• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Intramedullary Bone Saw

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Intramedullary Bone Sawsee related information
Date Initiated by FirmJuly 01, 2026
Date PostedAugust 03, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2869-2026
Recall Event ID 99346
Product Classification Orthopedic manual surgical instrument - Product Code LXH
ProductIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary Bone Saw Blade Assembly 14mm
Code Information UDI-DI (01)00887868507816(17)360221(10)67818764; Lot 67818764 Model No 475630; UDI-DI (01)00887868507816(17)360401(10)67818768; Lot 67818768 Model No 475630; UDI-DI (01)00887868507816(17)360428(10)67954534; Lot 67954534
FEI Number 1825034
Recalling Firm/
Manufacturer
Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact411 Technical Services
800-348-2759
Manufacturer Reason
for Recall
Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 6/22/26 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with this recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Calls received outside of call center operating hours will receive a vo
Quantity in Commerce26
DistributionRefer to the attached ZFA 2026-00096_Initial Consignee List and ZFA 2026-00096_Initial Distribution History containing the distribution and the identification of the affected consignees.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-