| Date Initiated by Firm | July 20, 2004 |
| Date Posted | September 15, 2004 |
| Recall Status1 |
Terminated 3 on February 09, 2006 |
| Recall Number | Z-1454-04 |
| Recall Event ID |
29855 |
| 510(K)Number | K921255 |
| Product Classification |
System, X-Ray, Mobile - Product Code IZL
|
| Product | BV 25 Gold Mobile X-Ray System |
| Code Information |
see site numbers listed above for BV 25 |
Recalling Firm/ Manufacturer |
Philips Medical Systems Sales & Service Region No. America 22100 Bothell Everett Hwy Bothell WA 98021
|
| For Additional Information Contact | Sarah Baxter 425-487-7665 |
Manufacturer Reason for Recall | Potential for c-arm to move due to loose screws |
FDA Determined Cause 2 | Other |
| Action | Beginning on 7/20/04 the firm sent a letter dated 7/19/04 via certified mail. Mailing was completed 7/27/04. The letter advises customers of the problem, advises how to determine of the screws are loose, and if so to contact Philips. The letter also advises that a respresentative will visit each consignee to inspect and correct the problem. |
| Quantity in Commerce | 1900 total devices (includes BV25, BV 25 Gold, BV26, BV29, BV212 |
| Distribution | The units are distributed to hospitals and medical centers throughout the U.S. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IZL
|