCIT5497413AWSSUR HCIT4429184AWS HCIT4825663AWS HCIT4579183AW 0004568789 HCIT4468687AWS HCIT4711469AWS CTDGV1800001AW HCIT4730432AWS HCIT4784198AWS AW52863172 MD1055AW02 MD1076AW01 MD1145AW01 AW52900192 AW52863192 AW52554872 AW52863072 AW52863123 AW52862781 MA1841AW01 AW49901282 MA1942AW01 AW48582681 AW52293331 MA1047AW03 MA1780MR01 MA1134AW03 MA1509AW01 MA1955AW01 MA1884AW02 AW50547782 MR1002AW02 MA1908AW01 MA1938AW01 MA1421AW04 AW51685322 AW51674972 AW52298392 AW50497272 MA1493AW04 AW52509941 MA1779AW01 AWS47942321 D90FD1F3-AW-AWS_PRIMARY AW53288172 AWS49900971 MR5340231AW1 AW53401992 AWS53057821 MA1148 AW52300402 AW52299022 AW52299021 AW52298832 AW52293211 AW5226643 AW1516 AW54169842H AW54169842E AW54169842B AW54169842A AW54169842 AW53198711 AW54169842G MA5361645AWS AW53932381 AW52299621 MA1339AWS01 MA1895AW01 MA1548AW05 AW51168922 AW52294081 AW49900202 MA1881AW03 MA1054AW03 MA1864AW01 MA1420CT02 AW54169842F MA1843AW03 MA1528AW03 MA1528AW02 MA1690AW04 MA1855AW01 MA1135AW02 AW54582352 MA1711AW04 MA1818MR01 MA1818AW01 MA1956AW01 AW50554842 MA1877AW06 MA5520791AWS NPAWX26627367 083049007184822 AMS03200 AMS1323 DBS0823 DVT01201 ENS0121 GRN1121 HLM06201 LWD0322 LDD0420 MST0121 RTD0192 RTD01206 RTD0722 SPK0125 TLB02202 ZWL0620 NZAWX25209812 85103833082 85178504240 85202521492 85103832554 85176909731 83180471123 85136254180 859AW000226CT1 85204792617 85170051006 85205055892 859AWS000189CT1 859AW000189CT2 85116842625 85109909109 85166750480 85AW0000011CT2 85AW0000011CT3 NO1054AW01 NO1194AW01 NO1146AW06 NO1011AW20 NO1146AW07 NO1070AW01 NO1136AW02 NO1018AW02 NO1028AW01 NO1102AW01 NO1087AW01 NO1012AW05 NO1051AW03 NO1043AW06 NO1047AW01 NO1260AW05 NO1043AW07 NO1016AW13 D26C48A0_AW_EDISON_AWCT NO1259AW03 NO1035AW05 NO1035AW06 NO1088AW04 NO1068AW12 NO1068AW11 OM1025AW01 AW52216701 DIWANSYSID1 AW53433002 D9311320-AW-AWS AWS54203001 OM1013AW02 OM1010AW03 PK1689AW01 AW51128742 AW51129812 PK1365AW01 PK1365AW02 PK1722AW01 PK1623AW01 AW53613072 AW52673302 AW53823302 AW54574622 AW54496572 PK1548AW02 PK1122AW01 PK1003AW01 PK1385AW02 PK1005AW03 PK1634AW01 PK1596AW01 PK1367AW03 PK1124AW01 PK1462AW01 PK1569AW01 PK1453AW01 HCIT4369313AWS HCIT4192962AWS HCIT4002500AWS To be provided HCIT4714397AW HCIT4663361AWS HCIT5153970AWS HCIT1505695AWS HCIT4469093 HCIT1581739AWS HCIT4274455AWS 1557AW0181CT2 1557AW0036MR1 1557AW0049MR1 1557AW0232CT1 1557AW0086CT2 1557AW0047MR2 PHAWX24817700 1557AW0158CT1 1557AW0038MR1 1557AW0236CT2 25D17424-AW-AWS_PRIMARY PHCAW10 1557AW0033MR1 P00141AW01 PL1203AW01 PL1398AW05 PL1398AW06 PL1398AW07 PL0398AW01 PL0353AW01 PL3057AW05 PL1476AW07 PL0158AW01 PL1476AW06 P00021AW01 PL0677AW01 12650913 PL1696AW05 PL1732AW04 PL0294AW01 PL1314AW02 PL0688AW01 PL1770AW03 PL1793AW02 PL1237AW02 PL3198AW02 PL1584AW07 PL2256AW01 PL1707AW03 PL1031AW03 PL1693AW02 15560095 PL1736AW01 600040AW01 PL0321AW01 PL1418AW01 PL0975AW01 PL1023AW07 PL1310AW04 PL1023AW08 P00041AW02 PLIMCGDA1AWS PL1345AW03 PL1754AW04 PL1580AW08 PL1580AW13 PL1580AW10 PL1753AW02 P00570AW01 PL1724AW03 PL1647AW05 PL1816AW03 PL3550AWS1 PL1673AW05 