| | Class 2 Device Recall Battle Bandage |  |
| Date Initiated by Firm | September 02, 2026 |
| Date Posted | October 02, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0081-2027 |
| Recall Event ID |
99814 |
| Product Classification |
Dressing, compression - Product Code MHW
|
| Product | Battle Bandage, REF: 32-700;
Patrol IFAK Bag, Black - Advanced, REF: 81-125-BL;
Patrol IFAK Bag, Multicam - Advanced, REF: 81-125-MC;
Patrol IFAK II, Olive Drab, Stocked (Option 1), 81-129-OD;
IFAK Bag, Intermediate, Black - Stocked, 81-133-BL;
IFAK Bag, Intermediate, Coyote Brown - Stocked, 81-133-CB;
IFAK Bag, Intermediate, Multicam - Stocked, 81-133-MC;
IFAK Bag, Advanced, Black - Stocked, 81-135-BL;
IFAK Bag, Advanced, Multicam - Stocked, 81-135-MC;
LEO Patrol Kit, 81-139;
IFAK Refill, 81-301;
IFAK Resupply, 81-303-01;
IFAK Resupply - CS/30, 81-303-30;
K9 Trauma Kit, 81-605;
Medic Enhancement Set, 82-504-03;
Medic Enhancement Resupply, 82-504-05;
Emergency Trauma Kit, 82-520;
First Responder - Intermediate, Coyote Brown, 83-303;
First Responder, Intermediate, Black, 83-303-BL;
DC-First Responder - Intermediate - Foliage Green, 83-303-FG;
First Responder, Intermediate, Multicam, 83-303-MC;
First Responder - Advanced- Coyote Brown, 83-305-CB;
Combat Life Saver - Intermediate - Coyote Brown, 83-323-CB;
Combat Life Saver - Intermediate - Multicam, 83-323-MC;
Multi Mission, Advanced, Black - Stocked, 84-446-01-BL;
Multi Mission, Advanced, Coyote Brown - Stocked, 84-446-01-CB;
Multi Mission, Advanced, Multicam - Stocked, 84-446-01-MC;
Direct Action Bag, Coyote Brown, Advanced, Stocked, 84-497-CB;
DC-DIRECT ACTION BAG, FOLIAGE GREEN, ADVANCED, STOCKED, 84-497-FG;
Direct Action Bag, Multicam, Advanced, Stocked, 84-497-MC;
Bleeding Control Kit - Advanced, Stocked, 87-155;
Stop the Bleed Kit (NJSP), 87-232;
Patrol Officer IFAK - Basic, 88-105-01;
Patrol Officer IFAK - Advanced, 88-105-03;
High Threat Kit - Basic, 88-108-01;
High Threat Kit - Intermediate, 88-108-02;
High Threat Kit - Advanced, 88-108-03;
High Threat Kit - PS, 88-108-05;
Micro Trauma Kit (without Tourniquet), Black, 88-147;
SDR EDC v1 - OTC, SDR-EDC1OTC;
DC-SDR EDC v1 - OTC, SDR-EDC1-OTC;
SDR Tactical Battle Belt IFAK - OTC - Black, SDR-TBBOTC;
DC-SDR Tactical Battle Belt IFAK - OTC, SDR-TBB-OTC;
SDR Tactical Battle Belt IFAK - OTC - Coyote Brown, SDR-TBBOTC-CB;
Emergency Trauma Kit (Employee only kit), SG0999/COM |
| Code Information |
REF/UDI-DI/Lot(Expiration):
32-700/M68932700070/PBO0005447(10/31/2029), PBO0006353(2/12/2030), PBO0004168(9/12/2029), PBO0006219(10/22/2029), PBO0005499(12/27/2028), PBO0005960(11/29/2029), PBO0004076(9/12/2029), PBO0004497(10/31/2029), PBO0004261(9/12/2029), PBO0004301(10/17/2029), PBO0005856(11/29/2029), PBO0005599(2/12/2030), PBO0006434(12/25/2029), PBO0009032(8/14/2031), PBO0008804(9/13/2030), PBO0010083(12/4/2031), PBO0010114(1/14/2032), PBO0007376(6/20/2030), PBO0005601(1/18/2030), PBO0007651(10/31/2029), PBO0007809(7/29/2030), PBO0008889(5/7/2030), PBO0010280(8/24/2031), PBO0010523(2/9/2032), PBO0010613(2/16/2032), PBO0007921(5/7/2030), PBO0009049(8/14/2031), PBO0008482(5/7/2030), PBO0009934(12/4/2031), PBO0009717(7/29/2030), PBO0008104(6/6/2030), PBO0006035(6/6/2030), PBO0009729(8/24/2031), PBO0008615(6/6/2030), PBO0008079(5/7/2030), PBO0008184(7/29/2030), PBO0009639(8/14/2031), PBO0009858(8/24/2031), PBO0004824(12/6/2029), PBO0008078(8/25/2030);
