• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Infinity Ammonia Standard Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Infinity Ammonia Standard Setsee related information
Date Initiated by FirmOctober 21, 2026
Date PostedNovember 23, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0533-2027
Recall Event ID 100164
510(K)NumberK012878 
Product Classification Electrode, ion-specific method, ammonia - Product Code JIG
ProductInfinity Ammonia Standard Set Model/Catalog Number: TR60901 Product Description: In Vitro Diagnostic calibrator and control
Code Information Lot Number: 382963, 364987, 307166 Expiration Date: Lot 307166 Exp. 9/30/2027, Lot 364987 Exp. 2/29/2028, Lot 382963 Exp. 5/31/2028 UDI: 00845275003314
FEI Number 1181121
Recalling Firm/
Manufacturer
Fisher Diagnostics
8365 Valley Pike
Middletown VA 22645-1905
Manufacturer Reason
for Recall
The ammonia reference standard used to manufacture the Infinity Ammonia Reagent was assigned an incorrect concentration, resulting in a low bias of approximately 6% to 15%. This low bias could lead to incorrect patient classification, delays in treatment initiation, and/or errors in clinical treatment decisions, which may cause patient harm.
FDA Determined
Cause 2
Process control
ActionThe firm began notifying customers on October 21, 2026. Users were instructed to inspect their inventory for affected lots of product. Affected product should be disposed of in accordance with local, regional, and national regulations. The recall notice should be passed to anyone within the customer's organization or any organization where the potentially affected lots have been transferred. If you have any additional questions regarding this recall, contact Fisher Diagnostics technical support by Toll Free 1800-528-0494 or by email: techsupport_diagnostics_mtn@thermofisher.com.
Quantity in Commerce132
DistributionWorldwide - US Nationwide distribution in the states and territory of CA, IL, PA, Puerto Rico and the countries of AUS, GER, KOR.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JIG
-
-