PL1661AW01 PL0856AW01 P00525AW01 P00752AW01 PL1689AW05 PL2711AW08 13192578 PL1394AW05 PL0580AW01 PL1602AW03 PL1394AW08 PLIMCKIE2AWS01 PL1742AW01 PL0760AW01 PL3895AW02 PL1575AW03 PL0659AW01 255A5D8B-AW-AWS_NODE 25B52C1C-AW-AWS_PRIMARY PL2822AW02 PL2822AW04 PL2822AW06 PL1772AW02 PL0302AW05 PL1609AW10 PL1609AW09 PL1019AW05 PL1698AW03 PL1764AW05 PL0302AW03 P00007AW01 P00007AW02 PL1016RW01 PL1972AW03 PLIMCKRO1AWS01 PL1709AW03 13327808 PL3007AW01 PL1607AW06 PL2190AW01 PL1751AW04 PL1142AW01 PL1321AW05 PL1321AW04 PL0700AW01 PL1111AW01 PL1008AW16 PL1312AW08 PL1810AW01 PL1312AW11 PL1312AW10 PL2832AW01 13092652 13092490 13829181 PL1178AW02 PL1178AW03 PL3334AW02 PL0483AW01 P00032AW01 PL0529AWS1 PL1590AW02 PL3243AW03 PL0215AW01 P00027AW01 PL1797AW02 PL1773AW02 13092479NANO PL1563AW02 PL1059AW01 P00495AW01 PL1103AW01 PL1426AW03 PL1719AW04 PLIMCKIE1AWS PL1165AW01 PL1494AW02 PL1552AW03 PL1524AW03 PL3116AW05 PL1601AW01 PL1502AW04 PL1502AW03 13137819 PLIMCRAD1AWS01 PL0030AW01 PL1190AW09 600034AW04 PL1491AW13 PL1337AW03 PL1190AW10 PL2027AW01 PL1445AW01 PL2342AW05 PL1651AW02 PL1651AW03 PL1651AW05 PL1107AW12 P00786AW01 PL3859AW01 PL1150AW11 25FDDCC0-AW-AWS_NODE PLDATWAR-AW2 600151AW06 PL3331AW05 PL0635AW01 PL1007AW03 PL1234AW04 PL1050AW03 P00326AW01 PL0080AW03 PL1840AW01 PL1640AW04 PL1640AW06 PL1599AW02 PL1684AW02 PL0217AW01 PL1692AW01 08081899 PL3182AW02 PL1655AW02 PL1774AW03 PL1606AW05 P00329AW02 PL0195AW06 P00467AW01 600092AW10 PL1176AW01 PL0399AW01 PL3811AW01 PL1169AW04 600092AW11 PL0714AW02 PL1010AW03 600042AW20 PL1147AW03 PL1600AW06 PL1600AW05 PL3529AW06 PL1316AW01 600062AW05 600062AW06 P00212AW01 PL2057AW01 PL1065AW08 PL1523AW04 PL1674AW08 PL1040AW03 P00549AW02 P00549AW01 PL1780AW07 P00792AW01 PL2773AW05 PL1709AW04 PL3811AW04 600042AW25 06987654-AW-AWS_PRIMARY PLDATWAR-AW1 PLIMCWAR3AWS PLIMCPOZ1AWS01 PLIMCSZC1AWS01 13864822 PL1725AWS1 PL1377AW01 PLIMCWRO1AWS01 PL1544AW05 PL1060AW01 PL0739AW01 PL3254AW06 PL3254AW05 PL1672AW15 P00674AW01 PL3124AW03 P00318AW01 P00694AW01 P00258AW01 PL2975AW01 PL2975RW01 600144AW08 PL1730AW02 PL1355AW02 PL1025AW06 PL0001AW03 PL1025AW07 PL2130AWS1 254C14CB-AW-AWS_NODE 255214A1-AW-AWS_PRIMARY PER00498 13377489 PPR99427 PER08330 PER59107 PER50336 PPR29813 PPR59078 PER00740 PPR88682 PPS51803 PPR590119 PPR59081 13377491 PER08017 PER08762 PPS79501 PPR51812 PPS17603 PER19030 PPS28216 13756110 13313881 PPS17714 PPS20812 PPS54000 12408774 HCIT5219986AWS HCIT5308057AWS AM14FGP168 QA1151AW02 RE1047AW02 RO1617AW01 RO1228AW02 13284370 RO1262AWS1 RO1227AWS1 RO1150AWS1 RODATBV-AW01 RO1472AWS1 RO1470AWS1 RO1562AWS1 RO1252AWS1 RO1328AWS1 RO1487AWS1 RO1487AWS2 RO1159AW02 RO1669AWS1 RO1496AW02 RO1581AWS1 RO1677AWS1 RO1336AWS1 RO1674AWS2 RO1668AWS1 RO1688AWS1 RO1593AWS1 RO1172AWS1 RO1098AW03 RO1098AW05 RO1672AWS1 RO1115AWS1 RO1079AW01 RO1332AW01 RO1417AWS1 RO1102AW04 RO1210AW01 RO1178AWS1 RO1015AWS1 RO1595AWS1 12770365 RO1255AWS1 RO1587AW01 RO1634AWS1 RO1602AW01 RO1219AWS1 RO1362AW01 RO1231AW03 RO1434AWS1 RO1693AWS1 13216826 14836460 RO1626AWS1 RO1205AWS2 RO1365AWS1 13132789 RO1314AWS1 