| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | December 12, 2025 |
| Date Posted | January 28, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1197-2026 |
| Recall Event ID |
98226 |
| Product Classification |
Gynecological laparoscopic kit - Product Code OHD
|
| Product | Medline medical procedure convenience kits labeled as:
1) RR-GYN LAPAROSCOPY PACK-LF, Kit SKU DYNJ0395022O;
2) PELVISCOPY PACK-LF, Kit SKU DYNJ0415776Q;
3) DIVA PACK, Kit SKU DYNJ04811O;
4) DA VINCI PROSTATE/COLPOPEXY, Kit SKU DYNJ44864S;
5) MH ROBOTICS PACK, Kit SKU DYNJ50880L;
6) DELNOR LAPAROSCOPIC GYN PACK, Kit SKU DYNJ82244B;
7) ROBOT PACK, Kit SKU DYNJ82761B;
8) GENERAL ROBOTICS PACK, Kit SKU DYNJT4964. |
| Code Information |
Medline Kit SKU DYNJ0395022O: UDI/DI 10195327634827 (EA) 40195327634828 (CS), Lot Number 25KME682;
Medline Kit SKU DYNJ0415776Q: UDI/DI 10198459566806 (EA) 40198459566807 (CS), Lot Number 25KMH125;
Medline Kit SKU DYNJ04811O: UDI/DI 10198459440953 (EA) 40198459440954 (CS), Lot Number 25KMD772;
Medline Kit SKU DYNJ44864S: UDI/DI 10195327127626 (EA) 40195327127627 (CS), Lot Number 25LMA115;
Medline Kit SKU DYNJ50880L: UDI/DI 10198459388002 (EA) 40198459388003 (CS), Lot Number 25LMA516;
Medline Kit SKU DYNJ82244B: UDI/DI 10198459344053 (EA) 40198459344054 (CS), Lot Number 25KMF328;
Medline Kit SKU DYNJ82761B: UDI/DI 10198459039232 (EA) 40198459039233 (CS), Lot Number 25LMA391;
Medline Kit SKU DYNJT4964: UDI/DI 10198459460098 (EA) 40198459460099 (CS), Lot Number 25KMJ341. |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following:
"REQUIRED ACTION:
1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.
2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form.
Website link: https://recalls.medline.com
Recall Reference #: R-25-251-FGX2
Recall Code:
3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.
4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product."
For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 1,928 total |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|