| | Class 2 Device Recall EBI |  |
| Date Initiated by Firm | June 17, 2003 |
| Date Posted | July 15, 2003 |
| Recall Status1 |
Terminated 3 on December 31, 2003 |
| Recall Number | Z-1006-03 |
| Recall Event ID |
26512 |
| 510(K)Number | K023233 K971083 K982484 K010179 |
| Product Classification |
Endoscope, Neurological - Product Code GWG
|
| Product | EBI VueCath Endoscopic Spinal System |
| Code Information |
Catalog Number: S002 Lot/Serial Nos: F10021, F10017, F10012, F10018, F10025, F10019, F10020, F10014, F10023, F10022, F10001, F10000, F10024, F10007, F10008, F00005, F10003, F10016, F10015, F10002, F10027. |
| FEI Number |
2242816
|
Recalling Firm/ Manufacturer |
EBI, L.P.
100 Interpace Parkway
Parsippany
NJ
07054-1079
|
| For Additional Information Contact | Barry Sands 973-331-3904 |
Manufacturer Reason for Recall | Non sterile device is labeled sterile. |
FDA Determined Cause 2 | Labeling False and Misleading |
| Action | The firm notified their customers via a field correction letter dated June 17, 2003. Instructions given are to sterilize the unit prior to use. A EBI field representative will be visiting to replace the boxes. |
| Quantity in Commerce | 39 |
| Distribution | Hospitals, clinics, and doctors nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = GWG 510(K)s with Product Code = GWG
|
|
|
|