| Date Initiated by Firm | January 08, 2026 |
| Date Posted | February 24, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1468-2026 |
| Recall Event ID |
98333 |
| Product Classification |
Dressing change tray - Product Code OXQ
|
| Product | Convenience kits containing Tego Connectors
Medline DIALYSIS DRSNG CHNG
SKU DYNDC3167 |
| Code Information |
UDI-DI (ea) 10195327015954
UDI-DI (case) 40195327015955
Kit lots
24IME953
25CMA708
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 800-633-5463 |
Manufacturer Reason for Recall | Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body. |
FDA Determined Cause 2 | Process control |
| Action | On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail.
Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned. |
| Quantity in Commerce | 150 |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|