 Class 2 Device Recall Baxter 1550 Hemodialysis Machinesee related information
Date Initiated by FirmFebruary 20, 2003
Date PostedMarch 20, 2003
Recall Status1 Terminated 3 on March 27, 2006
Recall NumberZ-0661-03
Recall Event ID 25607
510(K)NumberK883111 
Product Classification System, Dialysate Delivery, Single Patient - Product Code FKP
ProductBaxter 1550 Single Patient System Hemodialysis Machines; Baxter Healthcare Corporation, Deerfield, IL 60015
Code Information catalog #5M5538, 5M5538R, 5M5551, 5M5551R, 5M5575, FM4719, FM4774, FM4779, FM4779R; all serial numbers
FEI Number 1416980
Recalling Firm/
Manufacturer
Baxter Healthcare Corp.
Rt 120 & Wilson Rd
Round Lake IL 60073
For Additional Information ContactCenter for One Baxter
800-422-9837
Manufacturer Reason
for Recall
Excess or insufficient ultrafiltration may occur during manual control of ultrafiltration via adjustment of transmembrane pressure (TMP)
FDA Determined
Cause 2
Other
ActionBaxter sent Safety Alert letters dated 2/20/03 to the end users, informing them of the possibility of either excess or insufficient ultrafiltration when the operator manually controls ultrafiltration via adjustment of transmembrane pressure. The accounts were informed that turning off the UF controller during fault codes FL-04, FL-07 and FL-08 may affect the ability to safely perform manual ultrafiltration, and that treatment should be discontinued if those three fault codes are displayed. Changes to the instructions for use are in process and will be sent to the end users when completed. Additionally, the Baxter 1550 has a side panel that contains a 16-digit display. Currently, during all ultrafiltration system fault modes the display indicates 'Manual UF Control'. To avoid operator confusion, new software is being developed that will provide a display of 'Discontinue TX' when a FL-04, FL-07 or FL-08 fault code occurs. The SPS 550 does not have the side panel display and does not require the software upgrade.
DistributionNationwide and internationally to Algeria, AMerican Samoa, Argentina, Aruba, Australia, Barbados, Belgium, Bolivia, Brazil, Brunei, Burma, Canada, Chile, China, Colombia, Croatia, Cyprus, Dominican Republic, Ecuador, Egypt, El Salvador, Yugoslavia, Germany, Ghana, Guam, Guatemala, Honduras, Hong Kong, Hungary, INdia, Indonesia, Israel, Italy, Jamaica, Jordan, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Romania, Russia, South Korea, Singapore, Slovakia, Slovenia, South Africa, Spain, St. Maarten, Taiwan, Thailand, Tobago, Trinidad, Turkey, United Kingdom, Uruguay, Venezuela and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FKP
-
-