81-125-BL/M68981125030/PBO0010399(1/9/2029), PBO0006315(7/23/2029);
81-125-MC/M68981125030/PBO0006046(12/25/2028), PBO0006198(12/25/2028), PBO0007487(11/14/2028), PBO0007765(10/10/2027), PBO0008917(1/8/2029), PBO0010390(1/9/2029);
81-129-OD/M68981129060/PBO0010400(10/18/2029), PBO0010557(1/9/2029);
81-133-BL/M68981133080/PBO0004848 (9/5/2029);
81-133-CB/M68981133080/PBO0004849(9/11/2028);
81-133-MC/M68981133080/PBO0004850(9/11/2028);
81-135-BL/M68981135060/PBO0006810(1/23/2028), PBO0008450(12/31/2028);
81-135-MC/M68981135060/PBO0008217(12/31/2028), PBO0005853(5/9/2029), PBO0008944(11/28/2029);
81-139/M68981139010/PBO0005528(11/23/2027), PBO0010122(11/23/2027), PBO0007389(11/23/2027);
81-301/M68981301030/PBO0008729(1/8/2029), PBO0009012(1/8/2029), PBO0009785(1/9/2029), PBO0008634(1/8/2029);
81-303-01/M6898130301130/PBO0006031(1/23/2028), PBO0005527(10/3/2028), PBO0007324(1/23/2028), PBO0005163(12/31/2026), PBO0004331(7/17/2028), PBO0009143(12/31/2028), PBO0006564(1/23/2028), PBO0006871(1/23/2028), PBO0005598(11/28/2028), PBO0009321(12/31/2028), PBO0009438(12/31/2028), PBO0009417(9/25/2029), PBO0008167(12/31/2028), PBO0009959(9/25/2029), PBO0005343(9/11/2028), PBO0004484(8/14/2028), PBO0004620(8/14/2028), PBO0004672(8/14/2028), PBO0009328(2/8/2030), PBO0009607(9/25/2029), PBO0007584(12/31/2028), PBO0008805(12/31/2028), PBO0009258(12/31/2028), PBO0009286(11/28/2029), PBO0006327(1/23/2028), PBO0010076(9/25/2029), PBO0007790(12/31/2028), PBO0007810(12/31/2028), PBO0007890(12/31/2028), PBO0006841(1/23/2028), PBO0007937(12/31/2028), PBO0008111(12/31/2028), PBO0007628(12/31/2028), PBO0005321(9/11/2028), PBO0006493(1/23/2028), PBO0008992(11/28/2029), PBO0009051(12/31/2028), PBO0010410(9/25/2029), PBO0006908(1/23/2028), PBO0007234(1/23/2028), PBO0007272(1/23/2028), PBO0010508(9/25/2029), PBO0010580(9/25/2029), PBO0005838(12/25/2028), PBO0005885(5/28/2029), PBO0010009(9/25/2029), PBO0005597(11/28/2028), PBO0005508(10/3/2028), PBO0009296(12/31/2028), PBO0006165(1/23/2028), PBO0006010(1/23/2028), PBO0008084(12/31/2028);
81-303-30/M6898130330010/PBO0006441(1/23/2028), PBO0006326(1/23/2028), PBO0009138(12/31/2028), PBO0009293(12/31/2028), PBO0010158(9/25/2029), PBO0008112(12/31/2028), PBO0008085(12/31/2028), PBO0008072(12/31/2028), PBO0009165(12/31/2028), PBO0009832(9/25/2029), PBO0008439(12/31/2028), PBO0006599(1/23/2028), PBO0008836(12/31/2028), PBO0009329(2/8/2030), PBO0009685(9/25/2029), PBO0004465(8/14/2028), PBO0005391(10/3/2028), PBO0005394(10/3/2028), PBO0005681(10/3/2028), PBO0005750(12/27/2028), PBO0006001(5/28/2029), PBO0006581(1/23/2028), PBO0006649(1/23/2028), PBO0006736(1/23/2028), PBO0006822(1/23/2028), PBO0006872(1/23/2028), PBO0007106(1/23/2028), PBO0007251(1/23/2028), PBO0008825(12/31/2028), PBO0009022(11/28/2029), PBO0008872(12/31/2028), PBO0009740(9/25/2029), PBO0009782(9/25/2029), PBO0009768(9/25/2029), PBO0009831(9/25/2029), PBO0009937(12/31/2028), PBO0009918(9/25/2029), PBO0009951(9/25/2029), PBO0010073(9/25/2029), PBO0010097(9/25/2029), PBO0008957(12/31/2028), PBO0009958(9/25/2029), PBO0009642(9/25/2029), PBO0009664(9/25/2029), PBO0007597(1/23/2028), PBO0007627(12/31/2028), PBO0007988(12/31/2028), PBO0004412(1/20/2028), PBO0004125(12/31/2026), PBO0005322(11/19/2028), PBO0005887(12/27/2028), PBO0006066(1/23/2028), PBO0006117(12/25/2028), PBO0006011(5/28/2029), PBO0004673(8/14/2028), PBO0005169(12/31/2026), PBO0006139(1/23/2028), PBO0006116(5/28/2029), PBO0006227(1/23/2028), PBO0006272(1/23/2028), PBO0006273(1/23/2028), PBO0006191(1/23/2028), PBO0006339(1/23/2028), PBO0010746(9/25/2029), PBO0004359(12/31/2026), PBO0006379(1/23/2028), PBO0004737(7/17/2028), PBO0005576(10/3/2028), PBO0005680(11/28/2028), PBO0006777(1/23/2028), PBO0006669(1/23/2028);
81-605/M68981605020/PBO0005156(12/31/2026), PBO0006811(10/31/2027), PBO0008219(10/31/2027), PBO0004893(10/31/2025), PBO0009367(10/31/2027), PBO0008464(10/31/2027);
82-504-03/M6898250403050/PBO0006956(11/30/2025);
82-504-05/M6898250405060/PBO0006957(11/30/2025), PBO0008338(3/30/2026), PBO0006430(11/30/2025), PBO0010030(2/25/2028), PBO0005728(7/31/2026), PBO0008997(6/30/2026), PBO0005924(7/31/2026);
82-520/M68982520020/PBO0008664(10/29/2028), PBO0008767(12/31/2028);
83-303/M68983303040/PBO0009194(5/7/2030), PBO0010271(9/25/2029), PBO0005157(12/31/2026);
83-303-BL/M68983303040/PBO0004789(8/14/2028);
83-303-FG/M68983303040/PBO0004790(8/14/2028);
83-303-MC/M68983303040/PBO0007112(1/23/2028), PBO0008988(11/28/2029);
83-305-CB/M68983305040/PBO0004851(9/5/2029);
83-323-CB/M68983323100/PBO0004523(3/4/2024);
83-323-MC/M68983323100/PBO0004454(8/14/2028);
84-446-01-BL/M6898444601190, PBO0006368(4/3/2025), PBO0010159(2/18/2028), PBO0006836(6/13/2025), PBO0005726(5/31/2025), PBO0007075(11/30/2025), PBO0004998(10/31/2024);
84-446-01-CB/M6898444601190/PBO0005727(5/31/2025), PBO0007252(11/30/2025), PBO0009313(2/25/2028), PBO0010246(2/18/2028), PBO0008940(3/30/2026), PBO000642994/3/2025), PBO0004996(10/31/2024), PBO0006462(6/13/2025), PBO0006711(5/31/2025), PBO0008339(1/30/2026), PBO0007070(11/30/2025);
84-446-01-MC/M6898444601190/PBO0005202(10/31/2024), PBO0008340(1/30/2026), PBO0007071(11/30/2025), PBO0006428(4/3/2025), PBO0004997(10/31/2024), PBO0008423(1/30/2026), PBO0009303(2/25/2028), PBO0009314(2/25/2028), PBO0009719(2/18/2028), PBO0010160(2/18/2028), PBO0010493(2/25/2028);
84-497-CB/M68984497110/PBO0005812(4/30/2026);
84-497-FG/M68984497110/PBO0004990(10/31/2024);
84-497-MC/M68984497110/PBO0006832(11/30/2025), PBO0008367(3/30/2026);
87-155/M68987155030/PBO0004364(8/14/2028);
87-232/M68987232010/PBO0005412(8/31/2027), PBO0006639(12/31/2027);
