• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall American Contract Systems

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall American Contract Systemssee related information
Date Initiated by FirmSeptember 04, 2025
Date PostedOctober 16, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0177-2026
Recall Event ID 97659
Product Classification Central venous catheter tray - Product Code OFF
ProductMedical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION SUPPLY KIT UICL44R ADULT CENTRAL LINE INSERTION SUPPLY KIT
Code Information UICL44P UDI-DI 191072194268 Lot 8003911 Bag serial number N/A; UICL44R UDI-DI 191072234018 Lot 8317711 Bag serial number 68244081 * If bag serial number is listed as "n/a", then entire lot is affected
FEI Number 3007069947
Recalling Firm/
Manufacturer
American Contract Systems Inc.
1601 Se Gateway Dr Ste 120
Grimes IA 50111-5223
For Additional Information ContactMarlene Jones
470-280-4277
Manufacturer Reason
for Recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionFirm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Quantity in Commerce32,433 total
DistributionUS Nationwide distribution in the states of SD, IA, MN, WA, IL.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-