RO1629AWS1 RO1563AWS1 15659048 RO1692AWS1 RO1613CT01 RO1656AWS1 RO1459AWS1 12780454 13127930 12770363 RO1042AW01 RO1194AWS1 RO1627AW01 RO1675AWS2 12948042 RO1426AWS1 RO1096AWS1 RO1188AW04 RO1544AW02 RO1694AWS1 RO1564AWS1 13217490 RO1662AWS1 12820894 RU3093AWS1 RU8060AWS1 RU6522AWS1 RU7270AWS1 RU1829AWS1 RF0001AWS3 RU2137AWS1 RU1284AWS1 RU8719AWS1 600107AWS2 600107AWS1 RU6329AWS1 RU9255ID01 RU2104AWS1 RU2116AWS1 RF1305AWS1 RF0048AWS1 RU7746AWS1 RU3007AWS1 RU7850AWS1 RU0709AWS1 RU1035AWS1 RU2096AWS1 RF3976AWS1 RU5484AWS1 RU7813AWS1 RU6340AWS1 RU9368AWS1 RU6340AWS2 RU3923AWS1 RF2624AWS1 RU1203AWS1 RU0473AWS1 RU0472AWS1 RU7346AWS1 RU6265AWS1 RU2580AWS1 RF3270AWS1 RU7635AWS1 RU1733AWS1 RU1038AWS1 RU2527AWS1 RU2652AWS1 RF1477AWS1 RF1477AWS3 RF1477AWS2 RF0001AW08 RU0003ID01 RU1127AWS1 RU7524AWS1 RU1579AWS1 RF3031AWS1 RU1558AWS1 RU1689AWS2 RF1647AWS1 RU1041AWS1 RU8065AWS1 RU1011AWS1 RU7601AWS1 RU1086AWS1 RU1087AWS1 RU1717ID02 RU1112ID01 RU3052AWS1 RU1089AWS1 RU9333ID04 RU6547AWS1 RU1211AWS1 RF0832AWS1 RF0827AWS1 RU1198AWS1 RU9227AWS1 RU1479AWS1 RU3259AWS1 RU1095AWS1 RU9544AWS1 RF0931AWS1 RF2557ID01 RU1857AWS1 RU7021AWS1 RU1117AWS1 RU1591AWS1 RU1510AWS1 RU1091AWS1 RU1315AWS1 RU3143AWS1 RU7693AWS1 RF3256AWS1 RU1922AWS1 RU8069AWS1 RU1272AWS1 RU7703AWS1 RU6424AWS1 RF1316AWS1 RU3045AWS1 RU2428AWS1 RU8108AWS1 RU8902AWS1 RU1834AWS1 RU9190AWS1 RU7175AWS1 RU7956AWS1 RU8515AWS1 RU1424AWS1 RU7927AWS2 RU1109AWS1 RU1470AWS1 RU4075AWS1 RU6547AWS2 RU1435AWS1 RU7893AWS1 RU7986AWS1 RU1134AWS1 RU4089AWS1 RF1371AWS1 RU7274AWS1 RU3398AWS1 RU1516AWS1 RU1755AWS1 RF4234AWS1 600130AWS2 600130AWS1 RU1094AWS1 RU9000AWS1 RU4234AWS1 RU1309AWS1 RU0870AWS1 RU9594AWS1 25CE6E9D-AW-AWS_PRIMARY RURU6844AWS1 RU7973AWS1 RF3659AWS1 RO3052AWS2 RU8321AWS1 RU1128AWS1 RF1353AWS1 RU0731AWS1 RU3639AWS1 RF2791AWS1 RF1085AWS1 RU4440AWS2 RU4440AWS1 RU1685AWS1 RU2532AWS1 RU6235AWS1 RU3896AWS1 RF3595AWS1 RU8792AWS1 RU9450AWS1 RU1045AWS1 RU7316AWS1 RU3424AWS1 RU3462AWS1 RF0209AWS1 RF0908AWS1 RF1883AWS1 RU7837AWS1 RF0001AWS1 RU1693AWS1 RU1544AWS1 RU2641AWS1 RU5552AWS1 RF1293AWS1 RU8615AWS1 RU1026AWS1 RU1026AWS2 RU2526AWS1 RU9940AWS1 RU7312AWS1 RU4566AWS1 RU2275AWS2 RU9155AWS1 RU2741AWS1 RU1550AWS1 RU1550AW02 RU1012AWS1 RU3464AWS2 RU8243AWS1 RU1139AWS1 RU1377AWS1 RU6486AWS1 RU1692AWS1 RU1138AWS1 RF1383AWS1 RU1143AWS1 RF0554AWS1 RU2169AWS1 RF2412CT01 RU2353AWS1 RU1138AWS2 RU1449AWS1 RU1142AWS1 RU1052AWS1 RU1143AWS2 RU2250AWS1 RU9617AWS1 RU1934AWS1 RU1473AWS1 RU2438AW02 RU2679AWS3 RU9369AWS1 RU3427AWS1 RU5455AWS1 RU8356AWS1 RU1173AWS1 RU5952AWS1 RU0968AWS1 RU9595AWS1 600105AWS1 RU5495AWS2 RU5495AWS1 RU7827AWS1 RU9023AWS1 RU1718AWS1 RU3451AWS1 RU3469AWS1 RU3467AWS1 RU3951AWS1 RU3829AWS1 RF0918AWS1 RU1999AWS1 RU1999AWS2 RU7468AWS1 RF0449AWS1 RF0097AWS1 RU1217AWS1 RU1217AWS2 RF1157AWS1 RU4968AWS1 A5940829 SA2706PP05 SA2128 SA1060AW05 SA1216AW03 SA1232PP09 SA2079AW02 SA2921AW02 SA2468AW01 SA2051AW03 SA2102AW01 SA1041AW05 SA1220AW10 SA1037AW05 SA1114AW11 SA1089AW01 SA1304PP03 SA1035AW05 SA2896AW03 SA1036AW09 SA1036AW06 SA2315PP04 SA1223AW03 