• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MEDLINE CONVENIENCE KITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MEDLINE CONVENIENCE KITSsee related information
Date Initiated by FirmFebruary 27, 2026
Date PostedApril 17, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1885-2026
Recall Event ID 98601
Product Classification Gynecological laparoscopic kit - Product Code OHD
ProductMedline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes GU DAVINCI PACK - LF DYNJ909398A ROBOTIC PACK DYNJ908706D
Code Information DYNJ909398A UDI-DI 10195327338190 (EA) 40195327338191 (CS) LOTS 24GMH510 24HMG886 DYNJ908706D UDI-DI 10198459017773 (EA) 40198459017774 (CS) LOTS 24HMA783 24HMG308 24KMC651 24KMJ274 24LME931 25AME217 25BMI653
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-633-5463
Manufacturer Reason
for Recall
Unapproved design changes to the products outside of the 510(k) clearance.
FDA Determined
Cause 2
Component change control
ActionOn February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Quantity in Commerce270,311 total
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-