| Date Initiated by Firm | September 08, 2026 |
| Date Posted | October 22, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0283-2027 |
| Recall Event ID |
99968 |
| Product Classification |
General surgery tray - Product Code LRO
|
| Product | Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30. |
| Code Information |
Model/Catalog Number: SOAH30. UDI-DI: 10809160520246. Lot Code: Lot Numbers: 1705826, 1705827.
Expiration Date: 1/31/2028.
|
| FEI Number |
1047429
|
Recalling Firm/ Manufacturer |
AVID Medical, Inc.
9000 Westmont Dr
Toano
VA
23168-9351
|
| For Additional Information Contact | Marlene Jones +1-470-2804277 |
Manufacturer Reason for Recall | A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications. |
FDA Determined Cause 2 | Process control |
| Action | AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to identify, segregate, and quarantine all impacted kits identified, and to not use the impacted component located in the piggybacked portion of the convenience kit. Customers were also instructed to add warning labels to affected kits to ensure all users are aware of the actions to remove and discard the impacted component prior to use. distributors were instructed to notify consignees if affected units were further distributed. Additionally, consignees were requested to complete and return the Recall Response Form. |
| Quantity in Commerce | 80 kits |
| Distribution | US Nationwide distributor in the state of Arizona. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|