| | Class 2 Device Recall Infinity Ammonia Standard Set |  |
| Date Initiated by Firm | October 21, 2026 |
| Date Posted | November 23, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0533-2027 |
| Recall Event ID |
100164 |
| 510(K)Number | K012878 |
| Product Classification |
Electrode, ion-specific method, ammonia - Product Code JIG
|
| Product | Infinity Ammonia Standard Set
Model/Catalog Number: TR60901
Product Description: In Vitro Diagnostic calibrator and control |
| Code Information |
Lot Number: 382963, 364987, 307166
Expiration Date: Lot 307166 Exp. 9/30/2027, Lot 364987 Exp. 2/29/2028, Lot 382963 Exp. 5/31/2028
UDI: 00845275003314
|
| FEI Number |
1181121
|
Recalling Firm/ Manufacturer |
Fisher Diagnostics
8365 Valley Pike
Middletown
VA
22645-1905
|
Manufacturer Reason for Recall | The ammonia reference standard used to manufacture the Infinity Ammonia Reagent was assigned an incorrect concentration, resulting in a low bias of approximately 6% to 15%. This low bias could lead to incorrect patient classification, delays in treatment initiation, and/or errors in clinical treatment decisions, which may cause patient harm. |
FDA Determined Cause 2 | Process control |
| Action | The firm began notifying customers on October 21, 2026.
Users were instructed to inspect their inventory for affected lots of product. Affected product should be disposed of in accordance with local, regional, and national regulations. The recall notice should be passed to anyone within the customer's organization or any organization where the potentially affected lots have been transferred.
If you have any additional questions regarding this recall, contact Fisher Diagnostics technical support by Toll Free 1800-528-0494 or by email: techsupport_diagnostics_mtn@thermofisher.com. |
| Quantity in Commerce | 132 |
| Distribution | Worldwide - US Nationwide distribution in the states and territory of CA, IL, PA, Puerto Rico and the countries of AUS, GER, KOR. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JIG
|
|
|
|