SA2814AW02 SA3053AW01 SA1013AW20 SA2581AW02 SA2315AW01 SA1013PP12 SA1004AW07 SA1005 SA2097AW02 SA2236AW02 SA1208PP05 SA1006 SA1176PP07 SA1255AW02 SA2424AW02 SA2191PP02 SA1011AW32 SA5000AW09 SA2281PP04 SA2972AW01 SA1209AW03 SA1176 SA1025 SA2329 SA1277 06B89DE0-AW-AWS_PRIMARY SA1054AW02 SA2191 SA1054_YAM_AWS SA1054_PMAHAWS SA1054_MAJAWS SA1054_Duw SA1054_AW SA1054_ARTAWS SA1058 SA2958_AWS SA1233_AW SA1232_AWS SA1232AW10 AWSP SA1011_AWS SA2253PP05 SA2280AW01 SA1001AW14 SA1058PP26 SA1008PP06 SA1058AW37 SA2006PP05 SA1279AW03 SA1175AW02 SA1175PP04 SA2060AW02 SA2268PP01 RS4020AWS1 RS4226AWS1 RS4001AWS1 RS4001AWS2 RS4014AWS1 RS4048AWS1 RS4185AWS1 RS4060AWS1 RS4029AWS1 AW55222592 RS4199AWS1 RS4034AWS1 RS4031AWS1 RS4038AWS1 RS4039AWS1 RS4035AWS1 RS4199AWS5 RS4199AWS4 RS4199AWS3 RS4199AWS2 RS4124AWS1 RS4012AWS2 RS4012AWS1 RS4156AWS1 RS4156AWS3 RS4025AWS1 RS4182AWS1 RS4095AWS2 RS4095AWS1 RS4095AWS3 RS4017AWS1 RS4037AWS1 RS4024AWS1 RS4028AWS1 RS4028AWS2 AW54013922 RS4036AWS1 RS4032AWS1 RS4032AWS2 6073IT0001 0847550033 6073AW0043MR1 6073AW0054MR1 6073AW0097CT3 6073AW0098CT3 6073AW0038CT2 06A14355-AW AMAWS 0847550033_06ABCD04_AWS05 0847550033_06ABCD02_AWS2 06ABCD01-AW-AWS_PRIMARY 06A0DB4F-AW-AWS_NODE 06a099cd-AW-AWS_PRIMARY 06A08792-AW-AWS_NODE 06A0865C-AW-AWS_NODE 06A04EDF-AW-AWS_PRIMARY SGAWX26086968 6073AW0004CT2 6073AW0003MR1 0847490039 0847550004 6073AW0031CT3 0847550024 0847550006 0847550005 SK1113AW01 SK1056AW12 SK1109AW01 SK1140AW02 SK1146AW01 SK1064AW02 12628432 SK1057AW01 SK1121AW01 SK1067AW03 SI4026AW02 SI4120AW03 12370147 ZA2169AW07 10914192 ZA2241AW02 ZA2241AW03 12114537 ZA3353AW01 11960505 ZA1940AW02 ZA2895PP04 12210473 ZA2266AW03 ZA1046AW01 ZAJRPAW01 ZA2241AWTST 06845945 0692F753-AW-AWS_PRIMARY 11960888 YP158331 YE151134 YE162A79 YE162A82 YE364421 DPN85206 DP461336 DEL899103 DEL899121 DEL899119 DPN36300 NE465144 DEN272143 DEN272142 DEN272141 DEN272140 DEN272173 NPD29700 NPV85802 FEC41525 FEC41526 FEC41527 FEC41528 NP666105 NE465125 NPV60203 FEE12458 NPV39111 FEG60958 11903791 GEX00704 GEX00708 CE677645 ME6459101 ME219521 UE282698 UE2826135 MP126219 MPX94403 MPX35217 MPX431116 ME645996 ME645991 MPX70914 MPD30301 MEX71306 ME2083108 FPG86462 NPV73519 UP123616 UP925716 DPN29027 YE595540 FE416094 MP713118 MP713128 NP473634 NP473627 NE212438 NPV91413 NE212493 NE203146 NE203174 YE643412 FEG88321 DPN00035 FEB520225 FEB520228 DEN00850 ESDATVAL-AW3 NE740526 FEG48326 NE209993 NE209992 NE2099102 NE2099245 NE216241 NED00002 NE216224 NE670347 MPX33116 MPX98013 YP315030 DEL26456 DEL264116 NEV44710 DEN06347 FPD24004 CP223917 CP749714 CE270958 MEX71007 MP384009 MEX71119 13249415 FEC800285 FED55A16 FEC800310 DE641231 DPN49406 MEX70610 ME5043186 MEX70616 FE387836 EEM88720 EEG32537 MEX70113 MEX67451 FEG32779 FE421186 FPG41420 GEE34065 GEE340100 069183BA_AWS3 069103C9_AWS4 06917CC8_AWSCAR GPX17305 GPX17304 GPQ19340 GEE62259 GEC21872 CE561279 ESDATVAL-AW4 FED03937 MP385520 MP878912 MEX70508 NEV44805 UE929528 UPW07916 UE929538 UPW07918 FPD10803 YE600103 YE600185 DPN87724 DP226717 DPN17105 DED647132 DPD35500 