| | Class 2 Device Recall uCT 550 |  |
| Date Initiated by Firm | September 12, 2025 |
| Date Posted | October 21, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0218-2026 |
| Recall Event ID |
97668 |
| 510(K)Number | K181414 K200016 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | Shanghai United Imaging Healthcare Co., Ltd.
Computed Tomography X-ray System,
Model:uCT 550
MD, Rx Only |
| Code Information |
UDI:(01)06971576831036/
Serial Numbers: 353020 353021 353023 353045 353049 354001 354002 354003 354004 354005 354006 354007 354008 354009 354010 354011 354012 354013 354014 354015 354016 354017
|
Recalling Firm/ Manufacturer |
UIH Technologies LLC 9380 Kirby Dr Houston TX 77054-2515
|
| For Additional Information Contact | Customer Service 832-669-2799 |
Manufacturer Reason for Recall | Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. |
FDA Determined Cause 2 | Other |
| Action | On September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions:
1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts.
2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers.
3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799
4.Report adverse events: if any adverse events or quality issue occur, please report them to the FDA via the MedWatch program.
5. Please share this information with all relevant personnel at your facility, including technologist, service staff, and clinical teams. |
| Quantity in Commerce | 22 units |
| Distribution | US:
OUS: |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|