| Date Initiated by Firm | July 22, 2005 |
| Date Posted | September 02, 2005 |
| Recall Status1 |
Terminated 3 on June 20, 2007 |
| Recall Number | Z-1495-05 |
| Recall Event ID |
32694 |
| 510(K)Number | K032712 |
| Product Classification |
Catheter, Suprapubic (And Accessories) - Product Code KOB
|
| Product | Vaxcel Plus Chronic Dialysis Catheters - Catheter/CPU/15/28/Metal
Catalog Number : 45-548 |
| Code Information |
Lot Numbers: 986995 993827 1000956 1001291 1009508 1013920 1021335 1022504 1027123 |
| FEI Number |
3001451463
|
Recalling Firm/ Manufacturer |
Boston Scientific Corporation 1 Boston Scientific Pl Natick MA 01760-1536
|
| For Additional Information Contact | Cindy Finney 508-683-4678 |
Manufacturer Reason for Recall | The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use. |
FDA Determined Cause 2 | Other |
| Action | Boston Scientific intiated the recall on 7/22/05 to direct accounts and to distributors on 7/ 27 /05. Distributors are requested to conduct sub recalls to the medical facility level. A revised letter dated 8/2/05 was sent to 18 US accounts for one lot (#1020895) omitted in the letter of 7/22/05. |
| Distribution | Nationwide
VA Hopsitals
Foreign: Canada, Africa, Japan, Israel, Japan, Nertherlands, Venezula |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KOB
|