| Date Initiated by Firm | July 01, 2026 |
| Date Posted | August 05, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2896-2026 |
| Recall Event ID |
99344 |
| Product Classification |
Plasma, coagulation control - Product Code GGN
|
| Product | 5-PLASMA Bivalirudin Control, Model Number 5D-44905 |
| Code Information |
UDI-DI: 05070002420385,
Lot Number: KE132602 |
| FEI Number |
3005017664
|
Recalling Firm/ Manufacturer |
Aniara Diagnostica LLC 7768 Service Center Dr West Chester OH 45069-2442
|
| For Additional Information Contact | Michele Washing 513-505-5276 |
Manufacturer Reason for Recall | Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values. |
FDA Determined Cause 2 | Process control |
| Action | Email notifications were sent to customers on 6/17/26.
Please reply to this email as acknowledgment you have received this initial notification and please include the quantity of each product which is remaining in your possession.
A notification letter dated 7/1/26 was sent to customers.
Actions to be implemented:
" Destroy all stock of 5-PLASMA Bivalirudin Control, 5-PLASMA Bivalirudin Calibrator, & 5-PLASMA AMR Verification Kits
- Please permanently deface the box labels and vial labels before disposal.
" Complete and return the Awareness Acknowledgement form.
" If product has been consumed please add brief statement to the Comment section, complete the End User Information, and return the Awareness Acknowledgement form. |
| Quantity in Commerce | 9 units |
| Distribution | US Nationwide distribution in the states of IL, NC, TN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|