| | Class 2 Device Recall Custom Waste Management Kit Vascular Tray |  |
| Date Initiated by Firm | December 04, 2025 |
| Date Posted | January 16, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1125-2026 |
| Recall Event ID |
98139 |
| Product Classification |
Universal drainage tray - Product Code PPD
|
| Product | Merit Medical , Custom Waste Management Kit Vascular Tray
REF: K10T-05261 |
| Code Information |
UDI: 00884450320297/ Lot: T2736369 T2781930 T2898771S T3047503 T3090431 T3121937 T3146114 T3158708 T3204129 T3232415
|
Recalling Firm/ Manufacturer |
Merit Medical Systems, Inc. 1600 W Merit Pkwy South Jordan UT 84095-2416
|
| For Additional Information Contact | Merit Customer Service 1800-356-3748 |
Manufacturer Reason for Recall | Inflation device handle may detach from the syringe during procedure. |
FDA Determined Cause 2 | Process control |
| Action | On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions:
1. Please read and understand the recall notification.
2. Should the handle detach during use and the balloon not completely deflate, pull the
handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast
3. Ensure that applicable personnel within your organization are made aware of this field action.
3. If the product has been further distributed to other facilities, institutions, or manufacturers, please
ensure this notice is immediately shared with them.
4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within seven (7) calendar days |
| Quantity in Commerce | 864 units |
| Distribution | Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Korea, Republic of Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Madagascar, Mexico, Monaco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Trinidad and Tobago, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian, Republic of Viet Nam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|