88-105-01/M6898810501020/PBO0008466(5/7/2033);
88-105-03/M6898810503070/PBO0005280(11/1/2028), PBO0005908(12/25/2029), PBO0005998(2/12/2030), PBO0008259(10/18/2029);
88-108-01/M6898810801030/PBO0009562(11/23/2027);
88-108-02/M6898810802030/PBO0006951(11/23/2027), PBO0005757(11/23/2027), PBO0008782(11/23/2027), PBO0007373(11/23/2027);
88-108-03/M6898810803030/PBO0004612(11/23/2027), PBO0004942(11/23/2027), PBO0004525 (11/23/2027);
88-108-05/M6898810805010/PBO0006203(1/23/2028);
88-147/¿M68988147010/PBO0005267(11/23/2027), PBO0005095(11/23/2027), PBO0005937(11/23/2027), PBO0006812(11/23/2027);
SDR-EDC1OTC/M689SDREDC1OCT010/PBO0005185(5/16/2027), PBO0005872(5/16/2027);
SDR-EDC1-OTC/M689SDREDC1OTC050/PBO0004353(10/4/2026), PBO0004463(10/4/2026), PBO0004796(5/16/2027), PBO0004719(5/16/2027), PBO0004916(4/20/2027);
SDR-TBBOTC/M6898SDRTBB0TC020/PBO0005184(8/9/2027), PBO0005754(5/16/2027);
SDR-TBB-OTC/M689SDR-TBB-OTC/PBO0004797(5/16/2027), PBO0004718(5/16/2027), PBO0004915(5/16/2027);
SDR-TBBOTC-CB/M689SDR-TBB-OTC/PBO0005873(5/16/2027);
SG0999/COM/M689SG0999COM040/PBO0005076(12/31/2026), M689SG0999COM041/PBO0005331(8/9/2027), M689SG0999COM042/PBO0009635(8/9/2027), M689SG0999COM043/PBO0007404(8/9/2027), M689SG0999COM044/PBO0007568(8/9/2027), M689SG0999COM045/PBO0007899(8/9/2027), M689SG0999COM046/PBO0007897(8/9/2027), M689SG0999COM047/PBO0005744(5/16/2027), M689SG0999COM048/PBO0005777(5/16/2027),
|
| FEI Number |
3008258694
|
Recalling Firm/ Manufacturer |
Safeguard US Operating LLC
13359 Reese Blvd E
Huntersville
NC
28078-6591
|
| For Additional Information Contact | Kimberly Reed 904-322-3619 |
Manufacturer Reason for Recall | Compression bandage may be degraded, which may cause unwanted foreign material to be introduced inside the wound, which may lead to wound exploration and irrigation, formation of scar tissue, and infection. |
FDA Determined Cause 2 | Component change control |
| Action | On 9/4/2026, recall notices were mailed to customers who were asked to do the following:
1) Quarantine and remove all affected bandages from all points of use, including emergency kits, ambulances, storage and training areas.
2) If the affected bandage was purchased as a component of a kit, physically open the kit, remove the Battle Bandage and place into quarantine.
3) Complete and return the recall acknowledgement and receipt form via email to recalls@safeguardmedical.com
In addition, distributors were asked to:
Forward this information to any point of sale where you have sold the affected product(s).
Any questions directly associated with this recall should be directed to: recalls@safeguardmedical.com |
| Quantity in Commerce | 46,881 |
| Distribution | Worldwide - US Nationwide distribution including in the states of FL, OK, CA, WA, GA, TX, PA, IL, MD, NC, OR, DE, NJ, MI, RI, AR, WI, AZ, CO, IN, MO, OH, MA, NY, VA, KY, UT, NM, SC, TN, LA, AL, MN, MT, AK, WV, KS, NE, WY, MS, HI, NH, IA, DC, CT and the countries of BRA, BEL, SGP, AUS.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|