| | Class 2 Device Recall 1077 DPOAE/TEOAE PROBE |  |
| Date Initiated by Firm | November 06, 2025 |
| Date Posted | December 18, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0940-2026 |
| Recall Event ID |
97894 |
| 510(K)Number | K100661 |
| Product Classification |
Stimulator, auditory, evoked response - Product Code GWJ
|
| Product | MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) |
| Code Information |
Part Number: 8-69-41100;
UDI-DI: 04260223141355;
Serial Numbers: 1008852, 1008888, 1008916, 1008925,
1008935, 1008945, 1008951, 1008957,
1008960, 1008962, 1008966, 1008967,
1008987, 1008990, 1009054, 1009065,
1009066, 1009131, 1009137, 1009138,
1009139, 1009145, 1009148, 1009155,
1009159, 1009161, 1009169, 1009184,
1009188, 1009191, 1009305, 1009315,
1009316, 1009319, 1009321, 1009322,
1009325, 1009328, 1009343, 1009352,
1009356, 1009357, 1009359, 1009366,
1009372, 1009375, 1009376, 1009386,
1009388, 1009390, 1009396, 1009399,
1009400, 1009404, 1009411, 1009415,
1009419, 1009426, 1009427, 1009432,
1009439, 1009442, 1009454, 1009456,
1009460, 1009463, 1009464, 1009468,
1009475, 1009481, 1009488, 1009494,
1009495, 1009497, 1009848, 1009855. |
| FEI Number |
3008111228
|
Recalling Firm/ Manufacturer |
PATH Landsberger Str. 65 Germering Germany
|
| For Additional Information Contact | Dr. Johann Oswald +498980076502 |
Manufacturer Reason for Recall | Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome. |
FDA Determined Cause 2 | Component change control |
| Action | On November 6, 2025 URGENT: MEDICAL DEVICE Field Corrective Action letters were sent to their two distributors. Actions to be taken: - Users will be asked to confirm the DPOAE protocol in use.
- Verify the DPOAE Protocol used is not affected by the issue (DP1, DP2,
DP3, DP4, DP17, DP18, DP19, DP20, DP21).
- If any other protocol is used, please replace it by using AccuLink
Software. Follow instructions in IFU to do so.
- Remove all affected protocols (DP5-DP16) using AccuLink Software.
Follow instructions in IFU to do so.
- Complete and return the attached form to confirm these steps were
followed.
- The expected lifetime of the probes is two years from date of
manufacture. Please put the products out of use, once this is reached. |
| Distribution | Domestic: WI.
International: Ireland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = GWJ
|
|
|
|