DPD35501 DED00117 DP418884 DPN26333 DE492A59 DE492A79 DPD14207 DE5392105 DPN17107 DE2032105 DE6588190 DPN89605 DE539285 DED647139 DED647167 DED647162 DPD01100 DED647170 DED647134 DE203288 DEH135109 DEH135104 DE5284114 DE5284108 DE528481 11447010 ESAW24817618 12605232 12764181 12622331 15309315 FE4438102 FP721A22 FE4438120 FEB33861 FE4438141 MEX25551 MEX25569 NPV79369 FPA99430 FPG39207 FEG45946 MP284820 FEG51856 FEG51855 FEG51854 FEG51848 FEG51847 UPW25205 YE646941 ME881026 DEK78892 DPN36505 NPV60101 NEV27524 NEV27538 HEAWS NE723610 NE723617 NEV07032 FEG88311 CEZ01561 MPX04220 CEG94658 MEX61755 MEX61795 MEX61794 MEX689446 MEX689445 MEX689419 MEX689418 MEX689417 MEX689416 MEX689415 MEX689414 MEX689413 MEX689412 ME751803 0692958F-AW-AWS_PRIMARY 06920EA6-AW-AWS_PRIMARY 069293e6-AW-AWS_Primary DEN01130 YP308730 YE344738 YE344740 DPN886104 NE673154 NPD00401 GED64913 FEG35054 MPX18618 MPX18629 NE748619 ME0515101 ME0515115 ME0515124 NE673142 CE557052 CE557061 CPM80624 NEV07045 NEV13615 NEV13623 UE260531 MEX68623 MPX65412 NE300614 MP926422 GE448066 GE448092 UE535A129 UE743A03 UE535A178 UE535A168 YE645883 YED00101 NEV06843 FE457A137 FEB34283 FP167514 FE457A133 FE457A145 FEE08469 cc00wwaw020 cc00wwaw019 cc00wwaw018 cc00wwaw017 cc00wwaw016 cc00wwaw015 cc00wwaw014 cc00wwaw013 cc00wwaw012 cc00wwaw011 cc00wwaw010 cc00wwaw009 cc00wwaw008 cc00wwaw007 CC00WWAW006 cc00wwaw005 cc00wwaw004 cc00wwaw003 cc00wwaw002 cc00wwaw001 CEK53840 GPX22610 GPX45702 MPX69001 MPX69000 DE669660 DE669698 MEX71230 UEW00701 DEG27417 NE219648 NE304912 NE223250 NPV15980 NE219662 NPD08401 NE185074 NE185072 NE419145 NEV439241 NE419194 NE419180 NE419166 NEV439251 NEV439250 NE223280 NEV44791 vvs07aws12 vvs07aws09 vvs07aws08 vvs07aws07 vvd04aws02 VVD03AWS02 VVD02AWS02 vvd01aws02 vvd15aws02 vvd10aws02 vvs07aws11 vvs07aws10 vvd14aws02 vvd08aws02 vvd14oaws02 vvdmolaws02 vvd12aws02 vvd06aws02 vvd05aws02 CEE156130 ESDATVAL-AW2 ESDATVAL-AWS01 ESDATVAL-AW7 ESDATVAL-AW6 ESDATVAL-AW5 ESDATVAL-AWS02 ESDATVAL-AWS07 ESDATVAL-AWS06 ESDATVAL-AWS05 MP230737 YP065823 YE626954 YE6002127 NE741442 NE741453 NEV35843 NEV06815 v01aap-010642 v01aap-010641 v01aap-010640 v01aap-010639 v01aap-010638 v01aap-010637 v01aap-010636 v01aap-010635 v01aap-010634 v01aap-010633 v01aap-010632 v01aap-010631 v01aap-010630 v01aap-010629 v01aap-010628 v01aap-010627 15205354 NE688842 NE688831 CE671345 EP223317 EE536917 EPP04604 EEF51596 EE6728109 083049806708324 083049803187419 LKAWX26457137 083049805333620 L007AWS01 P003AWS1 P003AWS2 O051AWSE1 D007AWS1 5482805B O005SS02_AWE 5483570A_MICROAWS X001AWS1 O012AWS1 O005AWS1 O009AWS1 O005AWS01 O005AWS03 N001AWS2 N001AWS02 M007AWS01 M007SS01EDISON_AWCT 5446060C L001AWS01 S002SS01 O020AWS2 E001AWS1 BD003AWS1 M001AWS3 M001AWS2 M001AWS01 M001SS02EDISON_AWCT M004AWS01 Z001AWS1 5562890B R002AWS2 A042AWS1 A201SS02_AWE A201SS01_AWE 12597869 A020AWS1 A201AWS2 A201AWST A205AWS1 A232NOD5 A232NOD4 A232NOD3 A232NOD2 A232NOD1 A203AWS1 AWS_GON5361362 Y004AWS1 Y004AWS2 AC001AWS2 14738329 C011AWS3 C011AWS1 U001AWS2 G001AWS1 KSAGEVAWS001 