| | Class 2 Device Recall Tube Stand for the Dynarad Phantom Portable XRay System |  |
| Date Initiated by Firm | March 10, 2005 |
| Date Posted | May 11, 2005 |
| Recall Status1 |
Terminated 3 on November 14, 2005 |
| Recall Number | Z-0816-05 |
| Recall Event ID |
31456 |
| Product Classification |
Tube Mount, X-Ray, Diagnostic - Product Code IYB
|
| Product | Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131 |
| Code Information |
All Phantom Model Tube Stands with squared corners on all four corners of the stand. Catalog #501031, serial numbers 00743-0204, 00778-0204, 00997-0304, 01001-0304, 01103-0304, 01293-0304, 01371-0304, 01425-0404, 01452-0404, 01470-0404, 01475-0404, 01575-0404, 01592-0404, 01700-0404, 01704-0404, 01790-0404, 01841-0404, 01851-0404, 01898-0504, 02019-0504, 02023-0504, 02051-0504, 02096-0504, 02178-0504, 02188-0504, 02310-0604, 02357-0604, 02429-0604, 02517-0604, 02534-0604, 02584-0604, 02671-0604, 02752-0704, 02760-0704, 02776-0704, 02839-0704, 02891-0704, 02928-0704, 02938-0704, 02945-0704, 02959-0704, 02980-0704, 02991-0704, 03076-0704, 03084-0704, 03122-0704, 03375-0804, 03455-0804, 03476-0804, 03518-0804, 03543-0804, 03570-0904, 03579-0904, 03636-0904, 03700-0904, 03708-0904, 03722-0904, 03740-0904, 03952-0904, 04037-0904, 04134-1004, 04183-1004, 04347-1004, 04380-1004, 04407-1004, 04477-1004, 04481-1004, 04491-1004, 04524-1004, 04534-1004, 04592-1104, 04600-1104, 04605-1104, 04648-1104, 04694-1104, 04699-1104, 04703-1104, 0439-1104, 04783-1104, 04862-1104, 05190-1204, 05194-1204, 05198-1204, 05216-1204, 05364-1204, 05414-1204, 05418-1204, 05439-1204, 00283-0105, 00340-0105, 00481-0105, 00571-0205, 00575-0205, 00617-0205, 00649-0205, 00763-0205, 00841-0205, 00858-0205, 00891-0205, 00900-0205, 00913-0205 |
| FEI Number |
1418964
|
Recalling Firm/ Manufacturer |
Del Medical Systems Group 11550 West King Street Franklin Park IL 60131-1330
|
| For Additional Information Contact | Ms. Veronica Meridith 800-800-6006 |
Manufacturer Reason for Recall | The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator. |
FDA Determined Cause 2 | Other |
| Action | Del Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired. |
| Quantity in Commerce | 31 stands |
| Distribution | Nationwide, Mexico and Thailand. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|