• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Ventilation (Tympanostomy) Tubes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Ventilation (Tympanostomy) Tubessee related information
Date Initiated by FirmSeptember 24, 2026
Date PostedNovember 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0388-2027
Recall Event ID 100076
510(K)NumberK151067 
Product Classification Tube, tympanostomy - Product Code ETD
ProductBOBBIN VENTILATION TUBE 5PK 1.14 FLPL, REF: 1010201, Lot# 0231048277 (GTIN 20763000281612).
Code Information UDI-DI: 20763000281612. Lots: 0231048277
FEI Number 1045254
Recalling Firm/
Manufacturer
Medtronic Xomed, Inc.
6743 Southpoint Dr N
Jacksonville FL 32216-6218
For Additional Information ContactMarie Gentile
904-296-9600
Manufacturer Reason
for Recall
Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics.
FDA Determined
Cause 2
Process control
ActionOn 9/24/2026, recall notices were mailed and emailed to customers asking them to do the following: For patients who currently have implanted affected devices, as a precautionary measure, should continue to be monitored in accordance with standard clinical practices. 1) Identify, segregate, quarantine, and stop use of any affected products within your inventory. 2) Return affected unused product in your inventory to firm; see response form. 3) Share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Maintain a copy of this letter for your records. 4) Complete and return the Customer Confirmation Form via email to neuro.quality@medtronic.com For questions, contact your firm Representative or Customer Service at 800-874-5797 rs.jaxcustomerservice@medtronic.com
Quantity in Commerce2000
DistributionWorldwide - US Nationwide distribution including in the states of NY, AL, VA, MI, AZ, WA, TX, MA, IL, ME, NM, CA, IN, NC, OH, MO, MN, NH, IA, NV, GA, PA, FL, TN, WV, SC, OK, NE, OR, SD, AK, LA, CO, WI, MD, KS, UT, ID, WY, NJ and the countries of Canada, Australia, New Zealand, Hong Kong.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = ETD
-
-