| | Class 2 Device Recall |  |
| Date Initiated by Firm | February 20, 2002 |
| Date Posted | September 12, 2003 |
| Recall Status1 |
Terminated 3 on April 11, 2005 |
| Recall Number | Z-1204-03 |
| Recall Event ID |
26581 |
| 510(K)Number | K912470 |
| Product Classification |
Table, Radiographic, Non-Tilting, Powered - Product Code IZZ
|
| Product | Multidiagnost 96 |
| Code Information |
The firm utilizes a site number specific to each unit: 17197, 17368, 17325, 76449, 17247, 17339, 17304, 13444, 6616, 6558, 6559, 6426, 6539, 6541, 6583, 35366, 37859, 31000, 67642, 62363, 62362, 62278, 62300, 25769, 25901, 25884, 25768, 25900, 25888, 25834, 25967, 25710, 25907, 62279, 62310, 62403, 62391, 62309, 62404, 73921, 50135, 44742, 44670, 44703, 44760, 44699, 46921, 52354 |
| FEI Number |
1000524572
|
Recalling Firm/ Manufacturer |
Philips Medical Systems
22100 Bothell Everett Highway
Bothell
WA
98041
|
| For Additional Information Contact | 425-487-7000 |
Manufacturer Reason for Recall | Spontaneous movement of table without input from an operator |
FDA Determined Cause 2 | Other |
| Action | On 2/20/02 the firm issued Field Change Order to their field personnel, instructing them to visit each site and upgrade each unit. |
| Quantity in Commerce | 48 units |
| Distribution | The firm distributed units to hospitals and medical centers located throughout the US. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = IZZ
|
|
|
|