| | Class 2 Device Recall HomeFill Oxygen System |  |
| Date Initiated by Firm | July 28, 2026 |
| Date Posted | September 18, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3221-2026 |
| Recall Event ID |
99660 |
| Product Classification |
Cylinder, compressed gas, and valve - Product Code ECX
|
| Product | HF2P0ST9
HomeFill Oxygen System
HomeFill Cylinder M9 with CGA870 Compatible Post Valve |
| Code Information |
UDI:00850018761369/Serial #s: 24IKMA00532 24IKMA0053M 24IKMA0054N 24IKMA005G8 24IKMA005AP 24IKMA005QB 24IKMA005Q4 24IKMA005PM 24IKMA005QJ 24IKMA005QX 24IKMA005PJ 24IKMA005SW 24IKMA005RB 24IKMA00612 24IKMA0067E 24IKMA0066L 24IKMA0068U 24IKMA00685 24IKMA006AQ 24IKMA006B7 24IKMA006BG
|
| FEI Number |
3013095415
|
Recalling Firm/ Manufacturer |
Ventec Life Systems, Inc.
22002 26th Ave Se
Bothell
WA
98021-4903
|
| For Additional Information Contact | George Koos 1-863-2266285 |
Manufacturer Reason for Recall | Labeling; Oxygen Cylinders do not contain required medical gas labeling |
FDA Determined Cause 2 | Process control |
| Action | On July 28, 2026, Ventec Life Systems Inc. issued a Urgent: Medical Device Recall notification to affected consignees via: certified mail. Ventec ask consignees to take the following actions:
1. Immediately discontinue use of affected HomeFill Oxygen Cylinders.
2. Inspect current stock, quarantine/segregate any unused affected products, remove affected product from clinical service and segregate to prevent further use until the enclosed label correction has been completed.
3. Use an unaffected oxygen cylinder for patient support.
4. If you have distributed or transferred affected product to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification.
5.. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health WITHIN 7 DAYS OF RECEIPT.
6. Upon receipt of the Customer Response Form, React Health will arrange for shipment of the label and instructions to apply the label to affected cylinders along with an update confirmation letter.
7. Update the affected product and return the completed confirmation letter to React Health. Contact Customer Support at 863-226 -6285 to discuss completing the form, if necessary. |
| Quantity in Commerce | 21 units |
| Distribution | US Nationwide distribution in the states of FL, IA, OH, NE, TN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|