| Date Initiated by Firm | November 26, 2025 |
| Date Posted | January 12, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1053-2026 |
| Recall Event ID |
98154 |
| Product Classification |
Sewing machine - Product Code RDE
|
| Product | Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as:
RR-LAP DONOR NEP ACCESSORY, Medline kit SKU DYNJ63395B |
| Code Information |
UDI/DI 10195327634698 (EA) 40195327634699 (CS), Lot numbers: 25HLA435, 25HLB046.
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product.
For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 33 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|