SPZGEVAWS001 SPZGETAWS001 KSAGETAWS001 529170AW07 KSB-AWS-SEAMLESS 850010AW01 350010AW07 SWBRI1AWS 315258AW02 273354AW01 259130AW11 SWCAR1AWS 824158AW01 251C7662-AW-AWS_PRIMARY 262010AW06 475290SS03 475290SV02 475290SS01 475290SS04 068C9C69-AW-AWS_PRIMARY sphfrgeaws12 1234576 250620AW01 265050AW06 266150AW06 5523572C 121385SS01 426020AW09 260010AW38 315258AW01 168934AW01 12204644 608168AW04 146806AW01 680542AW02 140206AW01 584074AW01 584074SV02 847040AW01 249121AW02 121512AW01 121512SV02 725120AW01 334963AW01 779205AW01 25302EBE-AW-AWS_Primary 475430AW09 786090AW11 475430AW11 502080AW06 109547AWS01 690120AW04 225030AW05 225030AW04 529150AW07 529150SV05 529150SV06 S-000-4384 S-000-4383 S-000-4382 S-000-4381 S-000-4380 5519582B S-000-4385 475410AW01 730140SV03 107104AW01 109547SV03 SWVIS1AWS 222768AW01 560060AW03 115030AW05 673160AW01 477485AW02 900010AW37 541010AW08 221679AW01 850020AW17 850020AW18 0853551103 0853550986 0853450269AWS1 38474A69-AW-AWS 0853210412AWS1 0853270223AWS2 0853550912 0853551045 0853551053 0853270214AWS1 0853210405AWS1 0853450261AWS1 0853551xxx 0853550847 0853550848 0853550875 0853550785 0853550921 0853550893 0853550888 0853550900 0853550895 0853550961 0853550932 0853550987 0853550992 0853550991 0853551007 0853550199 0853550198 0853551073 0853550200 0853550201 0853550935 0853550981 0853550980 0853551024 0853551078 0853551047 0853551037 0853551048 0853551058 0853551077 0853551040 0853551090 0853551081 0853551131 0853450251AWS1 0853270209AWS1 0853550999 0853260271AWS1 0853550218 0853270220AWS1 0853210408AWS1 0853551063 0853450255AWS1 0853551066 0853210393AWS2 0853270232AWS1 0853450276AWS1 0853450275AWS1 0853210434AWS1 0853210430AWS1 0853450287AWS1 0853270236AWS1 0853210437AWS1 0853450291AWS1 0853450281AWS1 0853210448AWS2 0853450266AWS1 38512921-AW-AWS_PRIMARY 38512920-AW-AWS 0853550851 1573AW0100CT 0856210132 0856490132CTAW1 0856490169CTAW1 1573AW0077CT4 1573AW0079CT1 1573AW0080CT1 1573AW0093CT1 1573AW0126CT1 1573AW0051MR1 1573AW0110CT1 1573AW0113CT1 1573AW0042MR1 1573AW0053CT1 1573AW0073CT1 1573AW0142CT1 1573AW0130CT 1573AW0143CT1 1573AW0146CT1 1573AW0131CT1 0005466242A 0005437183A THAWX24476145 THAWX25246646 0856490157CTAW2 1573AW0041MR1 1573AW0070MR1 1573AW0066CT1 1573AW0133CT 1573AW0045MR2 1573AW0120CT1 1573AW0140CT1 1573AW0218CT1 THAWX25203423 1573AW0134CT 1573AW0075MR1 1573AW0004 1573AW0051CT1 1573AW0041CT1 1573AW0094CT1 0856490171CTAW1 1573AW0217CT1 1573AW0085CT1 0005476618A 0005406666A 0005476360A 0005467851A THAWX24513792 D932256C-AW-AWS_PRIMARY 1573AW0045CT1 1573AW0013MR2 1573AW0057CT1 1573AW0163CT1 1573AW0046CT1 TN1028AW06 TN1588AW01 TN1729AW03 TN1798AW01 07886305 07755113 TN2002AW02 TN1278AW01 TN1193AW01 TN1121AW03 TN1271AW01 TN1632AW01 TN1744AW04 TN1272AW01 07772629 TN1781AW01 TN1018AW08 TN2000AW03 TN1061AW02 TN1194AW01 TN1166AW06 TN1166AW05 TN1059AW07 11848503 10926501 10915147 10926500 06002DCS16 06002PAC36 06172DCS03 06218DCS01 06236DCS01 06162PAC12 06A93D70-AW-AWS_PRIMARY 06540AW01 06051PAC30 08124655 12068133 09943699 08957093 10027DCS01 09820451 