 Class 3 Device Recall AxSYM Folate Specimen Diluentsee related information
Date Initiated by FirmDecember 16, 2003
Date PostedJuly 20, 2004
Recall Status1 Terminated 3 on July 20, 2004
Recall NumberZ-0365-04
Recall Event ID 27978
510(K)NumberK972232 
Product Classification unknown device name - Product Code CGN--
ProductAxSYM Folate Specimen Diluent, list 3C81-50, 10 mL bottles containing TRIS buffer with protein stabilizer; Abbott Laboratories, Abbott Park, IL 60064 USA
Code Information List 3C81-50, lots 91555M100, 05920M200
FEI Number 1415939
Recalling Firm/
Manufacturer
Abbott Laboratories HPD/ADD/GPRD
100/200 Abbott Park Road
Abbott Park IL 60064
Manufacturer Reason
for Recall
The is an increase in the number of control values out of range low, and diluted patient specimens may yield lower than expected results.
FDA Determined
Cause 2
Other
ActionAbbott notified their international affiliates via e-mail on 12/16/03 and 1/5/04. Device recall letters dated 12/16/03 and 12/31/03 will be sent to the end users by the affiliates, informing them of the unapproved serum albumin used to manufacture the folate products, and listing the impact to performance for each affected components. There are separate customer recall letters for each assay platform -- ARCHITECT, AxSYM and IMx.
Quantity in Commerce375 kits
DistributionThere was no U.S. distribution. The products were shipped internationally through Abbott affiliates in Canada, Germany, Singapore, Hong Kong, Australia, Brazil, New Zealand, Columbia, Chile, Uruguay, United Kingdom, Venezuela, Dominican Republic, Argentina and Curacao.