| | Class 2 Device Recall Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers |  |
| Date Initiated by Firm | October 21, 2026 |
| Date Posted | November 23, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0534-2027 |
| Recall Event ID |
100164 |
| 510(K)Number | K974620 |
| Product Classification |
Enzymatic method, ammonia - Product Code JIF
|
| Product | Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers
Model/Catalog Number: OSR61154
Product Description: In Vitro Diagnostic reagent, calibrator, and control |
| Code Information |
Lot Number: 2915, 2916, 9511, 9510, 0644, 0645, 0614, 0613, 2016, 5301, 9901, 2525, 7602, 7603, 4111, 2393, 7411, 6624, 3223, 3231, 7779, 7780, 0206, 0209
Expiration Date: Lot 2915 Exp. 03/01/2028, Lot 2916 Exp. 03/01/2028, Lot 9511 Exp. 02/01/2028, Lot 9510 Exp. 02/01/2028, Lot 0644 Exp. 12/01/2027, Lot 0645 Exp. 12/01/2027, Lot 0614 Exp. 11/01/2027, Lot 0613 Exp. 11/01/2027, Lot 2016 Exp. 10/01/2027, Lot 5301 Exp. 10/01/2027, Lot 9901 Exp. 08/01/2027, Lot 2525 Exp. 08/01/2027, Lot 7602 Exp. 07/01/2027, Lot 7603 Exp. 07/01/2027, Lot 4111 Exp. 05/01/2027, Lot 2393 Exp. 05/01/2027, Lot 7411 Exp. 03/01/2027, Lot 6624 Exp. 03/01/2027, Lot 3223 Exp. 02/01/2027, Lot 3231 Exp. 02/01/2027, Lot 7779 Exp. 01/01/2027, Lot 7780 Exp. 01/01/2027, Lot 0206 Exp. 01/01/2027, Lot 0209 Exp. 12/01/2026
UDI: 00845275000016
|
| FEI Number |
1181121
|
Recalling Firm/ Manufacturer |
Fisher Diagnostics
8365 Valley Pike
Middletown
VA
22645-1905
|
Manufacturer Reason for Recall | The ammonia reference standard used to manufacture the Infinity Ammonia Reagent was assigned an incorrect concentration, resulting in a low bias of approximately 6% to 15%. This low bias could lead to incorrect patient classification, delays in treatment initiation, and/or errors in clinical treatment decisions, which may cause patient harm. |
FDA Determined Cause 2 | Process control |
| Action | The firm began notifying customers on October 21, 2026.
Users were instructed to inspect their inventory for affected lots of product. Affected product should be disposed of in accordance with local, regional, and national regulations. The recall notice should be passed to anyone within the customer's organization or any organization where the potentially affected lots have been transferred.
If you have any additional questions regarding this recall, contact Fisher Diagnostics technical support by Toll Free 1800-528-0494 or by email: techsupport_diagnostics_mtn@thermofisher.com. |
| Quantity in Commerce | 75420 |
| Distribution | Worldwide - US Nationwide distribution in the states and territory of CA, IL, PA, Puerto Rico and the countries of AUS, GER, KOR. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|