| | Class 2 Device Recall MEDLINE |  |
| Date Initiated by Firm | May 04, 2026 |
| Date Posted | June 17, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2498-2026 |
| Recall Event ID |
98994 |
| Product Classification |
Wound dressing kit - Product Code MCY
|
| Product | MEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ, Medline kit number/SKU DYNJ68808 |
| Code Information |
UDI/DI 10193489843187 (each), 40193489843188 (case), Lot Numbers: 21LBT965, 21IBU081, 21GBN159, 21DBC976.
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers.
For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 91 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|