 Class 2 Device Recall Medline Convenience Kitssee related information
Date Initiated by FirmMay 21, 2026
Date PostedJune 29, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2543-2026
Recall Event ID 99162
Product Classification I.V. start kit - Product Code LRS
ProductMedline Convenience Kits: 1) ARTERIAL LINE KIT, Model Number: ART890B; 2) ULTRASOUND GUIDED IV SECUREMENT SET, Model Number: IVS3350
Code Information 1) ART890B, UDI-DI: 10653160360834(each), 00653160360837(case), Lot Number: 2025102390; 2) IVS3350, UDI-DI: 10653160313045(each), 00653160313048(case), Lot Number: 2025102490
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 5/21/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-082-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce250 kits
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report