| | Class 2 Device Recall VACUETTE LH Lithium Heparin Separator Tubes |  |
| Date Initiated by Firm | July 29, 2026 |
| Date Posted | October 21, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0200-2027 |
| Recall Event ID |
99518 |
| 510(K)Number | K960857 |
| Product Classification |
Tubes, vials, systems, serum separators, blood collection - Product Code JKA
|
| Product | Brand Name: VACUETTE
Product Name: LH Lithium Heparin Separator
Model/Catalog Number: 454008P
Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml) |
| Code Information |
Lot Code: Item Number: 454008P
Lot Number: B26043FM
UDI: Rack: 29120017579032, Case: 39120017579039
Expiration Date: 2027-07-12
|
| FEI Number |
3001451379
|
Recalling Firm/ Manufacturer |
Greiner Bio-One North America, Inc.
4238 Capital Dr
Monroe
NC
28110-7681
|
| For Additional Information Contact | Angela Drysdale 001-704-7768646 |
Manufacturer Reason for Recall | Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On 07/29/2026, an "URGENT MEDICAL DEVICE RECALL NOTICE" was emailed to customers informing them that, Greiner Bio-One North America identified a potential ethylenediaminetetraacetic acid (EDTA) contamination issue in a specific lot of VACUETTE Lithium Heparin serum separator tubes. Use of affected tubes may lead to:
" Falsely elevated potassium (K+) results
" Falsely decreased calcium (Ca++) results
Customers are instructed to:
" Immediately identify and quarantine all remaining units of item 454008P, lot B26043FM.
" Stop using the affected lot for blood collection.
" Notify relevant staff (phlebotomy, lab, clinical teams) of this advisory.
" Review any recent patient results obtained using this lot. Consider repeat testing with a
different lot if results appear inconsistent with clinical presentation.
" Follow proper order of draw, and recollection procedures when re-testing.
For unused product disposal (if destroyed by customer/distributor)
" The unused blood collection tubes contain chemical additives (LH and serum
separator). Recommend following OSHA process, and/or your facility's environmental
health and safety (EHS) policies and procedures for disposal.
For questions or documentation (Acknowledgement and Receipt Form) to
Recalls.US@gbo.com or please call 800-515-8112. |
| Quantity in Commerce | 860450 |
| Distribution | U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|