| Date Initiated by Firm | February 20, 2004 |
| Date Posted | July 20, 2004 |
| Recall Status1 |
Terminated 3 on May 28, 2004 |
| Recall Number | Z-0728-04 |
| Recall Event ID |
28438 |
| 510(K)Number | K001341 |
| Product Classification |
System, X-Ray, Stationary - Product Code KPR
|
| Product | Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit. |
| Code Information |
Serial Numbers affected: 4193-5702, 4193-5704, 4193-5705, 4193-5706, 4193-5708 |
| FEI Number |
1315356
|
Recalling Firm/ Manufacturer |
Eastman Kodak Co 343 State St Rochester NY 14650-1131
|
| For Additional Information Contact | James Burns 585-726-6383 |
Manufacturer Reason for Recall | Defective component in overhead tube support balancer may cause the suspended load to fall. |
FDA Determined Cause 2 | Other |
| Action | 2/20/2004: Field representatives notified the customer in China that the unit should not be used until the OTS Balancer component can be repaired/replaced. |
| Quantity in Commerce | 3 units |
| Distribution | Amsterdam, China and Kuwait. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KPR
|