| Date Initiated by Firm | June 19, 2003 |
| Date Posted | August 06, 2003 |
| Recall Status1 |
Terminated 3 on December 01, 2004 |
| Recall Number | Z-1077-03 |
| Recall Event ID |
26722 |
| Product | Synchron LX Systems Primary Tube Sample Template
(a piece of ''labeling'' for use with the Synchron LX 20 and LX 20 Pro systems).
Part Number: 967178 |
| Code Information |
No range of serial numbers apply. |
| FEI Number |
2050012
|
Recalling Firm/ Manufacturer |
Beckman Coulter Inc
200 South Kraemer Blvd W 337
Brea
CA
92822
|
| For Additional Information Contact | Kathleen M. Jaker 714-961-3666 |
Manufacturer Reason for Recall | Dead volume on sight gauge for the Synchron LX 20 and LX 20 Pro Systems not stated, which may result in instrument error. |
FDA Determined Cause 2 | Other |
| Action | A revised template with a recall letter was sent to customers with a response request on 19 June 2003. |
| Quantity in Commerce | 502 |
| Distribution | USA and Canada. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|