Class 1 Device Recall ACUSON AcuNav Ultrasound Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall ACUSON AcuNav Ultrasound Cathetersee related information
Date Initiated by FirmDecember 22, 2025
Date PostedFebruary 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1325-2026
Recall Event ID 98277
510(K)NumberK234064 
Product Classification Reprocessed intravascular ultrasound catheter - Product Code OWQ
ProductReprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH
Code Information 10135910RH UDI-DI 10197344026500 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905; 10135936RH UDI-DI 10197344026517 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
FDA Determined
Cause 2
Under Investigation by firm
ActionOn December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
Quantity in Commerce511
DistributionDistribution US nationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = OWQ
-
-