| | Class 2 Device Recall Arial Call Stations |  |
| Date Initiated by Firm | October 12, 2026 |
| Date Posted | November 12, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0438-2027 |
| Recall Event ID |
99965 |
| Product Classification |
Fall prevention alarm/sensor combination attached or unattached - Product Code PJO
|
| Product | Arial Call Stations
Model/Catalog Number: CSK200-900, CSK200-900MR, CSK200, CSK200MR, CSK200-1069, CSK200-1069MR
Software Version: V1.6.4 (FW) |
| Code Information |
UDI-DIs:
CSK200-1069MR: 00840948102707
CSK200-1069: 00840948102691
CSK200-MR: 00840948102684
CSK200: 00840948102677
CSK200-900MR: 00840948102660
CSK200-900: 00840948102653
Software Version: V1.6.4 (FW)
|
| FEI Number |
1929691
|
Recalling Firm/ Manufacturer |
Securitas Healthcare LLC
4600 Vine St
Lincoln
NE
68503-2823
|
| For Additional Information Contact | Ameya Oke 1-812-3717115 |
Manufacturer Reason for Recall | Call station batteries may fully deplete without a low battery alert. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On October 12, 2026, the firm posted the notification to their website at https://www.securitashealthcare.com/product-advisories. The firm sent letters to customers on 10/22/2026.
Customers were instructed on the following:
1. If you have Panasonic or have installed other non-approved batteries in your call stations, or have Panasonic or non-approved batteries in inventory for future use, replace them with Duracell, Jauch, or GP CR123 size batteries
2. Reduce the Missing Device Timeout setting to 125 minutes. This may provide earlier notification of a call station with a depleted battery on some Arial systems. Refer to Appendix 1 for instructions.
3. Regularly monitor alerts and battery status. "Missing Device" alerts should be promptly investigated.
4. Implement supplemental measures for high-risk residents to reduce the potential impact of an undetected battery depletion event.
|
| Quantity in Commerce | 42781 |
| Distribution | Distribution US nationwide, Canada, Israel, Republic of Korea, and United Arab Emirates. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|