16018DCS02 09947315 12133760 26021DCS01 31020DCS03 34393DCS05 34225PAC10 34296PAC02 34394PAC16 34001PAC28 34461DCS02 34717DCS01 34722DCS01 34716DCS01 34715DCS01 34328DCS05 34418DCS12 34OKAWS01 07652 08912539 12988510 13786524 14798244 13331647 35003DCS07 35097DCS01 08505256 38022DCS02 08864893 00023DCS01 08089974 13663131 53008DCS01 15649246 55001DCS02 55029DCS02 13140401 06051DCS24 23002DCS06 41400PAC13 41400PAC09 34340AW01 34726DCS02 35064DCS02 34381PAC80 41009DCS05 AE1492AW01 SKMCAWS 12931121 AE1010AW01 AE1010AW04 AE1564PP06 AE1160PP07 AE1802PP01 AE1025AW07 CZ213800S6 PK1453AW02 AW49043341 AE1476AW02 AE1321AW02 13083935 00670AWS01 10127417 00145AWS01 01278AWS01 05262AWS01 00002AWS02 00002CTSSAW 00234AWS01 00617AWS01 IE1513AW03 81564AWS29 81564AWS27 11477AWS01 00487AWS02 00242AWS01 00094AWS02 CRHAWS02 CRHAWS01 DCCRHAWAP20 CRHAWS21 CRHAWS20 70001AWS01 00001AWS01 81001ACHAWS03 81001ACHAWS02 81001ACHAWS01 MFTTSTAWS01 EKHTSTAWS01 MTWTSTAWS01 MFTAWS01 MAIAWS1 EKHAWS02 EKHAWS01 06ba4613-AW-AW31 06B90083 00222AWS01 00243AWS02 00991AWS01 77055AWS01 00058AWS01 00361AWS01 00440AWS04 DDC01VM059 77257AWS01 09447008 80257VM113 80257VM112 RTGAWS06 RTGAWS05 RTGAWS04 RTGAWS03 RTGAWS02 RTGAWS01 RTGTSTAWS20 RJFAWS20 00468AWS01 00197AWS01 00052AWS01 00033AWS01 00298AWS01 00199AWS01 00006AWS01 RWDDAVM035 RWDDAVM033 00133AWS01 00379AWS01 00182AWS01 00463AWS01 5545631B 01344AWS01 00065AWS02 00243AWS03 00555AWS01 09446999 00203AWS01 00556AWS01 10692AWS01 01065AWS01 RNQ51VM022 RNQ51VM021 RNQ51VM010 KGHAWS06_AWS KGHAWS05_AWS KGHAWS04_AWS uhlaws20 KGHAWS22 KGHAWS21 KGHAWS20 DCKGHAWAP20_AWS 00039AWS01 00916AWS02 00165AWS01 10627AWS02 00615AWS01 RWEAAVM027 RWEAAVM026 RWEAAVM025 06a7e5a3 UHLTSTAWS20 UHLAWS22 UHLAWS2O UHLAWS21 77522AWS01 06A0100A-AW-AWS_PRIMARY ULHAWS07 ULHAWS06 ULHAWS05 ULHAWS22 ULHAWS21 ULHAWS20 dculhawap20 dcuhlawap20 DCRTGAWAP20 10307AWS01 01305AWS01 10344AWS01 00044AWS01 11247AWS01 00010AWS01 00060AWS01 77588AWS01 00023AWS01 00026AWS01 11187AWS01 10050AWS01 09447753 GOSH-GE20 09526317 09429950 513636SMART2 00314AWS01 00314AWS02 00450AWS01 10537AWS01 00136AWS01 00584AWS01 00139AWS01 80257VM039 RNS01VM019 RNS01VM018 NGHAWS06 NGHAWS05 NGHAWS04 NGHAWS22 NGHAWS21 NGHAWS20 DCNGHAWAP20 RNS01VM010 UHLAWS01 06A44F88-AW-AWS_PRIMARY NUHAWS03 NUHAWS02 NUHAWS01 DCTRNAWAP20 NUHAWS22 NUHAWS21 NUHAWS20 DCNUHAWAP20 DCTSTAWAP21 DCTSTAWAP20 IE1208AW01 10448AWS02 10448AWS03 00328AWS01 00216AWS02 00100AWS01 10493315 00100CTSS1AW 20297AWS01 21012AWS01 00131AWS01 00224AWS01 10127402 CWSAWS01 00318AWS02 00020AWS01 00824AWS02 85252VM022 05252AWS02 DDC01VM042 DDC01VM041 00094AWS01 RK5BCVM023 RK5BCVM022 RK5BCVM010 00260AWS01 DCSFHAWAP01 SFHAWS03 SFHAWS02 SFHAWS01 SFHAWS22 SFHAWS21 SFHAWS20 sfhaws04 DCSFHAWAP20 5406760B 00223AWS03 10537CTS01 00192AWS01 01119AWS01 00424RBKAWS03 00424RBKAWS02 00424RBKAWS01 rbktstaws rbkaws04 rbkaws03 10198980 00082AWS01 00440AWS03 00440AWS02 00440DW16 00084AWS02 00244AWS01 RRF01-AWServer-14767611 00488AWS01 12370134 00357AWS01 HCIT4825232AWS HCIT4939529AWS HCIT5303301AWS HCIT5303296AWS