Class 2 Device Recall Inspire Medical Systems, Inc.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Inspire Medical Systems, Inc.see related information
Date Initiated by FirmMay 11, 2026
Date PostedJune 18, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2515-2026
Recall Event ID 98990
PMA NumberP130008 
Product Classification Stimulator, hypoglossal nerve, implanted, apnea - Product Code MNQ
ProductBrand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Code Information Lot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562 Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217
FEI Number 3007666314
Recalling Firm/
Manufacturer
Inspire Medical Systems Inc.
5500 Wayzata Blvd Ste 1600
Golden Valley MN 55416-1237
For Additional Information Contact
763-205-7974
Manufacturer Reason
for Recall
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
FDA Determined
Cause 2
Process control
ActionFirm began notifying consignees on May 11, 2026. Letters were titled "URGENT Medical Device Correction." The firm requests consignees' assistance in removing affected product from clinical inventory and returning them to Inspire.
Quantity in Commerce50 impacted units
DistributionWorldwide distribution - US Nationwide and the country of Hong Kong.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.