HCIT5153914AWS UZ1007AW02 HCIT4271649AWS HCIT4741747AWS HCIT4409614AWS CT444241HM4AW 5117944AW41 To be provided To be provided 2565C84B-AW-AWS_PRIMARY 5973IT0005 068811CE-AW-AWS_PRIMARY 5973AW0389CT1 5973AW0292CT1 9830AW444CT1 5973IT0002 25301BD9-AW-AWS_PRIMARY 06A1783B-AW-AWS_PRIMARY 25120DD1-AW-AWS 255167AC 25C09F68-AW-AWS_PRIMARY 25EEEC94-AW-AWS_PRIMARY 25DB5E0A-AW-AWS_NODE 25A7D57E-AW-AWS_NODE 25A78C49-AW-AWS_PRIMARY 257B8528-AW-AWS_NODE 25333227-AW-AWS_NODE 252E7472-AW-AWS_NODE 250781F7-AW-AWS_NODE VMTCAW01 VMCT_AWS01 VinMecSmartCityAW01 PRECISIONMEDAWS2 PRECISIONMEDAWS1 VNAW5320632A (Updated 3/11/2026) Additional System IDs 708425CTRVEHLAWS 718226APEXAWS WPHTMP 216844SMARTSUBAWS1 210575AW2 AUAWX26631168 AW52743961 HCIT5574699AWS HCIT5476277AWS HCIT5317135AWS CBDMG2300090AWS 604682AWS2 5454275AW41 5454275AW61 CBDWG2400077HMAWS 82427030241 15585407AWS 9429976 15169963 15391391 07577111 14829089 14830987 14696567 15533573 15706077 14829090 15561104 15658050 15525972 14759893 83049002220925 INAWX26381846 INAWX26563044 INAWX26381882 INAWX26381851 INAWX26596378 11960519 12094344 tpacsaw01 EW0392 SH0022 15432088 0850450224AS1 3BD9E8F8-AW-AWS_PRIMARY LV4001AW10 BMCAWSN AW54169842I AW55624592 AW55686451 MA1864AWS01 NZAWX25209812 14497283 14795021 RU4929AWS1 RUAWX26411049 RU5494AWS1 6073AW0020CT1 SGAWX25083035 L001AWS02 12597870 THAWX24852951 THAWX24988201 13745670 13748593 15875772 ANKAWS HUAWS AWS55342661 13352728 13360177 HCIT5303285AWS HCIT5393306AWS
FEI Number 1000284313
Recalling Firm/
Manufacturer
GE Medical Systems, SCS
283 rue de la Miniere
Buc France
For Additional Information ContactGE Healthcare Service
800-4371171
Manufacturer Reason
for Recall
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data.
FDA Determined
Cause 2
Software in the Use Environment
ActionOn September 5, 2025, firm began deployment of customer letter. Customers were informed they could continue use of their device by following the instructions provided: 1. Ensure the AW Server is installed on a facility internal network, that is not accessible from any public network. If access to AW Server from outside the facility network is desired, this connectivity must be configured to be usable exclusively through the facility s Virtual Private Network (VPN) or equivalent network security solution as described in your AW Server s Pre-Installation Manual that is available in the Customer documentation portal. NOTE: Access management to the facility s internal network on which the AW Server is installed on, and management of the network security solution is under the strict responsibility of the facility. 2. Ensure strong, secure passwords are configured for the AW Server. 3. GE HealthCare recommends that organizations follow security and cybersecurity best practices, including securing the network where AW Server is connected.
Quantity in Commerce7,302 (1,381 US; 5,921 OUS)
DistributionDomestic US distribution nationwide. International distribution worldwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LLZ
-
-